Field Evaluation of a Point-of-Care CD4 Analyzer for Monitoring HIV Patients in the Interior of the Amazon Region, Brazil
OBJECTIVE: To evaluate the accuracy of the PIMA point-of-care CD4 analyzer (PIMA) under field conditions in comparison to the current CD4 count system (FACSCalibur), and to evaluate the operational suitability and acceptability of health professionals (HP) and HIV-patients in using the PIMA in healt...
Autores principales: | , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4408088/ https://www.ncbi.nlm.nih.gov/pubmed/25905775 http://dx.doi.org/10.1371/journal.pone.0121400 |
_version_ | 1782368015995633664 |
---|---|
author | Pinto, Ione Conceição Sabidó, Meritxell Pereira, Analice Barbosa Mello, Maeve B. de Melo Xavier Shimizu, Andrea Protti, Bruna Lovizutto Benzaken, Adele Schwartz |
author_facet | Pinto, Ione Conceição Sabidó, Meritxell Pereira, Analice Barbosa Mello, Maeve B. de Melo Xavier Shimizu, Andrea Protti, Bruna Lovizutto Benzaken, Adele Schwartz |
author_sort | Pinto, Ione Conceição |
collection | PubMed |
description | OBJECTIVE: To evaluate the accuracy of the PIMA point-of-care CD4 analyzer (PIMA) under field conditions in comparison to the current CD4 count system (FACSCalibur), and to evaluate the operational suitability and acceptability of health professionals (HP) and HIV-patients in using the PIMA in health clinics in the Amazon Region. METHODS: CD4 counts were measured onsite by the PIMA using fingerprick blood and in the reference laboratory by both the PIMA and FACSCalibur using venous blood. We used the Bland–Altman method to estimate the mean bias, and calculated the sensitivity and specificity at <200 and <500 cell/μL thresholds. Patients (n = 404) and HP (n = 7) were interviewed on the acceptability and operational suitability of the PIMA. RESULTS: Using fingerprick blood (n = 337), the PIMA showed a concordance correlation coefficient (Rc) of 0.81, mean difference of -111.9 cell/μL, 93.1%/98.5% sensitivity, and 89.2%/56.7% specificity at <200 and <500 cell/μL thresholds, respectively. Venous blood (n = 340) showed an Rc of 0.89, mean difference of -83.4 cell/μL, 98.3%/97.5% sensitivity, and 93.9%/66.0% specificity at <200 and <500 cell/μL thresholds, respectively. The capillary PIMA was well accepted and found operationally appropriate by patients and HP. CONCLUSIONS: The agreement between both instruments was poor and the PIMA underestimated CD4 cell counts, which was more pronounced at CD4 counts ≥500 cell/μl. The PIMA’s performance with fingerprick blood was less reliable than its performance with venous blood. In Brazil, where antiretroviral treatment is initiated regardless of CD4 counts, the PIMA’s systematic bias towards CD4 underestimation may limit its role for monitoring HIV-patients. |
format | Online Article Text |
id | pubmed-4408088 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-44080882015-05-04 Field Evaluation of a Point-of-Care CD4 Analyzer for Monitoring HIV Patients in the Interior of the Amazon Region, Brazil Pinto, Ione Conceição Sabidó, Meritxell Pereira, Analice Barbosa Mello, Maeve B. de Melo Xavier Shimizu, Andrea Protti, Bruna Lovizutto Benzaken, Adele Schwartz PLoS One Research Article OBJECTIVE: To evaluate the accuracy of the PIMA point-of-care CD4 analyzer (PIMA) under field conditions in comparison to the current CD4 count system (FACSCalibur), and to evaluate the operational suitability and acceptability of health professionals (HP) and HIV-patients in using the PIMA in health clinics in the Amazon Region. METHODS: CD4 counts were measured onsite by the PIMA using fingerprick blood and in the reference laboratory by both the PIMA and FACSCalibur using venous blood. We used the Bland–Altman method to estimate the mean bias, and calculated the sensitivity and specificity at <200 and <500 cell/μL thresholds. Patients (n = 404) and HP (n = 7) were interviewed on the acceptability and operational suitability of the PIMA. RESULTS: Using fingerprick blood (n = 337), the PIMA showed a concordance correlation coefficient (Rc) of 0.81, mean difference of -111.9 cell/μL, 93.1%/98.5% sensitivity, and 89.2%/56.7% specificity at <200 and <500 cell/μL thresholds, respectively. Venous blood (n = 340) showed an Rc of 0.89, mean difference of -83.4 cell/μL, 98.3%/97.5% sensitivity, and 93.9%/66.0% specificity at <200 and <500 cell/μL thresholds, respectively. The capillary PIMA was well accepted and found operationally appropriate by patients and HP. CONCLUSIONS: The agreement between both instruments was poor and the PIMA underestimated CD4 cell counts, which was more pronounced at CD4 counts ≥500 cell/μl. The PIMA’s performance with fingerprick blood was less reliable than its performance with venous blood. In Brazil, where antiretroviral treatment is initiated regardless of CD4 counts, the PIMA’s systematic bias towards CD4 underestimation may limit its role for monitoring HIV-patients. Public Library of Science 2015-04-23 /pmc/articles/PMC4408088/ /pubmed/25905775 http://dx.doi.org/10.1371/journal.pone.0121400 Text en © 2015 Pinto et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited. |
spellingShingle | Research Article Pinto, Ione Conceição Sabidó, Meritxell Pereira, Analice Barbosa Mello, Maeve B. de Melo Xavier Shimizu, Andrea Protti, Bruna Lovizutto Benzaken, Adele Schwartz Field Evaluation of a Point-of-Care CD4 Analyzer for Monitoring HIV Patients in the Interior of the Amazon Region, Brazil |
title | Field Evaluation of a Point-of-Care CD4 Analyzer for Monitoring HIV Patients in the Interior of the Amazon Region, Brazil |
title_full | Field Evaluation of a Point-of-Care CD4 Analyzer for Monitoring HIV Patients in the Interior of the Amazon Region, Brazil |
title_fullStr | Field Evaluation of a Point-of-Care CD4 Analyzer for Monitoring HIV Patients in the Interior of the Amazon Region, Brazil |
title_full_unstemmed | Field Evaluation of a Point-of-Care CD4 Analyzer for Monitoring HIV Patients in the Interior of the Amazon Region, Brazil |
title_short | Field Evaluation of a Point-of-Care CD4 Analyzer for Monitoring HIV Patients in the Interior of the Amazon Region, Brazil |
title_sort | field evaluation of a point-of-care cd4 analyzer for monitoring hiv patients in the interior of the amazon region, brazil |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4408088/ https://www.ncbi.nlm.nih.gov/pubmed/25905775 http://dx.doi.org/10.1371/journal.pone.0121400 |
work_keys_str_mv | AT pintoioneconceicao fieldevaluationofapointofcarecd4analyzerformonitoringhivpatientsintheinterioroftheamazonregionbrazil AT sabidomeritxell fieldevaluationofapointofcarecd4analyzerformonitoringhivpatientsintheinterioroftheamazonregionbrazil AT pereiraanalicebarbosa fieldevaluationofapointofcarecd4analyzerformonitoringhivpatientsintheinterioroftheamazonregionbrazil AT mellomaeveb fieldevaluationofapointofcarecd4analyzerformonitoringhivpatientsintheinterioroftheamazonregionbrazil AT demeloxaviershimizuandrea fieldevaluationofapointofcarecd4analyzerformonitoringhivpatientsintheinterioroftheamazonregionbrazil AT prottibrunalovizutto fieldevaluationofapointofcarecd4analyzerformonitoringhivpatientsintheinterioroftheamazonregionbrazil AT benzakenadeleschwartz fieldevaluationofapointofcarecd4analyzerformonitoringhivpatientsintheinterioroftheamazonregionbrazil |