Cargando…

Microreactor Array Device

We report a device to fill an array of small chemical reaction chambers (microreactors) with reagent and then seal them using pressurized viscous liquid acting through a flexible membrane. The device enables multiple, independent chemical reactions involving free floating intermediate molecules with...

Descripción completa

Detalles Bibliográficos
Autores principales: Wiktor, Peter, Brunner, Al, Kahn, Peter, Qiu, Ji, Magee, Mitch, Bian, Xiaofang, Karthikeyan, Kailash, LaBaer, Joshua
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Nature Publishing Group 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4348619/
https://www.ncbi.nlm.nih.gov/pubmed/25736721
http://dx.doi.org/10.1038/srep08736
Descripción
Sumario:We report a device to fill an array of small chemical reaction chambers (microreactors) with reagent and then seal them using pressurized viscous liquid acting through a flexible membrane. The device enables multiple, independent chemical reactions involving free floating intermediate molecules without interference from neighboring reactions or external environments. The device is validated by protein expressed in situ directly from DNA in a microarray of ~10,000 spots with no diffusion during three hours incubation. Using the device to probe for an autoantibody cancer biomarker in blood serum sample gave five times higher signal to background ratio compared to standard protein microarray expressed on a flat microscope slide. Physical design principles to effectively fill the array of microreactors with reagent and experimental results of alternate methods for sealing the microreactors are presented.