Evaluation of the Nova StatSensor(®) Xpress(TM) Creatinine Point-Of-Care Handheld Analyzer

Creatinine is a parameter that is required to monitor renal function and is important to follow in patients under treatment with potentially toxic renal drugs, such as the anti-HIV drug Tenofovir. A point of care instrument to measure creatinine would be useful for patients monitoring in resource-li...

Descripción completa

Detalles Bibliográficos
Autores principales: Kosack, Cara Simone, de Kieviet, Wim, Bayrak, Kubra, Milovic, Anastacija, Page, Anne Laure
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4401790/
https://www.ncbi.nlm.nih.gov/pubmed/25886375
http://dx.doi.org/10.1371/journal.pone.0122433
_version_ 1782367196884762624
author Kosack, Cara Simone
de Kieviet, Wim
Bayrak, Kubra
Milovic, Anastacija
Page, Anne Laure
author_facet Kosack, Cara Simone
de Kieviet, Wim
Bayrak, Kubra
Milovic, Anastacija
Page, Anne Laure
author_sort Kosack, Cara Simone
collection PubMed
description Creatinine is a parameter that is required to monitor renal function and is important to follow in patients under treatment with potentially toxic renal drugs, such as the anti-HIV drug Tenofovir. A point of care instrument to measure creatinine would be useful for patients monitoring in resource-limited settings, where more instruments that are sophisticated are not available. The StatSensor Xpress Creatinine (Nova Biomedical Cooperation, Waltham, MA, USA) point of care analyzer was evaluated for its diagnostic performance in indicating drug therapy change. Creatinine was measured in parallel using the Nova StatSensor Xpress Creatinine analyzer and the Vitros 5,1FS (Ortho Clinical Diagnostics, Inc, Rochester, USA), which served as reference standard. The precision (i.e., repeatability and reproducibility) and accuracy of the StatSensor Xpress Creatinine analyzer were calculated using a panel of specimens with normal, low pathological and high pathological values. Two different Nova StatSensor Xpress Creatinine analyzers were used for the assessment of accuracy using repeated measurements. The coefficient of variation of the StatSensor Xpress Creatinine analyzers ranged from 2.3 to 5.9% for repeatability and from 4.2 to 9.0% for between-run reproducibility. The concordance correlation agreement was good except for high values (>600 µmol/L). The Bland-Altman analysis in high pathological specimens suggests that the Nova StatSensor Xpress Creatinine test tends to underestimate high creatinine values (i.e., >600 µmol/L). The Nova StatSensor Xpress Creatinine analyzers showed acceptable to good results in terms of repeatability, inter-device reproducibility and between-run reproducibility over time using quality control reagents. The analyzer was found sufficiently accurate for detecting pathological values in patients (age >10 year) and can be used with a moderate risk of misclassification.
format Online
Article
Text
id pubmed-4401790
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-44017902015-04-21 Evaluation of the Nova StatSensor(®) Xpress(TM) Creatinine Point-Of-Care Handheld Analyzer Kosack, Cara Simone de Kieviet, Wim Bayrak, Kubra Milovic, Anastacija Page, Anne Laure PLoS One Research Article Creatinine is a parameter that is required to monitor renal function and is important to follow in patients under treatment with potentially toxic renal drugs, such as the anti-HIV drug Tenofovir. A point of care instrument to measure creatinine would be useful for patients monitoring in resource-limited settings, where more instruments that are sophisticated are not available. The StatSensor Xpress Creatinine (Nova Biomedical Cooperation, Waltham, MA, USA) point of care analyzer was evaluated for its diagnostic performance in indicating drug therapy change. Creatinine was measured in parallel using the Nova StatSensor Xpress Creatinine analyzer and the Vitros 5,1FS (Ortho Clinical Diagnostics, Inc, Rochester, USA), which served as reference standard. The precision (i.e., repeatability and reproducibility) and accuracy of the StatSensor Xpress Creatinine analyzer were calculated using a panel of specimens with normal, low pathological and high pathological values. Two different Nova StatSensor Xpress Creatinine analyzers were used for the assessment of accuracy using repeated measurements. The coefficient of variation of the StatSensor Xpress Creatinine analyzers ranged from 2.3 to 5.9% for repeatability and from 4.2 to 9.0% for between-run reproducibility. The concordance correlation agreement was good except for high values (>600 µmol/L). The Bland-Altman analysis in high pathological specimens suggests that the Nova StatSensor Xpress Creatinine test tends to underestimate high creatinine values (i.e., >600 µmol/L). The Nova StatSensor Xpress Creatinine analyzers showed acceptable to good results in terms of repeatability, inter-device reproducibility and between-run reproducibility over time using quality control reagents. The analyzer was found sufficiently accurate for detecting pathological values in patients (age >10 year) and can be used with a moderate risk of misclassification. Public Library of Science 2015-04-17 /pmc/articles/PMC4401790/ /pubmed/25886375 http://dx.doi.org/10.1371/journal.pone.0122433 Text en © 2015 Kosack et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited.
spellingShingle Research Article
Kosack, Cara Simone
de Kieviet, Wim
Bayrak, Kubra
Milovic, Anastacija
Page, Anne Laure
Evaluation of the Nova StatSensor(®) Xpress(TM) Creatinine Point-Of-Care Handheld Analyzer
title Evaluation of the Nova StatSensor(®) Xpress(TM) Creatinine Point-Of-Care Handheld Analyzer
title_full Evaluation of the Nova StatSensor(®) Xpress(TM) Creatinine Point-Of-Care Handheld Analyzer
title_fullStr Evaluation of the Nova StatSensor(®) Xpress(TM) Creatinine Point-Of-Care Handheld Analyzer
title_full_unstemmed Evaluation of the Nova StatSensor(®) Xpress(TM) Creatinine Point-Of-Care Handheld Analyzer
title_short Evaluation of the Nova StatSensor(®) Xpress(TM) Creatinine Point-Of-Care Handheld Analyzer
title_sort evaluation of the nova statsensor(®) xpress(tm) creatinine point-of-care handheld analyzer
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4401790/
https://www.ncbi.nlm.nih.gov/pubmed/25886375
http://dx.doi.org/10.1371/journal.pone.0122433
work_keys_str_mv AT kosackcarasimone evaluationofthenovastatsensorxpresstmcreatininepointofcarehandheldanalyzer
AT dekievietwim evaluationofthenovastatsensorxpresstmcreatininepointofcarehandheldanalyzer
AT bayrakkubra evaluationofthenovastatsensorxpresstmcreatininepointofcarehandheldanalyzer
AT milovicanastacija evaluationofthenovastatsensorxpresstmcreatininepointofcarehandheldanalyzer
AT pageannelaure evaluationofthenovastatsensorxpresstmcreatininepointofcarehandheldanalyzer