Quantitative analysis of high-resolution, contrast-enhanced, cone-beam CT for the detection of intracranial in-stent hyperplasia

BACKGROUND: Intracranial in-stent hyperplasia is a stroke-associated complication that requires routine surveillance. OBJECTIVE: To compare the results of in vivo experiments to determine the accuracy and precision of in-stent hyperplasia measurements obtained with modified C-arm contrast-enhanced,...

Descripción completa

Detalles Bibliográficos
Autores principales: Flood, Thomas F, van der Bom, Imramsjah M J, Strittmatter, Lara, Puri, Ajit S, Hendricks, Gregory M, Wakhloo, Ajay K, Gounis, Matthew J
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BMJ Publishing Group 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4316917/
https://www.ncbi.nlm.nih.gov/pubmed/24480728
http://dx.doi.org/10.1136/neurintsurg-2013-010950
_version_ 1782355639653105664
author Flood, Thomas F
van der Bom, Imramsjah M J
Strittmatter, Lara
Puri, Ajit S
Hendricks, Gregory M
Wakhloo, Ajay K
Gounis, Matthew J
author_facet Flood, Thomas F
van der Bom, Imramsjah M J
Strittmatter, Lara
Puri, Ajit S
Hendricks, Gregory M
Wakhloo, Ajay K
Gounis, Matthew J
author_sort Flood, Thomas F
collection PubMed
description BACKGROUND: Intracranial in-stent hyperplasia is a stroke-associated complication that requires routine surveillance. OBJECTIVE: To compare the results of in vivo experiments to determine the accuracy and precision of in-stent hyperplasia measurements obtained with modified C-arm contrast-enhanced, cone-beam CT (CE-CBCT) imaging with those obtained by ‘gold standard’ histomorphometry. Additionally, to carry out clinical analyses comparing this CE-CBCT protocol with digital subtraction angiography (DSA). METHODS: A non-binned CE-CBCT protocol (VasoCT) was used that acquires x-ray images with a small field-of-view and applies a full-scale reconstruction algorithm providing high-resolution three-dimensional (3D) imaging with 100 µm isotropic voxels. In an vivo porcine model, VasoCT cross-sectional area measurements were compared with gold standard vessel histology. VasoCT and DSA were used to calculate in-stent stenosis in 23 imaging studies. RESULTS: Porcine VasoCT cross-sectional stent, lumen, and in-stent hyperplasia areas strongly correlated with histological measurements (r(2)=0.97, 0.93, 0.90; slope=1.14, 1.07, and 0.76, respectively; p<0.0001). Clinical VasoCT percentage stenosis correlated well with DSA percentage stenosis (r(2)=0.84; slope=0.76), and the two techniques were free of consistent bias (Bland–Altman, bias=3.29%; 95% CI −14.75% to 21.33%). An illustrative clinical case demonstrated the advantages of VasoCT, including 3D capability and non-invasive IV contrast administration, for detection of in-stent hyperplasia. CONCLUSIONS: C-arm VasoCT is a high-resolution 3D capable imaging technique that has been validated in an animal model for measurement of in-stent tissue growth. Successful clinical implementation of the protocol was performed in a small case series.
format Online
Article
Text
id pubmed-4316917
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher BMJ Publishing Group
record_format MEDLINE/PubMed
spelling pubmed-43169172015-02-11 Quantitative analysis of high-resolution, contrast-enhanced, cone-beam CT for the detection of intracranial in-stent hyperplasia Flood, Thomas F van der Bom, Imramsjah M J Strittmatter, Lara Puri, Ajit S Hendricks, Gregory M Wakhloo, Ajay K Gounis, Matthew J J Neurointerv Surg Basic Science BACKGROUND: Intracranial in-stent hyperplasia is a stroke-associated complication that requires routine surveillance. OBJECTIVE: To compare the results of in vivo experiments to determine the accuracy and precision of in-stent hyperplasia measurements obtained with modified C-arm contrast-enhanced, cone-beam CT (CE-CBCT) imaging with those obtained by ‘gold standard’ histomorphometry. Additionally, to carry out clinical analyses comparing this CE-CBCT protocol with digital subtraction angiography (DSA). METHODS: A non-binned CE-CBCT protocol (VasoCT) was used that acquires x-ray images with a small field-of-view and applies a full-scale reconstruction algorithm providing high-resolution three-dimensional (3D) imaging with 100 µm isotropic voxels. In an vivo porcine model, VasoCT cross-sectional area measurements were compared with gold standard vessel histology. VasoCT and DSA were used to calculate in-stent stenosis in 23 imaging studies. RESULTS: Porcine VasoCT cross-sectional stent, lumen, and in-stent hyperplasia areas strongly correlated with histological measurements (r(2)=0.97, 0.93, 0.90; slope=1.14, 1.07, and 0.76, respectively; p<0.0001). Clinical VasoCT percentage stenosis correlated well with DSA percentage stenosis (r(2)=0.84; slope=0.76), and the two techniques were free of consistent bias (Bland–Altman, bias=3.29%; 95% CI −14.75% to 21.33%). An illustrative clinical case demonstrated the advantages of VasoCT, including 3D capability and non-invasive IV contrast administration, for detection of in-stent hyperplasia. CONCLUSIONS: C-arm VasoCT is a high-resolution 3D capable imaging technique that has been validated in an animal model for measurement of in-stent tissue growth. Successful clinical implementation of the protocol was performed in a small case series. BMJ Publishing Group 2015-02 2014-01-30 /pmc/articles/PMC4316917/ /pubmed/24480728 http://dx.doi.org/10.1136/neurintsurg-2013-010950 Text en Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://group.bmj.com/group/rights-licensing/permissions This is an Open Access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 3.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/3.0/
spellingShingle Basic Science
Flood, Thomas F
van der Bom, Imramsjah M J
Strittmatter, Lara
Puri, Ajit S
Hendricks, Gregory M
Wakhloo, Ajay K
Gounis, Matthew J
Quantitative analysis of high-resolution, contrast-enhanced, cone-beam CT for the detection of intracranial in-stent hyperplasia
title Quantitative analysis of high-resolution, contrast-enhanced, cone-beam CT for the detection of intracranial in-stent hyperplasia
title_full Quantitative analysis of high-resolution, contrast-enhanced, cone-beam CT for the detection of intracranial in-stent hyperplasia
title_fullStr Quantitative analysis of high-resolution, contrast-enhanced, cone-beam CT for the detection of intracranial in-stent hyperplasia
title_full_unstemmed Quantitative analysis of high-resolution, contrast-enhanced, cone-beam CT for the detection of intracranial in-stent hyperplasia
title_short Quantitative analysis of high-resolution, contrast-enhanced, cone-beam CT for the detection of intracranial in-stent hyperplasia
title_sort quantitative analysis of high-resolution, contrast-enhanced, cone-beam ct for the detection of intracranial in-stent hyperplasia
topic Basic Science
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4316917/
https://www.ncbi.nlm.nih.gov/pubmed/24480728
http://dx.doi.org/10.1136/neurintsurg-2013-010950
work_keys_str_mv AT floodthomasf quantitativeanalysisofhighresolutioncontrastenhancedconebeamctforthedetectionofintracranialinstenthyperplasia
AT vanderbomimramsjahmj quantitativeanalysisofhighresolutioncontrastenhancedconebeamctforthedetectionofintracranialinstenthyperplasia
AT strittmatterlara quantitativeanalysisofhighresolutioncontrastenhancedconebeamctforthedetectionofintracranialinstenthyperplasia
AT puriajits quantitativeanalysisofhighresolutioncontrastenhancedconebeamctforthedetectionofintracranialinstenthyperplasia
AT hendricksgregorym quantitativeanalysisofhighresolutioncontrastenhancedconebeamctforthedetectionofintracranialinstenthyperplasia
AT wakhlooajayk quantitativeanalysisofhighresolutioncontrastenhancedconebeamctforthedetectionofintracranialinstenthyperplasia
AT gounismatthewj quantitativeanalysisofhighresolutioncontrastenhancedconebeamctforthedetectionofintracranialinstenthyperplasia