Evaluation of IVOCT imaging of coronary artery metallic stents with neointimal coverage

Accuracy of IVOCT for measurement of neointimal thickness and effect of neointima in the appearance of metallic struts in IVOCT images was investigated. Phantom vessels were constructed and coronary stents were deployed and covered with thick (250–400 μm) and thin (30–70 μm) phantom neointima. High...

Descripción completa

Detalles Bibliográficos
Autores principales: Elahi, Sahar, Ho, Derek, Feldman, Marc D., Dijkstra, Jouke, Milner, Thomas E.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Netherlands 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4368839/
https://www.ncbi.nlm.nih.gov/pubmed/25395364
http://dx.doi.org/10.1007/s10554-014-0569-7
_version_ 1782362699375575040
author Elahi, Sahar
Ho, Derek
Feldman, Marc D.
Dijkstra, Jouke
Milner, Thomas E.
author_facet Elahi, Sahar
Ho, Derek
Feldman, Marc D.
Dijkstra, Jouke
Milner, Thomas E.
author_sort Elahi, Sahar
collection PubMed
description Accuracy of IVOCT for measurement of neointimal thickness and effect of neointima in the appearance of metallic struts in IVOCT images was investigated. Phantom vessels were constructed and coronary stents were deployed and covered with thick (250–400 μm) and thin (30–70 μm) phantom neointima. High resolution Micro-CT images of the stent struts were recorded as a gold standard. IVOCT images of the phantom vessels were acquired with various luminal blood scattering strengths and measured neointimal thicknesses from IVOCT and Micro-CT images were compared. In transparent lumen, comparison of IVOCT and Micro-CT neointima thickness measurements found no significant difference (p > 0.05) in the thick neointima phantom but a significant difference (p < 0.05) in the thin neointima phantom. For both thick and thin neointima, IVOCT neointimal thickness measurements varied from Micro-CT values by as much as ±35 %. Increased luminal scattering due to presence of blood at concentrations <5 % did not interfere with measurement of thin neointimas and was validated by ANOVA analysis (p = 0.95). IV-OCT measurement of strut feature size with an overlying thin neointima match true values determined with Micro-CT (p = 0.82). Presence of a thick neointima resulted in lateral elongation or merry-go-rounding of stent strut features in IVOCT images. Phantom IVOCT images suggest that thick neointimal layers can result in more than 40 % lateral elongation of stent strut features. Simulated IVOCT images of metallic stent struts with varying neointimal thickness suggest that neointimal light scattering can introduce the merry-go-round effect.
format Online
Article
Text
id pubmed-4368839
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Springer Netherlands
record_format MEDLINE/PubMed
spelling pubmed-43688392015-03-26 Evaluation of IVOCT imaging of coronary artery metallic stents with neointimal coverage Elahi, Sahar Ho, Derek Feldman, Marc D. Dijkstra, Jouke Milner, Thomas E. Int J Cardiovasc Imaging Original Paper Accuracy of IVOCT for measurement of neointimal thickness and effect of neointima in the appearance of metallic struts in IVOCT images was investigated. Phantom vessels were constructed and coronary stents were deployed and covered with thick (250–400 μm) and thin (30–70 μm) phantom neointima. High resolution Micro-CT images of the stent struts were recorded as a gold standard. IVOCT images of the phantom vessels were acquired with various luminal blood scattering strengths and measured neointimal thicknesses from IVOCT and Micro-CT images were compared. In transparent lumen, comparison of IVOCT and Micro-CT neointima thickness measurements found no significant difference (p > 0.05) in the thick neointima phantom but a significant difference (p < 0.05) in the thin neointima phantom. For both thick and thin neointima, IVOCT neointimal thickness measurements varied from Micro-CT values by as much as ±35 %. Increased luminal scattering due to presence of blood at concentrations <5 % did not interfere with measurement of thin neointimas and was validated by ANOVA analysis (p = 0.95). IV-OCT measurement of strut feature size with an overlying thin neointima match true values determined with Micro-CT (p = 0.82). Presence of a thick neointima resulted in lateral elongation or merry-go-rounding of stent strut features in IVOCT images. Phantom IVOCT images suggest that thick neointimal layers can result in more than 40 % lateral elongation of stent strut features. Simulated IVOCT images of metallic stent struts with varying neointimal thickness suggest that neointimal light scattering can introduce the merry-go-round effect. Springer Netherlands 2014-11-14 2015 /pmc/articles/PMC4368839/ /pubmed/25395364 http://dx.doi.org/10.1007/s10554-014-0569-7 Text en © The Author(s) 2014 https://creativecommons.org/licenses/by/4.0/ Open AccessThis article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.
spellingShingle Original Paper
Elahi, Sahar
Ho, Derek
Feldman, Marc D.
Dijkstra, Jouke
Milner, Thomas E.
Evaluation of IVOCT imaging of coronary artery metallic stents with neointimal coverage
title Evaluation of IVOCT imaging of coronary artery metallic stents with neointimal coverage
title_full Evaluation of IVOCT imaging of coronary artery metallic stents with neointimal coverage
title_fullStr Evaluation of IVOCT imaging of coronary artery metallic stents with neointimal coverage
title_full_unstemmed Evaluation of IVOCT imaging of coronary artery metallic stents with neointimal coverage
title_short Evaluation of IVOCT imaging of coronary artery metallic stents with neointimal coverage
title_sort evaluation of ivoct imaging of coronary artery metallic stents with neointimal coverage
topic Original Paper
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4368839/
https://www.ncbi.nlm.nih.gov/pubmed/25395364
http://dx.doi.org/10.1007/s10554-014-0569-7
work_keys_str_mv AT elahisahar evaluationofivoctimagingofcoronaryarterymetallicstentswithneointimalcoverage
AT hoderek evaluationofivoctimagingofcoronaryarterymetallicstentswithneointimalcoverage
AT feldmanmarcd evaluationofivoctimagingofcoronaryarterymetallicstentswithneointimalcoverage
AT dijkstrajouke evaluationofivoctimagingofcoronaryarterymetallicstentswithneointimalcoverage
AT milnerthomase evaluationofivoctimagingofcoronaryarterymetallicstentswithneointimalcoverage