Influence of Accelerometer Type and Placement on Physical Activity Energy Expenditure Prediction in Manual Wheelchair Users

PURPOSE: To assess the validity of two accelerometer devices, at two different anatomical locations, for the prediction of physical activity energy expenditure (PAEE) in manual wheelchair users (MWUs). METHODS: Seventeen MWUs (36 ± 10 yrs, 72 ± 11 kg) completed ten activities; resting, folding cloth...

Descripción completa

Detalles Bibliográficos
Autores principales: Nightingale, Tom Edward, Walhin, Jean-Philippe, Thompson, Dylan, Bilzon, James Lee John
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4425541/
https://www.ncbi.nlm.nih.gov/pubmed/25955304
http://dx.doi.org/10.1371/journal.pone.0126086
_version_ 1782370501626167296
author Nightingale, Tom Edward
Walhin, Jean-Philippe
Thompson, Dylan
Bilzon, James Lee John
author_facet Nightingale, Tom Edward
Walhin, Jean-Philippe
Thompson, Dylan
Bilzon, James Lee John
author_sort Nightingale, Tom Edward
collection PubMed
description PURPOSE: To assess the validity of two accelerometer devices, at two different anatomical locations, for the prediction of physical activity energy expenditure (PAEE) in manual wheelchair users (MWUs). METHODS: Seventeen MWUs (36 ± 10 yrs, 72 ± 11 kg) completed ten activities; resting, folding clothes, propulsion on a 1% gradient (3,4,5,6 and 7 km·hr(-1)) and propulsion at 4km·hr(-1 )(with an additional 8% body mass, 2% and 3% gradient) on a motorised wheelchair treadmill. GT3X+ and GENEActiv accelerometers were worn on the right wrist (W) and upper arm (UA). Linear regression analysis was conducted between outputs from each accelerometer and criterion PAEE, measured using indirect calorimetry. Subsequent error statistics were calculated for the derived regression equations for all four device/location combinations, using a leave-one-out cross-validation analysis. RESULTS: Accelerometer outputs at each anatomical location were significantly (p < .01) associated with PAEE (GT3X+-UA; r = 0.68 and GT3X+-W; r = 0.82. GENEActiv-UA; r = 0.87 and GENEActiv-W; r = 0.88). Mean ± SD PAEE estimation errors for all activities combined were 15 ± 45%, 14 ± 50%, 3 ± 25% and 4 ± 26% for GT3X+-UA, GT3X+-W, GENEActiv-UA and GENEActiv-W, respectively. Absolute PAEE estimation errors for devices varied, 19 to 66% for GT3X+-UA, 17 to 122% for GT3X+-W, 15 to 26% for GENEActiv-UA and from 17.0 to 32% for the GENEActiv-W. CONCLUSION: The results indicate that the GENEActiv device worn on either the upper arm or wrist provides the most valid prediction of PAEE in MWUs. Variation in error statistics between the two devices is a result of inherent differences in internal components, on-board filtering processes and outputs of each device.
format Online
Article
Text
id pubmed-4425541
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-44255412015-05-21 Influence of Accelerometer Type and Placement on Physical Activity Energy Expenditure Prediction in Manual Wheelchair Users Nightingale, Tom Edward Walhin, Jean-Philippe Thompson, Dylan Bilzon, James Lee John PLoS One Research Article PURPOSE: To assess the validity of two accelerometer devices, at two different anatomical locations, for the prediction of physical activity energy expenditure (PAEE) in manual wheelchair users (MWUs). METHODS: Seventeen MWUs (36 ± 10 yrs, 72 ± 11 kg) completed ten activities; resting, folding clothes, propulsion on a 1% gradient (3,4,5,6 and 7 km·hr(-1)) and propulsion at 4km·hr(-1 )(with an additional 8% body mass, 2% and 3% gradient) on a motorised wheelchair treadmill. GT3X+ and GENEActiv accelerometers were worn on the right wrist (W) and upper arm (UA). Linear regression analysis was conducted between outputs from each accelerometer and criterion PAEE, measured using indirect calorimetry. Subsequent error statistics were calculated for the derived regression equations for all four device/location combinations, using a leave-one-out cross-validation analysis. RESULTS: Accelerometer outputs at each anatomical location were significantly (p < .01) associated with PAEE (GT3X+-UA; r = 0.68 and GT3X+-W; r = 0.82. GENEActiv-UA; r = 0.87 and GENEActiv-W; r = 0.88). Mean ± SD PAEE estimation errors for all activities combined were 15 ± 45%, 14 ± 50%, 3 ± 25% and 4 ± 26% for GT3X+-UA, GT3X+-W, GENEActiv-UA and GENEActiv-W, respectively. Absolute PAEE estimation errors for devices varied, 19 to 66% for GT3X+-UA, 17 to 122% for GT3X+-W, 15 to 26% for GENEActiv-UA and from 17.0 to 32% for the GENEActiv-W. CONCLUSION: The results indicate that the GENEActiv device worn on either the upper arm or wrist provides the most valid prediction of PAEE in MWUs. Variation in error statistics between the two devices is a result of inherent differences in internal components, on-board filtering processes and outputs of each device. Public Library of Science 2015-05-08 /pmc/articles/PMC4425541/ /pubmed/25955304 http://dx.doi.org/10.1371/journal.pone.0126086 Text en © 2015 Nightingale et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited.
spellingShingle Research Article
Nightingale, Tom Edward
Walhin, Jean-Philippe
Thompson, Dylan
Bilzon, James Lee John
Influence of Accelerometer Type and Placement on Physical Activity Energy Expenditure Prediction in Manual Wheelchair Users
title Influence of Accelerometer Type and Placement on Physical Activity Energy Expenditure Prediction in Manual Wheelchair Users
title_full Influence of Accelerometer Type and Placement on Physical Activity Energy Expenditure Prediction in Manual Wheelchair Users
title_fullStr Influence of Accelerometer Type and Placement on Physical Activity Energy Expenditure Prediction in Manual Wheelchair Users
title_full_unstemmed Influence of Accelerometer Type and Placement on Physical Activity Energy Expenditure Prediction in Manual Wheelchair Users
title_short Influence of Accelerometer Type and Placement on Physical Activity Energy Expenditure Prediction in Manual Wheelchair Users
title_sort influence of accelerometer type and placement on physical activity energy expenditure prediction in manual wheelchair users
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4425541/
https://www.ncbi.nlm.nih.gov/pubmed/25955304
http://dx.doi.org/10.1371/journal.pone.0126086
work_keys_str_mv AT nightingaletomedward influenceofaccelerometertypeandplacementonphysicalactivityenergyexpenditurepredictioninmanualwheelchairusers
AT walhinjeanphilippe influenceofaccelerometertypeandplacementonphysicalactivityenergyexpenditurepredictioninmanualwheelchairusers
AT thompsondylan influenceofaccelerometertypeandplacementonphysicalactivityenergyexpenditurepredictioninmanualwheelchairusers
AT bilzonjamesleejohn influenceofaccelerometertypeandplacementonphysicalactivityenergyexpenditurepredictioninmanualwheelchairusers