Noninvasive Method for Measuring Local Pulse Wave Velocity by Dual Pulse Wave Doppler: In Vitro and In Vivo Studies

OBJECTIVES: To evaluate the validity and reproducibility of a noninvasive dual pulse wave Doppler (DPWD) method, which involves simultaneous recording of flow velocity of two independent sample volumes with a measurable distance, for measuring the local arterial pulse wave velocity (PWV) through in...

Descripción completa

Detalles Bibliográficos
Autores principales: Wang, Zhen, Yang, Yong, Yuan, Li-jun, Liu, Jie, Duan, Yun-you, Cao, Tie-sheng
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4364771/
https://www.ncbi.nlm.nih.gov/pubmed/25786124
http://dx.doi.org/10.1371/journal.pone.0120482
_version_ 1782362125578010624
author Wang, Zhen
Yang, Yong
Yuan, Li-jun
Liu, Jie
Duan, Yun-you
Cao, Tie-sheng
author_facet Wang, Zhen
Yang, Yong
Yuan, Li-jun
Liu, Jie
Duan, Yun-you
Cao, Tie-sheng
author_sort Wang, Zhen
collection PubMed
description OBJECTIVES: To evaluate the validity and reproducibility of a noninvasive dual pulse wave Doppler (DPWD) method, which involves simultaneous recording of flow velocity of two independent sample volumes with a measurable distance, for measuring the local arterial pulse wave velocity (PWV) through in vitro and in vivo studies. METHODS: The DPWD mode of Hitachi HI Vision Preirus ultrasound system with a 5–13MHz transducer was used. An in vitro model was designed to compare the PWV of a homogeneous rubber tubing with the local PWV of its middle part measured by DPWD method. In the in vivo study, local PWV of 45 hypertensive patients (25 male, 49.8±3.1 years) and 45 matched healthy subjects (25 male, 49.3±3.0 years) were investigated at the left common carotid artery (LCCA) by DPWD method. RESULTS: In the in vitro study, the local PWV measured by DPWP method and the PWV of the homogeneous rubber tubing did not show statistical difference (5.16 ± 0.28 m/s vs 5.03 ± 0.15 m/s, p = 0.075). The coefficient of variation (CV) of the intra- and inter- measurements for local PWV were 3.46% and 4.96%, for the PWV of the homogeneous rubber tubing were 0.99% and 1.98%. In the in vivo study, a significantly higher local PWV of LCCA was found in the hypertensive patients as compared to that in healthy subjects (6.29±1.04m/s vs. 5.31±0.72m/s, P = 0.019). The CV of the intra- and inter- measurements in hypertensive patients were 2.22% and 3.94%, in healthy subjects were 2.07% and 4.14%. CONCLUSIONS: This study demonstrated the feasibility of the noninvasive DPWD method to determine the local PWV, which was accurate and reproducible not only in vitro but also in vivo studies. This noninvasive echocardiographic method may be illuminating to clinical use.
format Online
Article
Text
id pubmed-4364771
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-43647712015-03-23 Noninvasive Method for Measuring Local Pulse Wave Velocity by Dual Pulse Wave Doppler: In Vitro and In Vivo Studies Wang, Zhen Yang, Yong Yuan, Li-jun Liu, Jie Duan, Yun-you Cao, Tie-sheng PLoS One Research Article OBJECTIVES: To evaluate the validity and reproducibility of a noninvasive dual pulse wave Doppler (DPWD) method, which involves simultaneous recording of flow velocity of two independent sample volumes with a measurable distance, for measuring the local arterial pulse wave velocity (PWV) through in vitro and in vivo studies. METHODS: The DPWD mode of Hitachi HI Vision Preirus ultrasound system with a 5–13MHz transducer was used. An in vitro model was designed to compare the PWV of a homogeneous rubber tubing with the local PWV of its middle part measured by DPWD method. In the in vivo study, local PWV of 45 hypertensive patients (25 male, 49.8±3.1 years) and 45 matched healthy subjects (25 male, 49.3±3.0 years) were investigated at the left common carotid artery (LCCA) by DPWD method. RESULTS: In the in vitro study, the local PWV measured by DPWP method and the PWV of the homogeneous rubber tubing did not show statistical difference (5.16 ± 0.28 m/s vs 5.03 ± 0.15 m/s, p = 0.075). The coefficient of variation (CV) of the intra- and inter- measurements for local PWV were 3.46% and 4.96%, for the PWV of the homogeneous rubber tubing were 0.99% and 1.98%. In the in vivo study, a significantly higher local PWV of LCCA was found in the hypertensive patients as compared to that in healthy subjects (6.29±1.04m/s vs. 5.31±0.72m/s, P = 0.019). The CV of the intra- and inter- measurements in hypertensive patients were 2.22% and 3.94%, in healthy subjects were 2.07% and 4.14%. CONCLUSIONS: This study demonstrated the feasibility of the noninvasive DPWD method to determine the local PWV, which was accurate and reproducible not only in vitro but also in vivo studies. This noninvasive echocardiographic method may be illuminating to clinical use. Public Library of Science 2015-03-18 /pmc/articles/PMC4364771/ /pubmed/25786124 http://dx.doi.org/10.1371/journal.pone.0120482 Text en © 2015 Wang et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited.
spellingShingle Research Article
Wang, Zhen
Yang, Yong
Yuan, Li-jun
Liu, Jie
Duan, Yun-you
Cao, Tie-sheng
Noninvasive Method for Measuring Local Pulse Wave Velocity by Dual Pulse Wave Doppler: In Vitro and In Vivo Studies
title Noninvasive Method for Measuring Local Pulse Wave Velocity by Dual Pulse Wave Doppler: In Vitro and In Vivo Studies
title_full Noninvasive Method for Measuring Local Pulse Wave Velocity by Dual Pulse Wave Doppler: In Vitro and In Vivo Studies
title_fullStr Noninvasive Method for Measuring Local Pulse Wave Velocity by Dual Pulse Wave Doppler: In Vitro and In Vivo Studies
title_full_unstemmed Noninvasive Method for Measuring Local Pulse Wave Velocity by Dual Pulse Wave Doppler: In Vitro and In Vivo Studies
title_short Noninvasive Method for Measuring Local Pulse Wave Velocity by Dual Pulse Wave Doppler: In Vitro and In Vivo Studies
title_sort noninvasive method for measuring local pulse wave velocity by dual pulse wave doppler: in vitro and in vivo studies
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4364771/
https://www.ncbi.nlm.nih.gov/pubmed/25786124
http://dx.doi.org/10.1371/journal.pone.0120482
work_keys_str_mv AT wangzhen noninvasivemethodformeasuringlocalpulsewavevelocitybydualpulsewavedopplerinvitroandinvivostudies
AT yangyong noninvasivemethodformeasuringlocalpulsewavevelocitybydualpulsewavedopplerinvitroandinvivostudies
AT yuanlijun noninvasivemethodformeasuringlocalpulsewavevelocitybydualpulsewavedopplerinvitroandinvivostudies
AT liujie noninvasivemethodformeasuringlocalpulsewavevelocitybydualpulsewavedopplerinvitroandinvivostudies
AT duanyunyou noninvasivemethodformeasuringlocalpulsewavevelocitybydualpulsewavedopplerinvitroandinvivostudies
AT caotiesheng noninvasivemethodformeasuringlocalpulsewavevelocitybydualpulsewavedopplerinvitroandinvivostudies