Cargando…

FRACTURE OF ABUTMENT SCREW SUPPORTING A CEMENTED IMPLANT-RETAINED PROSTHESIS WITH EXTERNAL HEXAGON CONNECTION: A CASE REPORT WITH SEM EVALUATION

One of the causes of implant failures in cemented implant-retained prostheses is the fracture of abutment screw or UCLA abutment. This article reports a case of simultaneous fracture of two UCLA abutments screws occurring in an implant-supported prosthesis placed in the mandibular molar region. The...

Descripción completa

Detalles Bibliográficos
Autores principales: Santos, Mariana Diniz Bisi, Pfeifer, Arthur Braga, Silva, Marcos Rogério Pupo, Sendyk, Claudio Luiz, Sendyk, WIlson Roberto
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Faculdade de Odontologia de Bauru da Universidade de São Paulo 2007
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4327248/
https://www.ncbi.nlm.nih.gov/pubmed/19089120
http://dx.doi.org/10.1590/S1678-77572007000200015
Descripción
Sumario:One of the causes of implant failures in cemented implant-retained prostheses is the fracture of abutment screw or UCLA abutment. This article reports a case of simultaneous fracture of two UCLA abutments screws occurring in an implant-supported prosthesis placed in the mandibular molar region. The fractured structures were examined under scanning electron microscopy to investigate the probable causes of the failure, which were not related to failures on materials or fabrication of the screws, but rather were due to shear forces. The misfit in cemented prostheses may be the most likely cause of shear force generation.