Field Evaluation of a Point-of-Care CD4 Analyzer for Monitoring HIV Patients in the Interior of the Amazon Region, Brazil

OBJECTIVE: To evaluate the accuracy of the PIMA point-of-care CD4 analyzer (PIMA) under field conditions in comparison to the current CD4 count system (FACSCalibur), and to evaluate the operational suitability and acceptability of health professionals (HP) and HIV-patients in using the PIMA in healt...

Descripción completa

Detalles Bibliográficos
Autores principales: Pinto, Ione Conceição, Sabidó, Meritxell, Pereira, Analice Barbosa, Mello, Maeve B., de Melo Xavier Shimizu, Andrea, Protti, Bruna Lovizutto, Benzaken, Adele Schwartz
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4408088/
https://www.ncbi.nlm.nih.gov/pubmed/25905775
http://dx.doi.org/10.1371/journal.pone.0121400
_version_ 1782368015995633664
author Pinto, Ione Conceição
Sabidó, Meritxell
Pereira, Analice Barbosa
Mello, Maeve B.
de Melo Xavier Shimizu, Andrea
Protti, Bruna Lovizutto
Benzaken, Adele Schwartz
author_facet Pinto, Ione Conceição
Sabidó, Meritxell
Pereira, Analice Barbosa
Mello, Maeve B.
de Melo Xavier Shimizu, Andrea
Protti, Bruna Lovizutto
Benzaken, Adele Schwartz
author_sort Pinto, Ione Conceição
collection PubMed
description OBJECTIVE: To evaluate the accuracy of the PIMA point-of-care CD4 analyzer (PIMA) under field conditions in comparison to the current CD4 count system (FACSCalibur), and to evaluate the operational suitability and acceptability of health professionals (HP) and HIV-patients in using the PIMA in health clinics in the Amazon Region. METHODS: CD4 counts were measured onsite by the PIMA using fingerprick blood and in the reference laboratory by both the PIMA and FACSCalibur using venous blood. We used the Bland–Altman method to estimate the mean bias, and calculated the sensitivity and specificity at <200 and <500 cell/μL thresholds. Patients (n = 404) and HP (n = 7) were interviewed on the acceptability and operational suitability of the PIMA. RESULTS: Using fingerprick blood (n = 337), the PIMA showed a concordance correlation coefficient (Rc) of 0.81, mean difference of -111.9 cell/μL, 93.1%/98.5% sensitivity, and 89.2%/56.7% specificity at <200 and <500 cell/μL thresholds, respectively. Venous blood (n = 340) showed an Rc of 0.89, mean difference of -83.4 cell/μL, 98.3%/97.5% sensitivity, and 93.9%/66.0% specificity at <200 and <500 cell/μL thresholds, respectively. The capillary PIMA was well accepted and found operationally appropriate by patients and HP. CONCLUSIONS: The agreement between both instruments was poor and the PIMA underestimated CD4 cell counts, which was more pronounced at CD4 counts ≥500 cell/μl. The PIMA’s performance with fingerprick blood was less reliable than its performance with venous blood. In Brazil, where antiretroviral treatment is initiated regardless of CD4 counts, the PIMA’s systematic bias towards CD4 underestimation may limit its role for monitoring HIV-patients.
format Online
Article
Text
id pubmed-4408088
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-44080882015-05-04 Field Evaluation of a Point-of-Care CD4 Analyzer for Monitoring HIV Patients in the Interior of the Amazon Region, Brazil Pinto, Ione Conceição Sabidó, Meritxell Pereira, Analice Barbosa Mello, Maeve B. de Melo Xavier Shimizu, Andrea Protti, Bruna Lovizutto Benzaken, Adele Schwartz PLoS One Research Article OBJECTIVE: To evaluate the accuracy of the PIMA point-of-care CD4 analyzer (PIMA) under field conditions in comparison to the current CD4 count system (FACSCalibur), and to evaluate the operational suitability and acceptability of health professionals (HP) and HIV-patients in using the PIMA in health clinics in the Amazon Region. METHODS: CD4 counts were measured onsite by the PIMA using fingerprick blood and in the reference laboratory by both the PIMA and FACSCalibur using venous blood. We used the Bland–Altman method to estimate the mean bias, and calculated the sensitivity and specificity at <200 and <500 cell/μL thresholds. Patients (n = 404) and HP (n = 7) were interviewed on the acceptability and operational suitability of the PIMA. RESULTS: Using fingerprick blood (n = 337), the PIMA showed a concordance correlation coefficient (Rc) of 0.81, mean difference of -111.9 cell/μL, 93.1%/98.5% sensitivity, and 89.2%/56.7% specificity at <200 and <500 cell/μL thresholds, respectively. Venous blood (n = 340) showed an Rc of 0.89, mean difference of -83.4 cell/μL, 98.3%/97.5% sensitivity, and 93.9%/66.0% specificity at <200 and <500 cell/μL thresholds, respectively. The capillary PIMA was well accepted and found operationally appropriate by patients and HP. CONCLUSIONS: The agreement between both instruments was poor and the PIMA underestimated CD4 cell counts, which was more pronounced at CD4 counts ≥500 cell/μl. The PIMA’s performance with fingerprick blood was less reliable than its performance with venous blood. In Brazil, where antiretroviral treatment is initiated regardless of CD4 counts, the PIMA’s systematic bias towards CD4 underestimation may limit its role for monitoring HIV-patients. Public Library of Science 2015-04-23 /pmc/articles/PMC4408088/ /pubmed/25905775 http://dx.doi.org/10.1371/journal.pone.0121400 Text en © 2015 Pinto et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited.
spellingShingle Research Article
Pinto, Ione Conceição
Sabidó, Meritxell
Pereira, Analice Barbosa
Mello, Maeve B.
de Melo Xavier Shimizu, Andrea
Protti, Bruna Lovizutto
Benzaken, Adele Schwartz
Field Evaluation of a Point-of-Care CD4 Analyzer for Monitoring HIV Patients in the Interior of the Amazon Region, Brazil
title Field Evaluation of a Point-of-Care CD4 Analyzer for Monitoring HIV Patients in the Interior of the Amazon Region, Brazil
title_full Field Evaluation of a Point-of-Care CD4 Analyzer for Monitoring HIV Patients in the Interior of the Amazon Region, Brazil
title_fullStr Field Evaluation of a Point-of-Care CD4 Analyzer for Monitoring HIV Patients in the Interior of the Amazon Region, Brazil
title_full_unstemmed Field Evaluation of a Point-of-Care CD4 Analyzer for Monitoring HIV Patients in the Interior of the Amazon Region, Brazil
title_short Field Evaluation of a Point-of-Care CD4 Analyzer for Monitoring HIV Patients in the Interior of the Amazon Region, Brazil
title_sort field evaluation of a point-of-care cd4 analyzer for monitoring hiv patients in the interior of the amazon region, brazil
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4408088/
https://www.ncbi.nlm.nih.gov/pubmed/25905775
http://dx.doi.org/10.1371/journal.pone.0121400
work_keys_str_mv AT pintoioneconceicao fieldevaluationofapointofcarecd4analyzerformonitoringhivpatientsintheinterioroftheamazonregionbrazil
AT sabidomeritxell fieldevaluationofapointofcarecd4analyzerformonitoringhivpatientsintheinterioroftheamazonregionbrazil
AT pereiraanalicebarbosa fieldevaluationofapointofcarecd4analyzerformonitoringhivpatientsintheinterioroftheamazonregionbrazil
AT mellomaeveb fieldevaluationofapointofcarecd4analyzerformonitoringhivpatientsintheinterioroftheamazonregionbrazil
AT demeloxaviershimizuandrea fieldevaluationofapointofcarecd4analyzerformonitoringhivpatientsintheinterioroftheamazonregionbrazil
AT prottibrunalovizutto fieldevaluationofapointofcarecd4analyzerformonitoringhivpatientsintheinterioroftheamazonregionbrazil
AT benzakenadeleschwartz fieldevaluationofapointofcarecd4analyzerformonitoringhivpatientsintheinterioroftheamazonregionbrazil