Mechanical and Biomechanical Characterization of a Polyurethane Nucleus Replacement Device Injected and Cured In Situ Within a Balloon

BACKGROUND: The DASCOR device has recently been introduced as an innovative nucleus replacement alternative for the treatment of low-back pain caused by degenerative intervertebral disc disease. The purpose of this study was to characterize, through a series of preclinical mechanical bench and biome...

Descripción completa

Detalles Bibliográficos
Autores principales: Tsantrizos, Anthony, Ordway, Nathaniel R., Myint, Khin, Martz, Erik, Yuan, Hansen A.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: RRY Publications, LLC 2008
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4365654/
https://www.ncbi.nlm.nih.gov/pubmed/25802599
http://dx.doi.org/10.1016/SASJ-2007-0113-RR
_version_ 1782362260811808768
author Tsantrizos, Anthony
Ordway, Nathaniel R.
Myint, Khin
Martz, Erik
Yuan, Hansen A.
author_facet Tsantrizos, Anthony
Ordway, Nathaniel R.
Myint, Khin
Martz, Erik
Yuan, Hansen A.
author_sort Tsantrizos, Anthony
collection PubMed
description BACKGROUND: The DASCOR device has recently been introduced as an innovative nucleus replacement alternative for the treatment of low-back pain caused by degenerative intervertebral disc disease. The purpose of this study was to characterize, through a series of preclinical mechanical bench and biomechanical tests, the effectiveness of this device. METHODS: A number of samples were created using similar preparation methods in order to characterize the nucleus replacement device in multiple mechanical bench tests, using ASTM-guided protocols, where appropriate. Mechanical bench testing included static testing to characterize the device's compressive, shear properties, and fatigue testing to determine the device's compressive fatigue strength, wear, and durability. Biomechanical testing, using human cadaveric lumbar spines, was also conducted to determine the ability of the device to restore multidirectional segmental flexibility and to determine its resulting endplate contact stress. RESULTS: The static compressive and shear moduli of the nucleus replacement device were determined to be between 4.2–5.6 MPa and 1.4–1.9 MPa, respectively. Similarly, the ultimate compressive and shear strength were 12,400 N and 6,993 N, respectively. The maximum axial compressive fatigue strength of the tested device that was able to withstand a runout without failure was determined to be approximately 3 MPa. The wear assessment determined that the device is durable and yielded minimal wear rates of 0.29mg/Mc. Finally, the biomechanical testing demonstrated that the device can restore the multidirectional segmental flexibility to a level seen in the intact condition while concurrently producing a uniform endplate contact stress. CONCLUSIONS: The results of the present study provided a mechanical justification supporting the clinical use of the nucleus replacement device and also help explain and support the positive clinical results obtained from two European studies and one US pilot study. CLINICAL RELEVANCE: Nucleus replacement devices are rapidly emerging to address specific conditions of degenerative disc disease. Preclinical testing of such devices is paramount in order to potentially ensure successful clinical outcomes post implantation
format Online
Article
Text
id pubmed-4365654
institution National Center for Biotechnology Information
language English
publishDate 2008
publisher RRY Publications, LLC
record_format MEDLINE/PubMed
spelling pubmed-43656542015-03-23 Mechanical and Biomechanical Characterization of a Polyurethane Nucleus Replacement Device Injected and Cured In Situ Within a Balloon Tsantrizos, Anthony Ordway, Nathaniel R. Myint, Khin Martz, Erik Yuan, Hansen A. SAS J Nucleus Replacement BACKGROUND: The DASCOR device has recently been introduced as an innovative nucleus replacement alternative for the treatment of low-back pain caused by degenerative intervertebral disc disease. The purpose of this study was to characterize, through a series of preclinical mechanical bench and biomechanical tests, the effectiveness of this device. METHODS: A number of samples were created using similar preparation methods in order to characterize the nucleus replacement device in multiple mechanical bench tests, using ASTM-guided protocols, where appropriate. Mechanical bench testing included static testing to characterize the device's compressive, shear properties, and fatigue testing to determine the device's compressive fatigue strength, wear, and durability. Biomechanical testing, using human cadaveric lumbar spines, was also conducted to determine the ability of the device to restore multidirectional segmental flexibility and to determine its resulting endplate contact stress. RESULTS: The static compressive and shear moduli of the nucleus replacement device were determined to be between 4.2–5.6 MPa and 1.4–1.9 MPa, respectively. Similarly, the ultimate compressive and shear strength were 12,400 N and 6,993 N, respectively. The maximum axial compressive fatigue strength of the tested device that was able to withstand a runout without failure was determined to be approximately 3 MPa. The wear assessment determined that the device is durable and yielded minimal wear rates of 0.29mg/Mc. Finally, the biomechanical testing demonstrated that the device can restore the multidirectional segmental flexibility to a level seen in the intact condition while concurrently producing a uniform endplate contact stress. CONCLUSIONS: The results of the present study provided a mechanical justification supporting the clinical use of the nucleus replacement device and also help explain and support the positive clinical results obtained from two European studies and one US pilot study. CLINICAL RELEVANCE: Nucleus replacement devices are rapidly emerging to address specific conditions of degenerative disc disease. Preclinical testing of such devices is paramount in order to potentially ensure successful clinical outcomes post implantation RRY Publications, LLC 2008-03-01 /pmc/articles/PMC4365654/ /pubmed/25802599 http://dx.doi.org/10.1016/SASJ-2007-0113-RR Text en Copyright SAS - Spine Arthroplasty Society 2008 http://creativecommons.org/licenses/by-nc-nd/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution-Noncommercial 3.0 Unported License, permitting all non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Nucleus Replacement
Tsantrizos, Anthony
Ordway, Nathaniel R.
Myint, Khin
Martz, Erik
Yuan, Hansen A.
Mechanical and Biomechanical Characterization of a Polyurethane Nucleus Replacement Device Injected and Cured In Situ Within a Balloon
title Mechanical and Biomechanical Characterization of a Polyurethane Nucleus Replacement Device Injected and Cured In Situ Within a Balloon
title_full Mechanical and Biomechanical Characterization of a Polyurethane Nucleus Replacement Device Injected and Cured In Situ Within a Balloon
title_fullStr Mechanical and Biomechanical Characterization of a Polyurethane Nucleus Replacement Device Injected and Cured In Situ Within a Balloon
title_full_unstemmed Mechanical and Biomechanical Characterization of a Polyurethane Nucleus Replacement Device Injected and Cured In Situ Within a Balloon
title_short Mechanical and Biomechanical Characterization of a Polyurethane Nucleus Replacement Device Injected and Cured In Situ Within a Balloon
title_sort mechanical and biomechanical characterization of a polyurethane nucleus replacement device injected and cured in situ within a balloon
topic Nucleus Replacement
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4365654/
https://www.ncbi.nlm.nih.gov/pubmed/25802599
http://dx.doi.org/10.1016/SASJ-2007-0113-RR
work_keys_str_mv AT tsantrizosanthony mechanicalandbiomechanicalcharacterizationofapolyurethanenucleusreplacementdeviceinjectedandcuredinsituwithinaballoon
AT ordwaynathanielr mechanicalandbiomechanicalcharacterizationofapolyurethanenucleusreplacementdeviceinjectedandcuredinsituwithinaballoon
AT myintkhin mechanicalandbiomechanicalcharacterizationofapolyurethanenucleusreplacementdeviceinjectedandcuredinsituwithinaballoon
AT martzerik mechanicalandbiomechanicalcharacterizationofapolyurethanenucleusreplacementdeviceinjectedandcuredinsituwithinaballoon
AT yuanhansena mechanicalandbiomechanicalcharacterizationofapolyurethanenucleusreplacementdeviceinjectedandcuredinsituwithinaballoon