Cargando…

Effect of Gold Nanoparticles on Prostate Dose Distribution under Ir-192 Internal and 18 MV External Radiotherapy Procedures Using Gel Dosimetry and Monte Carlo Method

BACKGROUND: Gel polymers are considered as new dosimeters for determining radiotherapy dose distribution in three dimensions. OBJECTIVE: The ability of a new formulation of MAGIC-f polymer gel was assessed by experimental measurement and Monte Carlo (MC) method for studying the effect of gold nanopa...

Descripción completa

Detalles Bibliográficos
Autores principales: Khosravi, H., Hashemi, B., Mahdavi, S. R., Hejazi, P.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Shiraz University of Medical Sciences 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4417617/
https://www.ncbi.nlm.nih.gov/pubmed/25973406
_version_ 1782369389881851904
author Khosravi, H.
Hashemi, B.
Mahdavi, S. R.
Hejazi, P.
author_facet Khosravi, H.
Hashemi, B.
Mahdavi, S. R.
Hejazi, P.
author_sort Khosravi, H.
collection PubMed
description BACKGROUND: Gel polymers are considered as new dosimeters for determining radiotherapy dose distribution in three dimensions. OBJECTIVE: The ability of a new formulation of MAGIC-f polymer gel was assessed by experimental measurement and Monte Carlo (MC) method for studying the effect of gold nanoparticles (GNPs) in prostate dose distributions under the internal Ir-192 and external 18MV radiotherapy practices. METHOD: A Plexiglas phantom was made representing human pelvis. The GNP shaving 15 nm in diameter and 0.1 mM concentration were synthesized using chemical reduction method. Then, a new formulation of MAGIC-f gel was synthesized. The fabricated gel was poured in the tubes located at the prostate (with and without the GNPs) and bladder locations of the phantom. The phantom was irradiated to an Ir-192 source and 18 MV beam of a Varian linac separately based on common radiotherapy procedures used for prostate cancer. After 24 hours, the irradiated gels were read using a Siemens 1.5 Tesla MRI scanner. The absolute doses at the reference points and isodose curves resulted from the experimental measurement of the gels and MC simulations following the internal and external radiotherapy practices were compared. RESULTS: The mean absorbed doses measured with the gel in the presence of the GNPs in prostate were 15% and 8 % higher than the corresponding values without the GNPs under the internal and external radiation therapies, respectively. MC simulations also indicated a dose increase of 14 % and 7 % due to presence of the GNPs, for the same experimental internal and external radiotherapy practices, respectively. CONCLUSION: There was a good agreement between the dose enhancement factors (DEFs) estimated with MC simulations and experiment gel measurements due to the GNPs. The results indicated that the polymer gel dosimetry method as developed and used in this study, can be recommended as a reliable method for investigating the DEF of GNPs in internal and external radiotherapy practices. 
format Online
Article
Text
id pubmed-4417617
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Shiraz University of Medical Sciences
record_format MEDLINE/PubMed
spelling pubmed-44176172015-05-13 Effect of Gold Nanoparticles on Prostate Dose Distribution under Ir-192 Internal and 18 MV External Radiotherapy Procedures Using Gel Dosimetry and Monte Carlo Method Khosravi, H. Hashemi, B. Mahdavi, S. R. Hejazi, P. J Biomed Phys Eng Original Article BACKGROUND: Gel polymers are considered as new dosimeters for determining radiotherapy dose distribution in three dimensions. OBJECTIVE: The ability of a new formulation of MAGIC-f polymer gel was assessed by experimental measurement and Monte Carlo (MC) method for studying the effect of gold nanoparticles (GNPs) in prostate dose distributions under the internal Ir-192 and external 18MV radiotherapy practices. METHOD: A Plexiglas phantom was made representing human pelvis. The GNP shaving 15 nm in diameter and 0.1 mM concentration were synthesized using chemical reduction method. Then, a new formulation of MAGIC-f gel was synthesized. The fabricated gel was poured in the tubes located at the prostate (with and without the GNPs) and bladder locations of the phantom. The phantom was irradiated to an Ir-192 source and 18 MV beam of a Varian linac separately based on common radiotherapy procedures used for prostate cancer. After 24 hours, the irradiated gels were read using a Siemens 1.5 Tesla MRI scanner. The absolute doses at the reference points and isodose curves resulted from the experimental measurement of the gels and MC simulations following the internal and external radiotherapy practices were compared. RESULTS: The mean absorbed doses measured with the gel in the presence of the GNPs in prostate were 15% and 8 % higher than the corresponding values without the GNPs under the internal and external radiation therapies, respectively. MC simulations also indicated a dose increase of 14 % and 7 % due to presence of the GNPs, for the same experimental internal and external radiotherapy practices, respectively. CONCLUSION: There was a good agreement between the dose enhancement factors (DEFs) estimated with MC simulations and experiment gel measurements due to the GNPs. The results indicated that the polymer gel dosimetry method as developed and used in this study, can be recommended as a reliable method for investigating the DEF of GNPs in internal and external radiotherapy practices.  Shiraz University of Medical Sciences 2015-03-04 /pmc/articles/PMC4417617/ /pubmed/25973406 Text en © 2015: Journal of Biomedical Physics and Engineering This is an Open Access article distributed under the terms of the Creative Commons Attribution-NonCommercial 3.0 Unported License (http://creativecommons.org/licenses/by-nc/3.0/deed.en_US), which permits unrestricted use, distribution and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Khosravi, H.
Hashemi, B.
Mahdavi, S. R.
Hejazi, P.
Effect of Gold Nanoparticles on Prostate Dose Distribution under Ir-192 Internal and 18 MV External Radiotherapy Procedures Using Gel Dosimetry and Monte Carlo Method
title Effect of Gold Nanoparticles on Prostate Dose Distribution under Ir-192 Internal and 18 MV External Radiotherapy Procedures Using Gel Dosimetry and Monte Carlo Method
title_full Effect of Gold Nanoparticles on Prostate Dose Distribution under Ir-192 Internal and 18 MV External Radiotherapy Procedures Using Gel Dosimetry and Monte Carlo Method
title_fullStr Effect of Gold Nanoparticles on Prostate Dose Distribution under Ir-192 Internal and 18 MV External Radiotherapy Procedures Using Gel Dosimetry and Monte Carlo Method
title_full_unstemmed Effect of Gold Nanoparticles on Prostate Dose Distribution under Ir-192 Internal and 18 MV External Radiotherapy Procedures Using Gel Dosimetry and Monte Carlo Method
title_short Effect of Gold Nanoparticles on Prostate Dose Distribution under Ir-192 Internal and 18 MV External Radiotherapy Procedures Using Gel Dosimetry and Monte Carlo Method
title_sort effect of gold nanoparticles on prostate dose distribution under ir-192 internal and 18 mv external radiotherapy procedures using gel dosimetry and monte carlo method
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4417617/
https://www.ncbi.nlm.nih.gov/pubmed/25973406
work_keys_str_mv AT khosravih effectofgoldnanoparticlesonprostatedosedistributionunderir192internaland18mvexternalradiotherapyproceduresusinggeldosimetryandmontecarlomethod
AT hashemib effectofgoldnanoparticlesonprostatedosedistributionunderir192internaland18mvexternalradiotherapyproceduresusinggeldosimetryandmontecarlomethod
AT mahdavisr effectofgoldnanoparticlesonprostatedosedistributionunderir192internaland18mvexternalradiotherapyproceduresusinggeldosimetryandmontecarlomethod
AT hejazip effectofgoldnanoparticlesonprostatedosedistributionunderir192internaland18mvexternalradiotherapyproceduresusinggeldosimetryandmontecarlomethod