Capacitance Variation Induced by Microfluidic Two-Phase Flow across Insulated Interdigital Electrodes in Lab-On-Chip Devices

Microfluidic two-phase flow detection has attracted plenty of interest in various areas of biology, medicine and chemistry. This work presents a capacitive sensor using insulated interdigital electrodes (IDEs) to detect the presence of droplets in a microchannel. This droplet sensor is composed of a...

Descripción completa

Detalles Bibliográficos
Autores principales: Dong, Tao, Barbosa, Cátia
Formato: Online Artículo Texto
Lenguaje:English
Publicado: MDPI 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4367328/
https://www.ncbi.nlm.nih.gov/pubmed/25629705
http://dx.doi.org/10.3390/s150202694
_version_ 1782362515493093376
author Dong, Tao
Barbosa, Cátia
author_facet Dong, Tao
Barbosa, Cátia
author_sort Dong, Tao
collection PubMed
description Microfluidic two-phase flow detection has attracted plenty of interest in various areas of biology, medicine and chemistry. This work presents a capacitive sensor using insulated interdigital electrodes (IDEs) to detect the presence of droplets in a microchannel. This droplet sensor is composed of a glass substrate, patterned gold electrodes and an insulation layer. A polydimethylsiloxane (PDMS) cover bonded to the multilayered structure forms a microchannel. Capacitance variation induced by the droplet passage was thoroughly investigated with both simulation and experimental work. Olive oil and deionized water were employed as the working fluids in the experiments to demonstrate the droplet sensor. The results show a good sensitivity of the droplet with the appropriate measurement connection. This capacitive droplet sensor is promising to be integrated into a lab-on-chip device for in situ monitoring/counting of droplets or bubbles.
format Online
Article
Text
id pubmed-4367328
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher MDPI
record_format MEDLINE/PubMed
spelling pubmed-43673282015-04-30 Capacitance Variation Induced by Microfluidic Two-Phase Flow across Insulated Interdigital Electrodes in Lab-On-Chip Devices Dong, Tao Barbosa, Cátia Sensors (Basel) Article Microfluidic two-phase flow detection has attracted plenty of interest in various areas of biology, medicine and chemistry. This work presents a capacitive sensor using insulated interdigital electrodes (IDEs) to detect the presence of droplets in a microchannel. This droplet sensor is composed of a glass substrate, patterned gold electrodes and an insulation layer. A polydimethylsiloxane (PDMS) cover bonded to the multilayered structure forms a microchannel. Capacitance variation induced by the droplet passage was thoroughly investigated with both simulation and experimental work. Olive oil and deionized water were employed as the working fluids in the experiments to demonstrate the droplet sensor. The results show a good sensitivity of the droplet with the appropriate measurement connection. This capacitive droplet sensor is promising to be integrated into a lab-on-chip device for in situ monitoring/counting of droplets or bubbles. MDPI 2015-01-26 /pmc/articles/PMC4367328/ /pubmed/25629705 http://dx.doi.org/10.3390/s150202694 Text en © 2015 by the authors; licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution license (http://creativecommons.org/licenses/by/3.0/)
spellingShingle Article
Dong, Tao
Barbosa, Cátia
Capacitance Variation Induced by Microfluidic Two-Phase Flow across Insulated Interdigital Electrodes in Lab-On-Chip Devices
title Capacitance Variation Induced by Microfluidic Two-Phase Flow across Insulated Interdigital Electrodes in Lab-On-Chip Devices
title_full Capacitance Variation Induced by Microfluidic Two-Phase Flow across Insulated Interdigital Electrodes in Lab-On-Chip Devices
title_fullStr Capacitance Variation Induced by Microfluidic Two-Phase Flow across Insulated Interdigital Electrodes in Lab-On-Chip Devices
title_full_unstemmed Capacitance Variation Induced by Microfluidic Two-Phase Flow across Insulated Interdigital Electrodes in Lab-On-Chip Devices
title_short Capacitance Variation Induced by Microfluidic Two-Phase Flow across Insulated Interdigital Electrodes in Lab-On-Chip Devices
title_sort capacitance variation induced by microfluidic two-phase flow across insulated interdigital electrodes in lab-on-chip devices
topic Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4367328/
https://www.ncbi.nlm.nih.gov/pubmed/25629705
http://dx.doi.org/10.3390/s150202694
work_keys_str_mv AT dongtao capacitancevariationinducedbymicrofluidictwophaseflowacrossinsulatedinterdigitalelectrodesinlabonchipdevices
AT barbosacatia capacitancevariationinducedbymicrofluidictwophaseflowacrossinsulatedinterdigitalelectrodesinlabonchipdevices