Operational evaluation of the earlobe arterialized blood collector in critically ill patients

BACKGROUND: The new Earlobe Arterialized Blood Collector (EABC®) is a minimally invasive prototype system able to perform capillary blood collection from the earlobe (EL) with minimal training and risk. This system could improve medical emergency management in extreme environments. Consequently, a p...

Descripción completa

Detalles Bibliográficos
Autores principales: Vaquer, Sergi, Masip, Jordi, Gili, Gisela, Gomà, Gemma, Oliva, Joan Carles, Frechette, Alexandre, Evetts, Simon, Russomano, Thais, Artigas, Antonio
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4382925/
https://www.ncbi.nlm.nih.gov/pubmed/25838884
http://dx.doi.org/10.1186/s13728-015-0025-x
Descripción
Sumario:BACKGROUND: The new Earlobe Arterialized Blood Collector (EABC®) is a minimally invasive prototype system able to perform capillary blood collection from the earlobe (EL) with minimal training and risk. This system could improve medical emergency management in extreme environments. Consequently, a prospective validation study was designed to evaluate operational performance of the EABC® in a cohort of critically ill patients. METHODS: Arterialized capillary blood was sampled from the EL of 55 invasively ventilated patients using the EABC® following a validated procedure. Operational characteristics such as the number of cuts and cartridges required, sampling failure/success ratio, bleeding complications, storage requirements and other auxiliary aspects were recorded. Result turnaround laboratory times (TAT) were compared with published references. RESULTS: Blood collection was as easily performed on one earlobe as the other. Twenty-six minutes (mean 25.8; SD = 3.8) were required to obtain results, 15 min for patient preparation (mean 15.3; SD = 2.6) + 11 min for sampling and analysis (mean 11.4; SD = 2.1), which is similar to published hospital reference laboratory TAT. The average number of cartridges required was 1.3 (1–3; mode = 1) with the mean number of cut attempts being 1.2 (1–4; mode = 1). Problems/difficulties occurred in 59% of cases but were mainly attributed to patient’s demographic characteristics, with only 10% attributable to the collector (superficial cut, blood leak, collector misalignment and obstructed vision). Haemostasis was quickly achieved with minimum complications. Storage of the complete sampling kit required a 300 × 300 × 300 mm box. Two 9-V batteries were used during the 2-year study period. CONCLUSIONS: The new EABC® system concept is safe, fast and easy to use. Observed problems/difficulties are easily amendable with certain design modifications. Definitive versions of the prototype have the potential for significant benefits for isolated and extreme environments in medicine.