Cargando…
Closed-Loop Control Better than Open-Loop Control of Profofol TCI Guided by BIS: A Randomized, Controlled, Multicenter Clinical Trial to Evaluate the CONCERT-CL Closed-Loop System
BACKGROUND: The CONCERT-CL closed-loop infusion system designed by VERYARK Technology Co., Ltd. (Guangxi, China) is an innovation using TCI combined with closed-loop controlled intravenous anesthesia under the guide of BIS. In this study we performed a randomized, controlled, multicenter study to co...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4401751/ https://www.ncbi.nlm.nih.gov/pubmed/25886041 http://dx.doi.org/10.1371/journal.pone.0123862 |
_version_ | 1782367187853377536 |
---|---|
author | Liu, Yu Li, Min Yang, Dong Zhang, Xuena Wu, Anshi Yao, Shanglong Xue, Zhanggang Yue, Yun |
author_facet | Liu, Yu Li, Min Yang, Dong Zhang, Xuena Wu, Anshi Yao, Shanglong Xue, Zhanggang Yue, Yun |
author_sort | Liu, Yu |
collection | PubMed |
description | BACKGROUND: The CONCERT-CL closed-loop infusion system designed by VERYARK Technology Co., Ltd. (Guangxi, China) is an innovation using TCI combined with closed-loop controlled intravenous anesthesia under the guide of BIS. In this study we performed a randomized, controlled, multicenter study to compare closed-loop control and open-loop control of propofol by using the CONCERT-CL closed-loop infusion system. METHODS: 180 surgical patients from three medical centers undergone TCI intravenous anesthesia with propofol and remifentanil were randomly assigned to propofol closed-loop group and propofol opened-loop groups. Primary outcome was global score (GS, GS = (MDAPE+Wobble)/% of time of bispectral index (BIS) 40-60). Secondary outcomes were doses of the anesthetics and emergence time from anesthesia, such as, time to tracheal extubation. RESULTS: There were 89 and 86 patients in the closed-loop and opened-loop groups, respectively. GS in the closed-loop groups (22.21±8.50) were lower than that in the opened-loop group (27.19±15.26) (p=0.009). The higher proportion of time of BIS between 40 and 60 was also observed in the closed-loop group (84.11±9.50%), while that was 79.92±13.17% in the opened-loop group, (p=0.016). No significant differences in propofol dose and time of tracheal extubation were observed. The frequency of propofol regulation in the closed-loop group (31.55±9.46 times/hr) was obverse higher than that in the opened-loop group (6.84±6.21 times/hr) (p=0.000). CONCLUSION: The CONCERT-CL closed-loop infusion system can automatically regulate the TCI of propofol, maintain the BIS value in an adequate range and reduce the workload of anesthesiologists better than open-loop system. TRIAL REGISTRATION: ChiCTR ChiCTR-OOR-14005551 |
format | Online Article Text |
id | pubmed-4401751 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-44017512015-04-21 Closed-Loop Control Better than Open-Loop Control of Profofol TCI Guided by BIS: A Randomized, Controlled, Multicenter Clinical Trial to Evaluate the CONCERT-CL Closed-Loop System Liu, Yu Li, Min Yang, Dong Zhang, Xuena Wu, Anshi Yao, Shanglong Xue, Zhanggang Yue, Yun PLoS One Research Article BACKGROUND: The CONCERT-CL closed-loop infusion system designed by VERYARK Technology Co., Ltd. (Guangxi, China) is an innovation using TCI combined with closed-loop controlled intravenous anesthesia under the guide of BIS. In this study we performed a randomized, controlled, multicenter study to compare closed-loop control and open-loop control of propofol by using the CONCERT-CL closed-loop infusion system. METHODS: 180 surgical patients from three medical centers undergone TCI intravenous anesthesia with propofol and remifentanil were randomly assigned to propofol closed-loop group and propofol opened-loop groups. Primary outcome was global score (GS, GS = (MDAPE+Wobble)/% of time of bispectral index (BIS) 40-60). Secondary outcomes were doses of the anesthetics and emergence time from anesthesia, such as, time to tracheal extubation. RESULTS: There were 89 and 86 patients in the closed-loop and opened-loop groups, respectively. GS in the closed-loop groups (22.21±8.50) were lower than that in the opened-loop group (27.19±15.26) (p=0.009). The higher proportion of time of BIS between 40 and 60 was also observed in the closed-loop group (84.11±9.50%), while that was 79.92±13.17% in the opened-loop group, (p=0.016). No significant differences in propofol dose and time of tracheal extubation were observed. The frequency of propofol regulation in the closed-loop group (31.55±9.46 times/hr) was obverse higher than that in the opened-loop group (6.84±6.21 times/hr) (p=0.000). CONCLUSION: The CONCERT-CL closed-loop infusion system can automatically regulate the TCI of propofol, maintain the BIS value in an adequate range and reduce the workload of anesthesiologists better than open-loop system. TRIAL REGISTRATION: ChiCTR ChiCTR-OOR-14005551 Public Library of Science 2015-04-17 /pmc/articles/PMC4401751/ /pubmed/25886041 http://dx.doi.org/10.1371/journal.pone.0123862 Text en © 2015 Liu et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited. |
spellingShingle | Research Article Liu, Yu Li, Min Yang, Dong Zhang, Xuena Wu, Anshi Yao, Shanglong Xue, Zhanggang Yue, Yun Closed-Loop Control Better than Open-Loop Control of Profofol TCI Guided by BIS: A Randomized, Controlled, Multicenter Clinical Trial to Evaluate the CONCERT-CL Closed-Loop System |
title | Closed-Loop Control Better than Open-Loop Control of Profofol TCI Guided by BIS: A Randomized, Controlled, Multicenter Clinical Trial to Evaluate the CONCERT-CL Closed-Loop System |
title_full | Closed-Loop Control Better than Open-Loop Control of Profofol TCI Guided by BIS: A Randomized, Controlled, Multicenter Clinical Trial to Evaluate the CONCERT-CL Closed-Loop System |
title_fullStr | Closed-Loop Control Better than Open-Loop Control of Profofol TCI Guided by BIS: A Randomized, Controlled, Multicenter Clinical Trial to Evaluate the CONCERT-CL Closed-Loop System |
title_full_unstemmed | Closed-Loop Control Better than Open-Loop Control of Profofol TCI Guided by BIS: A Randomized, Controlled, Multicenter Clinical Trial to Evaluate the CONCERT-CL Closed-Loop System |
title_short | Closed-Loop Control Better than Open-Loop Control of Profofol TCI Guided by BIS: A Randomized, Controlled, Multicenter Clinical Trial to Evaluate the CONCERT-CL Closed-Loop System |
title_sort | closed-loop control better than open-loop control of profofol tci guided by bis: a randomized, controlled, multicenter clinical trial to evaluate the concert-cl closed-loop system |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4401751/ https://www.ncbi.nlm.nih.gov/pubmed/25886041 http://dx.doi.org/10.1371/journal.pone.0123862 |
work_keys_str_mv | AT liuyu closedloopcontrolbetterthanopenloopcontrolofprofofoltciguidedbybisarandomizedcontrolledmulticenterclinicaltrialtoevaluatetheconcertclclosedloopsystem AT limin closedloopcontrolbetterthanopenloopcontrolofprofofoltciguidedbybisarandomizedcontrolledmulticenterclinicaltrialtoevaluatetheconcertclclosedloopsystem AT yangdong closedloopcontrolbetterthanopenloopcontrolofprofofoltciguidedbybisarandomizedcontrolledmulticenterclinicaltrialtoevaluatetheconcertclclosedloopsystem AT zhangxuena closedloopcontrolbetterthanopenloopcontrolofprofofoltciguidedbybisarandomizedcontrolledmulticenterclinicaltrialtoevaluatetheconcertclclosedloopsystem AT wuanshi closedloopcontrolbetterthanopenloopcontrolofprofofoltciguidedbybisarandomizedcontrolledmulticenterclinicaltrialtoevaluatetheconcertclclosedloopsystem AT yaoshanglong closedloopcontrolbetterthanopenloopcontrolofprofofoltciguidedbybisarandomizedcontrolledmulticenterclinicaltrialtoevaluatetheconcertclclosedloopsystem AT xuezhanggang closedloopcontrolbetterthanopenloopcontrolofprofofoltciguidedbybisarandomizedcontrolledmulticenterclinicaltrialtoevaluatetheconcertclclosedloopsystem AT yueyun closedloopcontrolbetterthanopenloopcontrolofprofofoltciguidedbybisarandomizedcontrolledmulticenterclinicaltrialtoevaluatetheconcertclclosedloopsystem |