Effects of pH, lactate, hematocrit and potassium level on the accuracy of continuous glucose monitoring (CGM) in pediatric intensive care unit
BACKGROUND: Continuous glucose monitoring (CGM) originally was developed for diabetic patients and it may be a useful tool for monitoring glucose changes in pediatric intensive care unit (PICU). Its use is, however, limited by the lack of sufficient data on its reliability at insufficient peripheral...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4372234/ https://www.ncbi.nlm.nih.gov/pubmed/25888142 http://dx.doi.org/10.1186/s13052-015-0122-x |
_version_ | 1782363145608626176 |
---|---|
author | Marics, Gábor Koncz, Levente Eitler, Katalin Vatai, Barbara Szénási, Boglárka Zakariás, David Mikos, Borbála Körner, Anna Tóth-Heyn, Péter |
author_facet | Marics, Gábor Koncz, Levente Eitler, Katalin Vatai, Barbara Szénási, Boglárka Zakariás, David Mikos, Borbála Körner, Anna Tóth-Heyn, Péter |
author_sort | Marics, Gábor |
collection | PubMed |
description | BACKGROUND: Continuous glucose monitoring (CGM) originally was developed for diabetic patients and it may be a useful tool for monitoring glucose changes in pediatric intensive care unit (PICU). Its use is, however, limited by the lack of sufficient data on its reliability at insufficient peripheral perfusion. We aimed to correlate the accuracy of CGM with laboratory markers relevant to disturbed tissue perfusion. PATIENTS AND METHODS: In 38 pediatric patients (age range, 0–18 years) requiring intensive care we tested the effect of pH, lactate, hematocrit and serum potassium on the difference between CGM and meter glucose measurements. Guardian® (Medtronic®) CGM results were compared to GEM 3000 (Instrumentation laboratory®) and point-of-care measurements. The clinical accuracy of CGM was evaluated by Clarke Error Grid -, Bland-Altman analysis and Pearson’s correlation. We used Friedman test for statistical analysis (statistical significance was established as a p < 0.05). RESULTS: CGM values exhibited a considerable variability without any correlation with the examined laboratory parameters. Clarke, Bland-Altman analysis and Pearson’s correlation coefficient demonstrated a good clinical accuracy of CGM (zone A and B = 96%; the mean difference between reference and CGM glucose was 1,3 mg/dL, 48 from the 780 calibration pairs overrunning the 2 standard deviation; Pearson’s correlation coefficient: 0.83). CONCLUSIONS: The accuracy of CGM measurements is independent of laboratory parameters relevant to tissue hypoperfusion. CGM may prove a reliable tool for continuous monitoring of glucose changes in PICUs, not much influenced by tissue perfusion, but still not appropriate for being the base for clinical decisions. |
format | Online Article Text |
id | pubmed-4372234 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | BioMed Central |
record_format | MEDLINE/PubMed |
spelling | pubmed-43722342015-03-25 Effects of pH, lactate, hematocrit and potassium level on the accuracy of continuous glucose monitoring (CGM) in pediatric intensive care unit Marics, Gábor Koncz, Levente Eitler, Katalin Vatai, Barbara Szénási, Boglárka Zakariás, David Mikos, Borbála Körner, Anna Tóth-Heyn, Péter Ital J Pediatr Research BACKGROUND: Continuous glucose monitoring (CGM) originally was developed for diabetic patients and it may be a useful tool for monitoring glucose changes in pediatric intensive care unit (PICU). Its use is, however, limited by the lack of sufficient data on its reliability at insufficient peripheral perfusion. We aimed to correlate the accuracy of CGM with laboratory markers relevant to disturbed tissue perfusion. PATIENTS AND METHODS: In 38 pediatric patients (age range, 0–18 years) requiring intensive care we tested the effect of pH, lactate, hematocrit and serum potassium on the difference between CGM and meter glucose measurements. Guardian® (Medtronic®) CGM results were compared to GEM 3000 (Instrumentation laboratory®) and point-of-care measurements. The clinical accuracy of CGM was evaluated by Clarke Error Grid -, Bland-Altman analysis and Pearson’s correlation. We used Friedman test for statistical analysis (statistical significance was established as a p < 0.05). RESULTS: CGM values exhibited a considerable variability without any correlation with the examined laboratory parameters. Clarke, Bland-Altman analysis and Pearson’s correlation coefficient demonstrated a good clinical accuracy of CGM (zone A and B = 96%; the mean difference between reference and CGM glucose was 1,3 mg/dL, 48 from the 780 calibration pairs overrunning the 2 standard deviation; Pearson’s correlation coefficient: 0.83). CONCLUSIONS: The accuracy of CGM measurements is independent of laboratory parameters relevant to tissue hypoperfusion. CGM may prove a reliable tool for continuous monitoring of glucose changes in PICUs, not much influenced by tissue perfusion, but still not appropriate for being the base for clinical decisions. BioMed Central 2015-03-19 /pmc/articles/PMC4372234/ /pubmed/25888142 http://dx.doi.org/10.1186/s13052-015-0122-x Text en © Marics et al.; licensee BioMed Central. 2015 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated. |
spellingShingle | Research Marics, Gábor Koncz, Levente Eitler, Katalin Vatai, Barbara Szénási, Boglárka Zakariás, David Mikos, Borbála Körner, Anna Tóth-Heyn, Péter Effects of pH, lactate, hematocrit and potassium level on the accuracy of continuous glucose monitoring (CGM) in pediatric intensive care unit |
title | Effects of pH, lactate, hematocrit and potassium level on the accuracy of continuous glucose monitoring (CGM) in pediatric intensive care unit |
title_full | Effects of pH, lactate, hematocrit and potassium level on the accuracy of continuous glucose monitoring (CGM) in pediatric intensive care unit |
title_fullStr | Effects of pH, lactate, hematocrit and potassium level on the accuracy of continuous glucose monitoring (CGM) in pediatric intensive care unit |
title_full_unstemmed | Effects of pH, lactate, hematocrit and potassium level on the accuracy of continuous glucose monitoring (CGM) in pediatric intensive care unit |
title_short | Effects of pH, lactate, hematocrit and potassium level on the accuracy of continuous glucose monitoring (CGM) in pediatric intensive care unit |
title_sort | effects of ph, lactate, hematocrit and potassium level on the accuracy of continuous glucose monitoring (cgm) in pediatric intensive care unit |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4372234/ https://www.ncbi.nlm.nih.gov/pubmed/25888142 http://dx.doi.org/10.1186/s13052-015-0122-x |
work_keys_str_mv | AT maricsgabor effectsofphlactatehematocritandpotassiumlevelontheaccuracyofcontinuousglucosemonitoringcgminpediatricintensivecareunit AT konczlevente effectsofphlactatehematocritandpotassiumlevelontheaccuracyofcontinuousglucosemonitoringcgminpediatricintensivecareunit AT eitlerkatalin effectsofphlactatehematocritandpotassiumlevelontheaccuracyofcontinuousglucosemonitoringcgminpediatricintensivecareunit AT vataibarbara effectsofphlactatehematocritandpotassiumlevelontheaccuracyofcontinuousglucosemonitoringcgminpediatricintensivecareunit AT szenasiboglarka effectsofphlactatehematocritandpotassiumlevelontheaccuracyofcontinuousglucosemonitoringcgminpediatricintensivecareunit AT zakariasdavid effectsofphlactatehematocritandpotassiumlevelontheaccuracyofcontinuousglucosemonitoringcgminpediatricintensivecareunit AT mikosborbala effectsofphlactatehematocritandpotassiumlevelontheaccuracyofcontinuousglucosemonitoringcgminpediatricintensivecareunit AT korneranna effectsofphlactatehematocritandpotassiumlevelontheaccuracyofcontinuousglucosemonitoringcgminpediatricintensivecareunit AT tothheynpeter effectsofphlactatehematocritandpotassiumlevelontheaccuracyofcontinuousglucosemonitoringcgminpediatricintensivecareunit |