An in vitro comparison of tracheostomy tube cuffs
INTRODUCTION: The Shiley™ Flexible adult tracheostomy tube with TaperGuard™ cuff has been designed through its geometry, materials, diameter, and wall thickness to minimize micro-aspiration of fluids past the cuff and to provide an effective air seal in the trachea while also minimizing the risk of...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Dove Medical Press
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4411023/ https://www.ncbi.nlm.nih.gov/pubmed/25960679 http://dx.doi.org/10.2147/MDER.S76960 |
_version_ | 1782368403900596224 |
---|---|
author | Maguire, Seamus Haury, Frances Jew, Korinne |
author_facet | Maguire, Seamus Haury, Frances Jew, Korinne |
author_sort | Maguire, Seamus |
collection | PubMed |
description | INTRODUCTION: The Shiley™ Flexible adult tracheostomy tube with TaperGuard™ cuff has been designed through its geometry, materials, diameter, and wall thickness to minimize micro-aspiration of fluids past the cuff and to provide an effective air seal in the trachea while also minimizing the risk of excessive contact pressure on the tracheal mucosa. The cuff also has a deflated profile that may allow for easier insertion through the stoma site. This unique design is known as the TaperGuard™ cuff. The purpose of the observational, in vitro study reported here was to compare the TaperGuard™ taper-shaped cuff to a conventional high-volume low-pressure cylindrical-shaped cuff (Shiley™ Disposable Inner Cannula Tracheostomy Tube [DCT]) with respect to applied tracheal wall pressure, air and fluid sealing efficacy, and insertion force. METHODS: Three sizes of tracheostomy tubes with the two cuff types were placed in appropriately sized tracheal models and lateral wall pressure was measured via pressure-sensing elements on the inner surface. Fluid sealing performance was assessed by inflating the cuffs within the tracheal models (25 cmH(2)O), instilling water above the cuff, and measuring fluid leakage past the cuff. To measure air leak, tubes were attached to a test lung and ventilator, and leak was calculated by subtracting the average exhaled tidal volume from the average delivered tidal volume. A tensile test machine was used to measure insertion force for each tube with the cuff deflated to simulate clinical insertion through a stoma site. RESULTS: The average pressure exerted on the lateral wall of the model trachea was lower for the taper-shaped cuff than for the cylindrical cuff under all test conditions (P<0.05). The taper-shaped cuff also demonstrated a more even, lower pressure distribution along the lateral wall of the model trachea. The average air and fluid seal performance with the taper-shaped cuff was significantly improved, when compared to the cylindrical-shaped cuff, for each tube size tested (P<0.05). The insertion force for the taper-shaped cuff was ~40% less than that for the cylindrical-shaped cuff. CONCLUSION: In a model trachea, the Shiley™ Flexible Adult tracheostomy tube with TaperGuard™ cuff, when compared to the Shiley™ Disposable Inner Cannula Tracheostomy tube with cylindrical cuff, exerted a lower average lateral wall pressure and a more evenly distributed pressure. In addition, it provided more effective fluid and air seals and required less force to insert. |
format | Online Article Text |
id | pubmed-4411023 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | Dove Medical Press |
record_format | MEDLINE/PubMed |
spelling | pubmed-44110232015-05-08 An in vitro comparison of tracheostomy tube cuffs Maguire, Seamus Haury, Frances Jew, Korinne Med Devices (Auckl) Original Research INTRODUCTION: The Shiley™ Flexible adult tracheostomy tube with TaperGuard™ cuff has been designed through its geometry, materials, diameter, and wall thickness to minimize micro-aspiration of fluids past the cuff and to provide an effective air seal in the trachea while also minimizing the risk of excessive contact pressure on the tracheal mucosa. The cuff also has a deflated profile that may allow for easier insertion through the stoma site. This unique design is known as the TaperGuard™ cuff. The purpose of the observational, in vitro study reported here was to compare the TaperGuard™ taper-shaped cuff to a conventional high-volume low-pressure cylindrical-shaped cuff (Shiley™ Disposable Inner Cannula Tracheostomy Tube [DCT]) with respect to applied tracheal wall pressure, air and fluid sealing efficacy, and insertion force. METHODS: Three sizes of tracheostomy tubes with the two cuff types were placed in appropriately sized tracheal models and lateral wall pressure was measured via pressure-sensing elements on the inner surface. Fluid sealing performance was assessed by inflating the cuffs within the tracheal models (25 cmH(2)O), instilling water above the cuff, and measuring fluid leakage past the cuff. To measure air leak, tubes were attached to a test lung and ventilator, and leak was calculated by subtracting the average exhaled tidal volume from the average delivered tidal volume. A tensile test machine was used to measure insertion force for each tube with the cuff deflated to simulate clinical insertion through a stoma site. RESULTS: The average pressure exerted on the lateral wall of the model trachea was lower for the taper-shaped cuff than for the cylindrical cuff under all test conditions (P<0.05). The taper-shaped cuff also demonstrated a more even, lower pressure distribution along the lateral wall of the model trachea. The average air and fluid seal performance with the taper-shaped cuff was significantly improved, when compared to the cylindrical-shaped cuff, for each tube size tested (P<0.05). The insertion force for the taper-shaped cuff was ~40% less than that for the cylindrical-shaped cuff. CONCLUSION: In a model trachea, the Shiley™ Flexible Adult tracheostomy tube with TaperGuard™ cuff, when compared to the Shiley™ Disposable Inner Cannula Tracheostomy tube with cylindrical cuff, exerted a lower average lateral wall pressure and a more evenly distributed pressure. In addition, it provided more effective fluid and air seals and required less force to insert. Dove Medical Press 2015-04-21 /pmc/articles/PMC4411023/ /pubmed/25960679 http://dx.doi.org/10.2147/MDER.S76960 Text en © 2015 Maguire et al. This work is published by Dove Medical Press Limited, and licensed under Creative Commons Attribution – Non Commercial (unported, v3.0) License The full terms of the License are available at http://creativecommons.org/licenses/by-nc/3.0/. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. |
spellingShingle | Original Research Maguire, Seamus Haury, Frances Jew, Korinne An in vitro comparison of tracheostomy tube cuffs |
title | An in vitro comparison of tracheostomy tube cuffs |
title_full | An in vitro comparison of tracheostomy tube cuffs |
title_fullStr | An in vitro comparison of tracheostomy tube cuffs |
title_full_unstemmed | An in vitro comparison of tracheostomy tube cuffs |
title_short | An in vitro comparison of tracheostomy tube cuffs |
title_sort | in vitro comparison of tracheostomy tube cuffs |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4411023/ https://www.ncbi.nlm.nih.gov/pubmed/25960679 http://dx.doi.org/10.2147/MDER.S76960 |
work_keys_str_mv | AT maguireseamus aninvitrocomparisonoftracheostomytubecuffs AT hauryfrances aninvitrocomparisonoftracheostomytubecuffs AT jewkorinne aninvitrocomparisonoftracheostomytubecuffs AT maguireseamus invitrocomparisonoftracheostomytubecuffs AT hauryfrances invitrocomparisonoftracheostomytubecuffs AT jewkorinne invitrocomparisonoftracheostomytubecuffs |