Accurate Measurement of Canal Length during Root Canal Treatment: An In Vivo Study
OBJECTIVES: To assess the consistency and accuracy of Electronic Apex Locator (EAL) (Root ZXII) in individual canals and its association with other clinical variables. STUDY DESIGN: Cross-Sectional study. PLACE OF STUDY: Dental section of the Aga Khan University Hospital, Karachi, Pakistan. MATERIAL...
Autores principales: | , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Master Publishing Group
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4392562/ |
Sumario: | OBJECTIVES: To assess the consistency and accuracy of Electronic Apex Locator (EAL) (Root ZXII) in individual canals and its association with other clinical variables. STUDY DESIGN: Cross-Sectional study. PLACE OF STUDY: Dental section of the Aga Khan University Hospital, Karachi, Pakistan. MATERIALS AND METHODS: Working length was measured by EAL in 180 patients requiring endodontic therapy in molar and premolar teeth. The effects of clinical variables e.g. gender and pulpal status on the consistency and accuracy of EAL were recorded. Performance of apex locator was considered “Consistent” when the scale bar was stable and moved only in correspondence to the movement of file in the root canal. Accuracy was determined by inserting the file at the working length determined by the EAL and periapical view of radiograph was taken using paralleling technique. Estimated working length was considered accurate when the file tip was located 0-2mm short of the radiographic apex. If the file was overextended from the radiographic apex, it showed dysfunction of the EAL. RESULTS: Consistency of EAL was found 97.6% in distobuccal canals, 91.1% in palatal canals, 73.7% in mesiolingual canals, 83.3% in mesiobuccal and 80.2% in distal canals. Accuracy of EAL was 91.4% in mesiolingual canal, 92% in mesiobuccal, and 90.2% in Palatal and 93.2% in distal canal. CONCLUSION: Consistency of electronic apex locator vary in different canals, however consistent measurements are highly accurate. No significant association was found between other clinical variables with the consistency and accuracy of EAL. |
---|