The diagnostic performance evaluation of the SD BIOLINE HIV/syphilis Duo rapid test in southern Ethiopia: a cross-sectional study

OBJECTIVE: To determine the diagnostic performance of the SD BIOLINE HIV/syphilis Duo rapid test. DESIGN: A hospital-based cross-sectional study. SETTING: This evaluation was conducted at one of the largest hospitals in southern Ethiopia. PARTICIPANTS: Serum samples obtained from clients attending t...

Descripción completa

Detalles Bibliográficos
Autores principales: Shimelis, Techalew, Tadesse, Endale
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BMJ Publishing Group 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4410125/
https://www.ncbi.nlm.nih.gov/pubmed/25908677
http://dx.doi.org/10.1136/bmjopen-2014-007371
_version_ 1782368275606274048
author Shimelis, Techalew
Tadesse, Endale
author_facet Shimelis, Techalew
Tadesse, Endale
author_sort Shimelis, Techalew
collection PubMed
description OBJECTIVE: To determine the diagnostic performance of the SD BIOLINE HIV/syphilis Duo rapid test. DESIGN: A hospital-based cross-sectional study. SETTING: This evaluation was conducted at one of the largest hospitals in southern Ethiopia. PARTICIPANTS: Serum samples obtained from clients attending the antiretroviral therapy and voluntary counselling and testing centres were used. Sera were originally collected for the purpose of investigating syphilis epidemiology. The performance of the test to detect HIV was evaluated using 400 sera (200 HIV positives and 200 HIV negatives). Also, its performance to detect syphilis was evaluated using 85 syphilis positive and 100 syphilis negative serum samples. Individuals <15 years of age or syphilis treated or those with ≤50 cells/µL CD4 cell count were originally excluded. OUTCOME MEASURES: HIV screening was carried out according to the national rapid diagnostic testing (RDT) algorithm: Shenghai Kehua Bioengineering (KHB) test kit as a screening test, followed by the HIV1/2 STAT-PAK assay if positive. Where the result of the STAT-PAK is discordant with KHB, Unigold HIV is used as a tiebreaker to determine the result. We also used ELISA to resolve discordant HIV results. Syphilis serostatus was determined using the Treponema pallidum haemagglutination assay (TPHA). RESULTS: The respective sensitivity, specificity, positive predictive value and negative predictive value of the SD BIOLINE HIV/syphilis Duo test were 100, 99.5, 99.5 and 100% for HIV and 97.6, 96, 95.4 and 98% for syphilis testing, respectively. In reference to TPHA, the test kit reported 4 false positives and 2 false negative results for syphilis. The κ values were 0.99 for HIV testing and 0.94 for syphilis testing. CONCLUSIONS: The excellent performance of the SD BIOLINE HIV/syphilis Duo test to detect HIV as well as syphilis facilitates the integration of syphilis testing and treatment to the already established HIV prevention programme, ultimately contributing to the dual HIV and syphilis elimination goal.
format Online
Article
Text
id pubmed-4410125
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher BMJ Publishing Group
record_format MEDLINE/PubMed
spelling pubmed-44101252015-05-01 The diagnostic performance evaluation of the SD BIOLINE HIV/syphilis Duo rapid test in southern Ethiopia: a cross-sectional study Shimelis, Techalew Tadesse, Endale BMJ Open Diagnostics OBJECTIVE: To determine the diagnostic performance of the SD BIOLINE HIV/syphilis Duo rapid test. DESIGN: A hospital-based cross-sectional study. SETTING: This evaluation was conducted at one of the largest hospitals in southern Ethiopia. PARTICIPANTS: Serum samples obtained from clients attending the antiretroviral therapy and voluntary counselling and testing centres were used. Sera were originally collected for the purpose of investigating syphilis epidemiology. The performance of the test to detect HIV was evaluated using 400 sera (200 HIV positives and 200 HIV negatives). Also, its performance to detect syphilis was evaluated using 85 syphilis positive and 100 syphilis negative serum samples. Individuals <15 years of age or syphilis treated or those with ≤50 cells/µL CD4 cell count were originally excluded. OUTCOME MEASURES: HIV screening was carried out according to the national rapid diagnostic testing (RDT) algorithm: Shenghai Kehua Bioengineering (KHB) test kit as a screening test, followed by the HIV1/2 STAT-PAK assay if positive. Where the result of the STAT-PAK is discordant with KHB, Unigold HIV is used as a tiebreaker to determine the result. We also used ELISA to resolve discordant HIV results. Syphilis serostatus was determined using the Treponema pallidum haemagglutination assay (TPHA). RESULTS: The respective sensitivity, specificity, positive predictive value and negative predictive value of the SD BIOLINE HIV/syphilis Duo test were 100, 99.5, 99.5 and 100% for HIV and 97.6, 96, 95.4 and 98% for syphilis testing, respectively. In reference to TPHA, the test kit reported 4 false positives and 2 false negative results for syphilis. The κ values were 0.99 for HIV testing and 0.94 for syphilis testing. CONCLUSIONS: The excellent performance of the SD BIOLINE HIV/syphilis Duo test to detect HIV as well as syphilis facilitates the integration of syphilis testing and treatment to the already established HIV prevention programme, ultimately contributing to the dual HIV and syphilis elimination goal. BMJ Publishing Group 2015-04-23 /pmc/articles/PMC4410125/ /pubmed/25908677 http://dx.doi.org/10.1136/bmjopen-2014-007371 Text en Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://group.bmj.com/group/rights-licensing/permissions This is an Open Access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/
spellingShingle Diagnostics
Shimelis, Techalew
Tadesse, Endale
The diagnostic performance evaluation of the SD BIOLINE HIV/syphilis Duo rapid test in southern Ethiopia: a cross-sectional study
title The diagnostic performance evaluation of the SD BIOLINE HIV/syphilis Duo rapid test in southern Ethiopia: a cross-sectional study
title_full The diagnostic performance evaluation of the SD BIOLINE HIV/syphilis Duo rapid test in southern Ethiopia: a cross-sectional study
title_fullStr The diagnostic performance evaluation of the SD BIOLINE HIV/syphilis Duo rapid test in southern Ethiopia: a cross-sectional study
title_full_unstemmed The diagnostic performance evaluation of the SD BIOLINE HIV/syphilis Duo rapid test in southern Ethiopia: a cross-sectional study
title_short The diagnostic performance evaluation of the SD BIOLINE HIV/syphilis Duo rapid test in southern Ethiopia: a cross-sectional study
title_sort diagnostic performance evaluation of the sd bioline hiv/syphilis duo rapid test in southern ethiopia: a cross-sectional study
topic Diagnostics
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4410125/
https://www.ncbi.nlm.nih.gov/pubmed/25908677
http://dx.doi.org/10.1136/bmjopen-2014-007371
work_keys_str_mv AT shimelistechalew thediagnosticperformanceevaluationofthesdbiolinehivsyphilisduorapidtestinsouthernethiopiaacrosssectionalstudy
AT tadesseendale thediagnosticperformanceevaluationofthesdbiolinehivsyphilisduorapidtestinsouthernethiopiaacrosssectionalstudy
AT shimelistechalew diagnosticperformanceevaluationofthesdbiolinehivsyphilisduorapidtestinsouthernethiopiaacrosssectionalstudy
AT tadesseendale diagnosticperformanceevaluationofthesdbiolinehivsyphilisduorapidtestinsouthernethiopiaacrosssectionalstudy