Cargando…

Design and Evaluation of a Cochlear Implant Strategy Based on a “Phantom” Channel

Unbalanced bipolar stimulation, delivered using charge balanced pulses, was used to produce “Phantom stimulation”, stimulation beyond the most apical contact of a cochlear implant’s electrode array. The Phantom channel was allocated audio frequencies below 300Hz in a speech coding strategy, conveyin...

Descripción completa

Detalles Bibliográficos
Autores principales: Nogueira, Waldo, Litvak, Leonid M., Saoji, Aniket A., Büchner, Andreas
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4373925/
https://www.ncbi.nlm.nih.gov/pubmed/25806818
http://dx.doi.org/10.1371/journal.pone.0120148
Descripción
Sumario:Unbalanced bipolar stimulation, delivered using charge balanced pulses, was used to produce “Phantom stimulation”, stimulation beyond the most apical contact of a cochlear implant’s electrode array. The Phantom channel was allocated audio frequencies below 300Hz in a speech coding strategy, conveying energy some two octaves lower than the clinical strategy and hence delivering the fundamental frequency of speech and of many musical tones. A group of 12 Advanced Bionics cochlear implant recipients took part in a chronic study investigating the fitting of the Phantom strategy and speech and music perception when using Phantom. The evaluation of speech in noise was performed immediately after fitting Phantom for the first time (Session 1) and after one month of take-home experience (Session 2). A repeated measures of analysis of variance (ANOVA) within factors strategy (Clinical, Phantom) and interaction time (Session 1, Session 2) revealed a significant effect for the interaction time and strategy. Phantom obtained a significant improvement in speech intelligibility after one month of use. Furthermore, a trend towards a better performance with Phantom (48%) with respect to F120 (37%) after 1 month of use failed to reach significance after type 1 error correction. Questionnaire results show a preference for Phantom when listening to music, likely driven by an improved balance between high and low frequencies.