Calibration of an automated California mastitis test with focus on the device-dependent variation

The aim of the paper was to estimate the accuracy of the metrology of an installed indirect on-line sensor system based on the automated California Mastitis Test (CMT) with focus on the prior established device-dependent variation. A sensor calibration was implemented. Therefore, seven sensors were...

Descripción completa

Detalles Bibliográficos
Autores principales: Neitzel, Anne-Christin, Stamer, Eckhard, Junge, Wolfgang, Thaller, Georg
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer International Publishing 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4320165/
https://www.ncbi.nlm.nih.gov/pubmed/25674485
http://dx.doi.org/10.1186/2193-1801-3-760
Descripción
Sumario:The aim of the paper was to estimate the accuracy of the metrology of an installed indirect on-line sensor system based on the automated California Mastitis Test (CMT) with focus on the prior established device-dependent variation. A sensor calibration was implemented. Therefore, seven sensors were tested with similar trials on the dairy research farm Karkendamm (Germany) on two days in July 2011 and January 2012. Thereby, 18 mixed milk samples from serial dilutions were fourfold recorded at every sensor. For the validation, independent sensor records with corresponding lab somatic cell score records (LSCS) in the period between both trials were used (n = 1,357). From these records for each sensor a polynomial regression function was calculated. The predicted SCS (PSCS) was obtained for each sensor with the previously determined regression coefficients. Pearson correlation coefficients between PSCS and LSCS were established for each sensor and ranged between r = 0.57 and r = 0.67. Comparing the results with the correlation coefficients between the on-line SCS (OSCS) and the LSCS (r = 0.20 to 0.57) for every sensor, the calibration showed the tendency to improve the installed sensor system.