Performance and user evaluation of a novel capacitance-based automatic urinometer compared with a manual standard urinometer after elective cardiac surgery

INTRODUCTION: In the intensive care setting, most physiologic parameters are monitored automatically. However, urine output (UO) is still monitored hourly by manually handled urinometers. In this study, we evaluated an automatic urinometer (AU) and compared it with a manual urinometer (MU). METHODS:...

Descripción completa

Detalles Bibliográficos
Autores principales: Eklund, Anton, Slettengren, Martin, van der Linden, Jan
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4416393/
https://www.ncbi.nlm.nih.gov/pubmed/25895503
http://dx.doi.org/10.1186/s13054-015-0899-4
_version_ 1782369234062409728
author Eklund, Anton
Slettengren, Martin
van der Linden, Jan
author_facet Eklund, Anton
Slettengren, Martin
van der Linden, Jan
author_sort Eklund, Anton
collection PubMed
description INTRODUCTION: In the intensive care setting, most physiologic parameters are monitored automatically. However, urine output (UO) is still monitored hourly by manually handled urinometers. In this study, we evaluated an automatic urinometer (AU) and compared it with a manual urinometer (MU). METHODS: This prospective study was carried out in the intensive care unit of a cardiothoracic surgical clinic. In postoperative patients (n = 34) with indwelling urinary catheters and an expected stay of 24 hours or more, hourly UO samples were measured with an AU (Sippi, n = 220; Observe Medical, Gothenburg, Sweden) or an MU (UnoMeter™ 500, n = 188; Unomedical, Birkerød, Denmark) and thereafter validated by cylinder measurements. Malposition of the instrument at the time of reading excluded measurement. Data were analyzed with the Bland-Altman method. The performance of the MU was used as the minimum criterion of acceptance when the AU was evaluated. The loss of precision with the MU due to temporal deviation from fixed hourly measurements was recorded (n = 108). A questionnaire filled out by the ward staff (n = 28) was used to evaluate the ease of use of the AU compared with the MU. RESULTS: Bland-Altman analysis showed a smaller mean bias for the AU (+1.9 ml) compared with the MU (+5.3 ml) (P <0.0001). There was no statistical difference in measurement precision between the two urinometers, as defined by their limits of agreement (±15.2 ml vs. ±16.6 ml, P = 0.11). The mean temporal variation with the MU was ±7.4 minutes (±12.4%), and the limits of agreement were ±23.9 minutes (±39.8%), compared with no temporal variation with the AU (P <0.0001). The ward staff considered the AU easy to learn to use and rated it higher than the MU (P <0.0001). CONCLUSIONS: The AU was not inferior to the MU and was significantly better in terms of bias, temporal deviation and staff opinion, although the clinical relevance of these findings may be open to discussion. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1186/s13054-015-0899-4) contains supplementary material, which is available to authorized users.
format Online
Article
Text
id pubmed-4416393
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-44163932015-05-02 Performance and user evaluation of a novel capacitance-based automatic urinometer compared with a manual standard urinometer after elective cardiac surgery Eklund, Anton Slettengren, Martin van der Linden, Jan Crit Care Research INTRODUCTION: In the intensive care setting, most physiologic parameters are monitored automatically. However, urine output (UO) is still monitored hourly by manually handled urinometers. In this study, we evaluated an automatic urinometer (AU) and compared it with a manual urinometer (MU). METHODS: This prospective study was carried out in the intensive care unit of a cardiothoracic surgical clinic. In postoperative patients (n = 34) with indwelling urinary catheters and an expected stay of 24 hours or more, hourly UO samples were measured with an AU (Sippi, n = 220; Observe Medical, Gothenburg, Sweden) or an MU (UnoMeter™ 500, n = 188; Unomedical, Birkerød, Denmark) and thereafter validated by cylinder measurements. Malposition of the instrument at the time of reading excluded measurement. Data were analyzed with the Bland-Altman method. The performance of the MU was used as the minimum criterion of acceptance when the AU was evaluated. The loss of precision with the MU due to temporal deviation from fixed hourly measurements was recorded (n = 108). A questionnaire filled out by the ward staff (n = 28) was used to evaluate the ease of use of the AU compared with the MU. RESULTS: Bland-Altman analysis showed a smaller mean bias for the AU (+1.9 ml) compared with the MU (+5.3 ml) (P <0.0001). There was no statistical difference in measurement precision between the two urinometers, as defined by their limits of agreement (±15.2 ml vs. ±16.6 ml, P = 0.11). The mean temporal variation with the MU was ±7.4 minutes (±12.4%), and the limits of agreement were ±23.9 minutes (±39.8%), compared with no temporal variation with the AU (P <0.0001). The ward staff considered the AU easy to learn to use and rated it higher than the MU (P <0.0001). CONCLUSIONS: The AU was not inferior to the MU and was significantly better in terms of bias, temporal deviation and staff opinion, although the clinical relevance of these findings may be open to discussion. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1186/s13054-015-0899-4) contains supplementary material, which is available to authorized users. BioMed Central 2015-04-21 2015 /pmc/articles/PMC4416393/ /pubmed/25895503 http://dx.doi.org/10.1186/s13054-015-0899-4 Text en © Eklund et al.; licensee BioMed Central. 2015 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research
Eklund, Anton
Slettengren, Martin
van der Linden, Jan
Performance and user evaluation of a novel capacitance-based automatic urinometer compared with a manual standard urinometer after elective cardiac surgery
title Performance and user evaluation of a novel capacitance-based automatic urinometer compared with a manual standard urinometer after elective cardiac surgery
title_full Performance and user evaluation of a novel capacitance-based automatic urinometer compared with a manual standard urinometer after elective cardiac surgery
title_fullStr Performance and user evaluation of a novel capacitance-based automatic urinometer compared with a manual standard urinometer after elective cardiac surgery
title_full_unstemmed Performance and user evaluation of a novel capacitance-based automatic urinometer compared with a manual standard urinometer after elective cardiac surgery
title_short Performance and user evaluation of a novel capacitance-based automatic urinometer compared with a manual standard urinometer after elective cardiac surgery
title_sort performance and user evaluation of a novel capacitance-based automatic urinometer compared with a manual standard urinometer after elective cardiac surgery
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4416393/
https://www.ncbi.nlm.nih.gov/pubmed/25895503
http://dx.doi.org/10.1186/s13054-015-0899-4
work_keys_str_mv AT eklundanton performanceanduserevaluationofanovelcapacitancebasedautomaticurinometercomparedwithamanualstandardurinometerafterelectivecardiacsurgery
AT slettengrenmartin performanceanduserevaluationofanovelcapacitancebasedautomaticurinometercomparedwithamanualstandardurinometerafterelectivecardiacsurgery
AT vanderlindenjan performanceanduserevaluationofanovelcapacitancebasedautomaticurinometercomparedwithamanualstandardurinometerafterelectivecardiacsurgery