Repeatability and Reproducibility of a Double-Pass Optical Quality Analysis Device
PURPOSE: To evaluate the repeatability and reproducibility of a double-pass instrument (OQASII, Visiomereics SL, Spain), which objectively measures overall optical quality of the human eyes. METHODS: The right eye of 119 healthy subjects with best corrected visual acuity of 20/25 or better was inclu...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4315520/ https://www.ncbi.nlm.nih.gov/pubmed/25646846 http://dx.doi.org/10.1371/journal.pone.0117587 |
_version_ | 1782355486492852224 |
---|---|
author | Xu, Chen-Chen Xue, Ting Wang, Qin-Mei Zhou, Yi-Ni Huang, Jin-Hai Yu, A-Yong |
author_facet | Xu, Chen-Chen Xue, Ting Wang, Qin-Mei Zhou, Yi-Ni Huang, Jin-Hai Yu, A-Yong |
author_sort | Xu, Chen-Chen |
collection | PubMed |
description | PURPOSE: To evaluate the repeatability and reproducibility of a double-pass instrument (OQASII, Visiomereics SL, Spain), which objectively measures overall optical quality of the human eyes. METHODS: The right eye of 119 healthy subjects with best corrected visual acuity of 20/25 or better was included in this prospective, comparative, observational study. Two separate tests with OQASII were conducted sequentially on the same day by two different examiners. A week later, the first examiner conducted the third measurement. All subjects underwent three consecutive tests during each session. The repeatability and reproducibility of the modulation transfer function cut off frequency (MTF cutoff), the Strehl ratio, the OQAS values (OVs) at contrasts of 100%, 20% and 9%, and the objective scatter index (OSI) were analyzed. RESULTS: For MTF cutoff, Strehl ratio, OV100%, OV20%, OV9%, and OSI, the mean values were 39.32±9.75cpd, 0.22±0.06, 1.31±0.33, 1.33±0.39, 1.33±0.41, 0.60±0.42, respectively. Repeatability and reproducibility were good with a very low coefficient of variation and high interclass correlation coefficients (>0.88) for all parameters. Bland-Altman plots showed good correlation with 95% limits of agreement ranged from -6.04 to 6.78cpd, -0.05 to 0.05, -0.20 to 0.23, -0.29 to 0.32, -0.40 to 0.42, -0.23 to 0.21 in inter-observer, and -6.56 to 7.42cpd, -0.06 to 0.06, -0.22 to 0.24, -0.30 to 0.32, -0.35 to 0.34, -0.24 to 0.23 in inter-visit, respectively. CONCLUSION: The OQASII system yields excellent repeatability and good reproducibility for objective measurements of overall optical quality in clinic. |
format | Online Article Text |
id | pubmed-4315520 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-43155202015-02-13 Repeatability and Reproducibility of a Double-Pass Optical Quality Analysis Device Xu, Chen-Chen Xue, Ting Wang, Qin-Mei Zhou, Yi-Ni Huang, Jin-Hai Yu, A-Yong PLoS One Research Article PURPOSE: To evaluate the repeatability and reproducibility of a double-pass instrument (OQASII, Visiomereics SL, Spain), which objectively measures overall optical quality of the human eyes. METHODS: The right eye of 119 healthy subjects with best corrected visual acuity of 20/25 or better was included in this prospective, comparative, observational study. Two separate tests with OQASII were conducted sequentially on the same day by two different examiners. A week later, the first examiner conducted the third measurement. All subjects underwent three consecutive tests during each session. The repeatability and reproducibility of the modulation transfer function cut off frequency (MTF cutoff), the Strehl ratio, the OQAS values (OVs) at contrasts of 100%, 20% and 9%, and the objective scatter index (OSI) were analyzed. RESULTS: For MTF cutoff, Strehl ratio, OV100%, OV20%, OV9%, and OSI, the mean values were 39.32±9.75cpd, 0.22±0.06, 1.31±0.33, 1.33±0.39, 1.33±0.41, 0.60±0.42, respectively. Repeatability and reproducibility were good with a very low coefficient of variation and high interclass correlation coefficients (>0.88) for all parameters. Bland-Altman plots showed good correlation with 95% limits of agreement ranged from -6.04 to 6.78cpd, -0.05 to 0.05, -0.20 to 0.23, -0.29 to 0.32, -0.40 to 0.42, -0.23 to 0.21 in inter-observer, and -6.56 to 7.42cpd, -0.06 to 0.06, -0.22 to 0.24, -0.30 to 0.32, -0.35 to 0.34, -0.24 to 0.23 in inter-visit, respectively. CONCLUSION: The OQASII system yields excellent repeatability and good reproducibility for objective measurements of overall optical quality in clinic. Public Library of Science 2015-02-03 /pmc/articles/PMC4315520/ /pubmed/25646846 http://dx.doi.org/10.1371/journal.pone.0117587 Text en © 2015 Xu et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited. |
spellingShingle | Research Article Xu, Chen-Chen Xue, Ting Wang, Qin-Mei Zhou, Yi-Ni Huang, Jin-Hai Yu, A-Yong Repeatability and Reproducibility of a Double-Pass Optical Quality Analysis Device |
title | Repeatability and Reproducibility of a Double-Pass Optical Quality Analysis Device |
title_full | Repeatability and Reproducibility of a Double-Pass Optical Quality Analysis Device |
title_fullStr | Repeatability and Reproducibility of a Double-Pass Optical Quality Analysis Device |
title_full_unstemmed | Repeatability and Reproducibility of a Double-Pass Optical Quality Analysis Device |
title_short | Repeatability and Reproducibility of a Double-Pass Optical Quality Analysis Device |
title_sort | repeatability and reproducibility of a double-pass optical quality analysis device |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4315520/ https://www.ncbi.nlm.nih.gov/pubmed/25646846 http://dx.doi.org/10.1371/journal.pone.0117587 |
work_keys_str_mv | AT xuchenchen repeatabilityandreproducibilityofadoublepassopticalqualityanalysisdevice AT xueting repeatabilityandreproducibilityofadoublepassopticalqualityanalysisdevice AT wangqinmei repeatabilityandreproducibilityofadoublepassopticalqualityanalysisdevice AT zhouyini repeatabilityandreproducibilityofadoublepassopticalqualityanalysisdevice AT huangjinhai repeatabilityandreproducibilityofadoublepassopticalqualityanalysisdevice AT yuayong repeatabilityandreproducibilityofadoublepassopticalqualityanalysisdevice |