Comparison of standardized uptake values measured on F-NaF PET/CT scans using three different tube current intensities
OBJECTIVE: To analyze standardized uptake values (SUVs) using three different tube current intensities for attenuation correction on (18)FNaF PET/CT scans. MATERIALS AND METHODS: A total of 254 (18)F-NaF PET/CT studies were analyzed using 10, 20 and 30 mAs. The SUVs were calculated in volumes of int...
Autores principales: | , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Colégio Brasileiro de Radiologia e Diagnóstico por
Imagem
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4366024/ https://www.ncbi.nlm.nih.gov/pubmed/25798003 http://dx.doi.org/10.1590/0100-3984.2014.0034 |
Sumario: | OBJECTIVE: To analyze standardized uptake values (SUVs) using three different tube current intensities for attenuation correction on (18)FNaF PET/CT scans. MATERIALS AND METHODS: A total of 254 (18)F-NaF PET/CT studies were analyzed using 10, 20 and 30 mAs. The SUVs were calculated in volumes of interest (VOIs) drawn on three skeletal regions, namely, right proximal humeral diaphysis (RH), right proximal femoral diaphysis (RF), and first lumbar vertebra (LV1) in a total of 712 VOIs. The analyses covered 675 regions classified as normal (236 RH, 232 RF, and 207 LV1). RESULTS: Mean SUV for each skeletal region was 3.8, 5.4 and 14.4 for RH, RF, and LV1, respectively. As the studies were grouped according to mAs value, the mean SUV values were 3.8, 3.9 and 3.7 for 10, 20 and 30 mAs, respectively, in the RH region; 5.4, 5.5 and 5.4 for 10, 20 and 30 mAs, respectively, in the RF region; 13.8, 14.9 and 14.5 for 10, 20 and 30 mAs, respectively, in the LV1 region. CONCLUSION: The three tube current values yielded similar results for SUV calculation. |
---|