FLASH proton density imaging for improved surface coil intensity correction in quantitative and semi-quantitative SSFP perfusion cardiovascular magnetic resonance

BACKGROUND: A low excitation flip angle (α < 10°) steady-state free precession (SSFP) proton-density (PD) reference scan is often used to estimate the B(1)-field inhomogeneity for surface coil intensity correction (SCIC) of the saturation-recovery (SR) prepared high flip angle (α = 40-50°) SSFP m...

Descripción completa

Detalles Bibliográficos
Autores principales: Nielles-Vallespin, Sonia, Kellman, Peter, Hsu, Li-Yueh, Arai, Andrew E
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4331176/
https://www.ncbi.nlm.nih.gov/pubmed/25827180
http://dx.doi.org/10.1186/s12968-015-0120-6
_version_ 1782357677010059264
author Nielles-Vallespin, Sonia
Kellman, Peter
Hsu, Li-Yueh
Arai, Andrew E
author_facet Nielles-Vallespin, Sonia
Kellman, Peter
Hsu, Li-Yueh
Arai, Andrew E
author_sort Nielles-Vallespin, Sonia
collection PubMed
description BACKGROUND: A low excitation flip angle (α < 10°) steady-state free precession (SSFP) proton-density (PD) reference scan is often used to estimate the B(1)-field inhomogeneity for surface coil intensity correction (SCIC) of the saturation-recovery (SR) prepared high flip angle (α = 40-50°) SSFP myocardial perfusion images. The different SSFP off-resonance response for these two flip angles might lead to suboptimal SCIC when there is a spatial variation in the background B(0)-field. The low flip angle SSFP-PD frames are more prone to parallel imaging banding artifacts in the presence of off-resonance. The use of FLASH-PD frames would eliminate both the banding artifacts and the uneven frequency response in the presence of off-resonance in the surface coil inhomogeneity estimate and improve homogeneity of semi-quantitative and quantitative perfusion measurements. METHODS: B(0)-field maps, SSFP and FLASH-PD frames were acquired in 10 healthy volunteers to analyze the SSFP off-resonance response. Furthermore, perfusion scans preceded by both FLASH and SSFP-PD frames from 10 patients with no myocardial infarction were analyzed semi-quantitatively and quantitatively (rest n = 10 and stress n = 1). Intra-subject myocardial blood flow (MBF) coefficient of variation (CoV) over the whole left ventricle (LV), as well as intra-subject peak contrast (CE) and upslope (SLP) standard deviation (SD) over 6 LV sectors were investigated. RESULTS: In the 6 out of 10 cases where artifacts were apparent in the LV ROI of the SSFP-PD images, all three variability metrics were statistically significantly lower when using the FLASH-PD frames as input for the SCIC (CoV(MBF-FLASH) = 0.3 ± 0.1, CoV(MBF-SSFP) = 0.4 ± 0.1, p = 0.03; SD(CE-FLASH) = 10 ± 2, SD(CE-SSFP) = 32 ± 7, p = 0.01; SD(SLP-FLASH) = 0.02 ± 0.01, SD(SLP-SSFP) = 0.06 ± 0.02, p = 0.03). Example rest and stress data sets from the patient pool demonstrate that the low flip angle SSFP protocol can exhibit severe ghosting artifacts originating from off-resonance banding artifacts at the edges of the field of view that parallel imaging is not able to unfold. These artifacts lead to errors in the quantitative perfusion maps and the semi-quantitative perfusion indexes, such as false positives. It is shown that this can be avoided by using FLASH-PD frames as input for the SCIC. CONCLUSIONS: FLASH-PD images are recommended as input for SCIC of SSFP perfusion images instead of low flip angle SSFP-PD images.
format Online
Article
Text
id pubmed-4331176
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-43311762015-02-18 FLASH proton density imaging for improved surface coil intensity correction in quantitative and semi-quantitative SSFP perfusion cardiovascular magnetic resonance Nielles-Vallespin, Sonia Kellman, Peter Hsu, Li-Yueh Arai, Andrew E J Cardiovasc Magn Reson Research BACKGROUND: A low excitation flip angle (α < 10°) steady-state free precession (SSFP) proton-density (PD) reference scan is often used to estimate the B(1)-field inhomogeneity for surface coil intensity correction (SCIC) of the saturation-recovery (SR) prepared high flip angle (α = 40-50°) SSFP myocardial perfusion images. The different SSFP off-resonance response for these two flip angles might lead to suboptimal SCIC when there is a spatial variation in the background B(0)-field. The low flip angle SSFP-PD frames are more prone to parallel imaging banding artifacts in the presence of off-resonance. The use of FLASH-PD frames would eliminate both the banding artifacts and the uneven frequency response in the presence of off-resonance in the surface coil inhomogeneity estimate and improve homogeneity of semi-quantitative and quantitative perfusion measurements. METHODS: B(0)-field maps, SSFP and FLASH-PD frames were acquired in 10 healthy volunteers to analyze the SSFP off-resonance response. Furthermore, perfusion scans preceded by both FLASH and SSFP-PD frames from 10 patients with no myocardial infarction were analyzed semi-quantitatively and quantitatively (rest n = 10 and stress n = 1). Intra-subject myocardial blood flow (MBF) coefficient of variation (CoV) over the whole left ventricle (LV), as well as intra-subject peak contrast (CE) and upslope (SLP) standard deviation (SD) over 6 LV sectors were investigated. RESULTS: In the 6 out of 10 cases where artifacts were apparent in the LV ROI of the SSFP-PD images, all three variability metrics were statistically significantly lower when using the FLASH-PD frames as input for the SCIC (CoV(MBF-FLASH) = 0.3 ± 0.1, CoV(MBF-SSFP) = 0.4 ± 0.1, p = 0.03; SD(CE-FLASH) = 10 ± 2, SD(CE-SSFP) = 32 ± 7, p = 0.01; SD(SLP-FLASH) = 0.02 ± 0.01, SD(SLP-SSFP) = 0.06 ± 0.02, p = 0.03). Example rest and stress data sets from the patient pool demonstrate that the low flip angle SSFP protocol can exhibit severe ghosting artifacts originating from off-resonance banding artifacts at the edges of the field of view that parallel imaging is not able to unfold. These artifacts lead to errors in the quantitative perfusion maps and the semi-quantitative perfusion indexes, such as false positives. It is shown that this can be avoided by using FLASH-PD frames as input for the SCIC. CONCLUSIONS: FLASH-PD images are recommended as input for SCIC of SSFP perfusion images instead of low flip angle SSFP-PD images. BioMed Central 2015-02-17 /pmc/articles/PMC4331176/ /pubmed/25827180 http://dx.doi.org/10.1186/s12968-015-0120-6 Text en © Nielles-Vallespin et al.; licensee BioMed Central. 2015 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research
Nielles-Vallespin, Sonia
Kellman, Peter
Hsu, Li-Yueh
Arai, Andrew E
FLASH proton density imaging for improved surface coil intensity correction in quantitative and semi-quantitative SSFP perfusion cardiovascular magnetic resonance
title FLASH proton density imaging for improved surface coil intensity correction in quantitative and semi-quantitative SSFP perfusion cardiovascular magnetic resonance
title_full FLASH proton density imaging for improved surface coil intensity correction in quantitative and semi-quantitative SSFP perfusion cardiovascular magnetic resonance
title_fullStr FLASH proton density imaging for improved surface coil intensity correction in quantitative and semi-quantitative SSFP perfusion cardiovascular magnetic resonance
title_full_unstemmed FLASH proton density imaging for improved surface coil intensity correction in quantitative and semi-quantitative SSFP perfusion cardiovascular magnetic resonance
title_short FLASH proton density imaging for improved surface coil intensity correction in quantitative and semi-quantitative SSFP perfusion cardiovascular magnetic resonance
title_sort flash proton density imaging for improved surface coil intensity correction in quantitative and semi-quantitative ssfp perfusion cardiovascular magnetic resonance
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4331176/
https://www.ncbi.nlm.nih.gov/pubmed/25827180
http://dx.doi.org/10.1186/s12968-015-0120-6
work_keys_str_mv AT niellesvallespinsonia flashprotondensityimagingforimprovedsurfacecoilintensitycorrectioninquantitativeandsemiquantitativessfpperfusioncardiovascularmagneticresonance
AT kellmanpeter flashprotondensityimagingforimprovedsurfacecoilintensitycorrectioninquantitativeandsemiquantitativessfpperfusioncardiovascularmagneticresonance
AT hsuliyueh flashprotondensityimagingforimprovedsurfacecoilintensitycorrectioninquantitativeandsemiquantitativessfpperfusioncardiovascularmagneticresonance
AT araiandrewe flashprotondensityimagingforimprovedsurfacecoilintensitycorrectioninquantitativeandsemiquantitativessfpperfusioncardiovascularmagneticresonance