Deviation in the predefined calibration factors of the dose calibrators and the associated inaccuracy in the radioactivity measurements of beta-gamma emitters

AIM: To determine whether the predefined calibration factors of the dose calibrators can provide accurate radioactivity measurements of beta-gamma emitters used in routine therapeutic nuclear medicine procedures. MATERIALS AND METHODS: Two models of dose calibrators were used in the present study fo...

Descripción completa

Detalles Bibliográficos
Autores principales: Sharma, Sarika, Singh, Baljinder, Koul, Ashwani, Mittal, Bhagwant Rai
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4379670/
https://www.ncbi.nlm.nih.gov/pubmed/25829729
http://dx.doi.org/10.4103/0972-3919.152972
_version_ 1782364229225938944
author Sharma, Sarika
Singh, Baljinder
Koul, Ashwani
Mittal, Bhagwant Rai
author_facet Sharma, Sarika
Singh, Baljinder
Koul, Ashwani
Mittal, Bhagwant Rai
author_sort Sharma, Sarika
collection PubMed
description AIM: To determine whether the predefined calibration factors of the dose calibrators can provide accurate radioactivity measurements of beta-gamma emitters used in routine therapeutic nuclear medicine procedures. MATERIALS AND METHODS: Two models of dose calibrators were used in the present study for radioactivity measurements of (153)Sm ethylenediamine-N, N, N’, N’-tetrakis methylene phosphonic acid (EDTMP) and (177)Lu (EDTMP). A known (precalibrated) activity of each of the two beta emitters received by us from our National Supplier for administration to the patients with extensive bony metastases for bone pain palliation, was used for experiments. RESULTS: When we used the manufacturers’ provided dial setting of 450 × 10, each of the dose calibrators underestimated the radioactivity of (177)Lu by about 9.0%. Dial settings of 403 × 10 and 408 × 10 for (177)Lu on CRC-15R and CRC-ultra dose calibrators respectively were calculated experimentally using an iterative approach. The radioactivity measurements made at these settings provided an excellent agreement with the specified values. Likewise, a dial setting of 230 for each of the two dose calibrators was calculated for (153)Sm, which provided a good agreement between the experimentally derived radioactivity values and the certified values. A deviation of ± 5.0% was observed when radioactivity of (177)Lu and (153)Sm was measured over a wide range (4.0 MBq to 2.1 GBq) for time intervals equivalent to 4.5 half-lives of each of the two radionuclides. A deviation of ± 5% was observed when radioactivity was counted in different dilution volumes and in syringes of varying size. CONCLUSION: These variations could lead to a cumulative error of about 20.0% toward the inaccuracy in the radioactivity measurements of the beta-gamma emitters and thus predefined calibration factors of the dose calibrators may require experimental re-setting of these parameters and periodic checking to provide accurate radioactivity estimates of beta-gamma emitters in a given clinical setting.
format Online
Article
Text
id pubmed-4379670
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-43796702015-04-01 Deviation in the predefined calibration factors of the dose calibrators and the associated inaccuracy in the radioactivity measurements of beta-gamma emitters Sharma, Sarika Singh, Baljinder Koul, Ashwani Mittal, Bhagwant Rai Indian J Nucl Med Original Article AIM: To determine whether the predefined calibration factors of the dose calibrators can provide accurate radioactivity measurements of beta-gamma emitters used in routine therapeutic nuclear medicine procedures. MATERIALS AND METHODS: Two models of dose calibrators were used in the present study for radioactivity measurements of (153)Sm ethylenediamine-N, N, N’, N’-tetrakis methylene phosphonic acid (EDTMP) and (177)Lu (EDTMP). A known (precalibrated) activity of each of the two beta emitters received by us from our National Supplier for administration to the patients with extensive bony metastases for bone pain palliation, was used for experiments. RESULTS: When we used the manufacturers’ provided dial setting of 450 × 10, each of the dose calibrators underestimated the radioactivity of (177)Lu by about 9.0%. Dial settings of 403 × 10 and 408 × 10 for (177)Lu on CRC-15R and CRC-ultra dose calibrators respectively were calculated experimentally using an iterative approach. The radioactivity measurements made at these settings provided an excellent agreement with the specified values. Likewise, a dial setting of 230 for each of the two dose calibrators was calculated for (153)Sm, which provided a good agreement between the experimentally derived radioactivity values and the certified values. A deviation of ± 5.0% was observed when radioactivity of (177)Lu and (153)Sm was measured over a wide range (4.0 MBq to 2.1 GBq) for time intervals equivalent to 4.5 half-lives of each of the two radionuclides. A deviation of ± 5% was observed when radioactivity was counted in different dilution volumes and in syringes of varying size. CONCLUSION: These variations could lead to a cumulative error of about 20.0% toward the inaccuracy in the radioactivity measurements of the beta-gamma emitters and thus predefined calibration factors of the dose calibrators may require experimental re-setting of these parameters and periodic checking to provide accurate radioactivity estimates of beta-gamma emitters in a given clinical setting. Medknow Publications & Media Pvt Ltd 2015 /pmc/articles/PMC4379670/ /pubmed/25829729 http://dx.doi.org/10.4103/0972-3919.152972 Text en Copyright: © Indian Journal of Nuclear Medicine http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Sharma, Sarika
Singh, Baljinder
Koul, Ashwani
Mittal, Bhagwant Rai
Deviation in the predefined calibration factors of the dose calibrators and the associated inaccuracy in the radioactivity measurements of beta-gamma emitters
title Deviation in the predefined calibration factors of the dose calibrators and the associated inaccuracy in the radioactivity measurements of beta-gamma emitters
title_full Deviation in the predefined calibration factors of the dose calibrators and the associated inaccuracy in the radioactivity measurements of beta-gamma emitters
title_fullStr Deviation in the predefined calibration factors of the dose calibrators and the associated inaccuracy in the radioactivity measurements of beta-gamma emitters
title_full_unstemmed Deviation in the predefined calibration factors of the dose calibrators and the associated inaccuracy in the radioactivity measurements of beta-gamma emitters
title_short Deviation in the predefined calibration factors of the dose calibrators and the associated inaccuracy in the radioactivity measurements of beta-gamma emitters
title_sort deviation in the predefined calibration factors of the dose calibrators and the associated inaccuracy in the radioactivity measurements of beta-gamma emitters
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4379670/
https://www.ncbi.nlm.nih.gov/pubmed/25829729
http://dx.doi.org/10.4103/0972-3919.152972
work_keys_str_mv AT sharmasarika deviationinthepredefinedcalibrationfactorsofthedosecalibratorsandtheassociatedinaccuracyintheradioactivitymeasurementsofbetagammaemitters
AT singhbaljinder deviationinthepredefinedcalibrationfactorsofthedosecalibratorsandtheassociatedinaccuracyintheradioactivitymeasurementsofbetagammaemitters
AT koulashwani deviationinthepredefinedcalibrationfactorsofthedosecalibratorsandtheassociatedinaccuracyintheradioactivitymeasurementsofbetagammaemitters
AT mittalbhagwantrai deviationinthepredefinedcalibrationfactorsofthedosecalibratorsandtheassociatedinaccuracyintheradioactivitymeasurementsofbetagammaemitters