Cargando…

Detection of Cerebral Hemorrhage in Rabbits by Time-Difference Magnetic Inductive Phase Shift Spectroscopy

Cerebral hemorrhage, a difficult issue in clinical practice, is often detected and studied with computed tomography (CT), magnetic resonance imaging (MRI), and positron emission tomography (PET). However, these expensive devices are not readily available in economically underdeveloped regions, and h...

Descripción completa

Detalles Bibliográficos
Autores principales: Pan, Wencai, Yan, Qingguang, Qin, Mingxin, Jin, Gui, Sun, Jian, Ning, Xu, Zhuang, Wei, Peng, Bin, Li, Gen
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4441421/
https://www.ncbi.nlm.nih.gov/pubmed/26001112
http://dx.doi.org/10.1371/journal.pone.0128127
_version_ 1782372786806718464
author Pan, Wencai
Yan, Qingguang
Qin, Mingxin
Jin, Gui
Sun, Jian
Ning, Xu
Zhuang, Wei
Peng, Bin
Li, Gen
author_facet Pan, Wencai
Yan, Qingguang
Qin, Mingxin
Jin, Gui
Sun, Jian
Ning, Xu
Zhuang, Wei
Peng, Bin
Li, Gen
author_sort Pan, Wencai
collection PubMed
description Cerebral hemorrhage, a difficult issue in clinical practice, is often detected and studied with computed tomography (CT), magnetic resonance imaging (MRI), and positron emission tomography (PET). However, these expensive devices are not readily available in economically underdeveloped regions, and hence are unable to provide bedside and emergency on-site monitoring. The magnetic inductive phase shift (MIPS) is an emerging technology that may become a new tool to detect cerebral hemorrhage and to serve as an inexpensive partial substitute to medical imaging. In order to study a wider band of cerebral hemorrhage MIPS and to provide more useful information for measuring cerebral hemorrhage, we established a cerebral hemorrhage magnetic induction phase shift spectroscopy (MIPSS) detection system. Thirteen rabbits with five cerebral hemorrhage states were studied using a single coil-coil within a 1 MHz-200 MHz frequency range in linear sweep. A feature band (FB) with the highest detection sensitivity and the greatest stability was selected for further analysis and processing. In addition, a maximum conductivity cerebrospinal fluid (CSF) MRI was performed to verify and interpret the MIPSS result. The average phase shift change induced by a 3 ml injection of autologous blood under FB was -7.7503° ± 1.4204°, which was considerably larger than our previous work. Data analysis with a non-parametric statistical Friedman M test showed that in the FB, MIPSS could distinguish the five states of cerebral hemorrhage in rabbits, with a statistical significance of p<0.05. A B-F distribution profile was designed according to the MIPSS under FB that can provide instantaneous diagnostic information about the cerebral hemorrhage severity from a single set of measurements. The results illustrate that the MIPSS detection method is able to provide a new possibility for real-time monitoring and diagnosis of the severity of cerebral hemorrhage.
format Online
Article
Text
id pubmed-4441421
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-44414212015-05-28 Detection of Cerebral Hemorrhage in Rabbits by Time-Difference Magnetic Inductive Phase Shift Spectroscopy Pan, Wencai Yan, Qingguang Qin, Mingxin Jin, Gui Sun, Jian Ning, Xu Zhuang, Wei Peng, Bin Li, Gen PLoS One Research Article Cerebral hemorrhage, a difficult issue in clinical practice, is often detected and studied with computed tomography (CT), magnetic resonance imaging (MRI), and positron emission tomography (PET). However, these expensive devices are not readily available in economically underdeveloped regions, and hence are unable to provide bedside and emergency on-site monitoring. The magnetic inductive phase shift (MIPS) is an emerging technology that may become a new tool to detect cerebral hemorrhage and to serve as an inexpensive partial substitute to medical imaging. In order to study a wider band of cerebral hemorrhage MIPS and to provide more useful information for measuring cerebral hemorrhage, we established a cerebral hemorrhage magnetic induction phase shift spectroscopy (MIPSS) detection system. Thirteen rabbits with five cerebral hemorrhage states were studied using a single coil-coil within a 1 MHz-200 MHz frequency range in linear sweep. A feature band (FB) with the highest detection sensitivity and the greatest stability was selected for further analysis and processing. In addition, a maximum conductivity cerebrospinal fluid (CSF) MRI was performed to verify and interpret the MIPSS result. The average phase shift change induced by a 3 ml injection of autologous blood under FB was -7.7503° ± 1.4204°, which was considerably larger than our previous work. Data analysis with a non-parametric statistical Friedman M test showed that in the FB, MIPSS could distinguish the five states of cerebral hemorrhage in rabbits, with a statistical significance of p<0.05. A B-F distribution profile was designed according to the MIPSS under FB that can provide instantaneous diagnostic information about the cerebral hemorrhage severity from a single set of measurements. The results illustrate that the MIPSS detection method is able to provide a new possibility for real-time monitoring and diagnosis of the severity of cerebral hemorrhage. Public Library of Science 2015-05-22 /pmc/articles/PMC4441421/ /pubmed/26001112 http://dx.doi.org/10.1371/journal.pone.0128127 Text en © 2015 Pan et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited.
spellingShingle Research Article
Pan, Wencai
Yan, Qingguang
Qin, Mingxin
Jin, Gui
Sun, Jian
Ning, Xu
Zhuang, Wei
Peng, Bin
Li, Gen
Detection of Cerebral Hemorrhage in Rabbits by Time-Difference Magnetic Inductive Phase Shift Spectroscopy
title Detection of Cerebral Hemorrhage in Rabbits by Time-Difference Magnetic Inductive Phase Shift Spectroscopy
title_full Detection of Cerebral Hemorrhage in Rabbits by Time-Difference Magnetic Inductive Phase Shift Spectroscopy
title_fullStr Detection of Cerebral Hemorrhage in Rabbits by Time-Difference Magnetic Inductive Phase Shift Spectroscopy
title_full_unstemmed Detection of Cerebral Hemorrhage in Rabbits by Time-Difference Magnetic Inductive Phase Shift Spectroscopy
title_short Detection of Cerebral Hemorrhage in Rabbits by Time-Difference Magnetic Inductive Phase Shift Spectroscopy
title_sort detection of cerebral hemorrhage in rabbits by time-difference magnetic inductive phase shift spectroscopy
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4441421/
https://www.ncbi.nlm.nih.gov/pubmed/26001112
http://dx.doi.org/10.1371/journal.pone.0128127
work_keys_str_mv AT panwencai detectionofcerebralhemorrhageinrabbitsbytimedifferencemagneticinductivephaseshiftspectroscopy
AT yanqingguang detectionofcerebralhemorrhageinrabbitsbytimedifferencemagneticinductivephaseshiftspectroscopy
AT qinmingxin detectionofcerebralhemorrhageinrabbitsbytimedifferencemagneticinductivephaseshiftspectroscopy
AT jingui detectionofcerebralhemorrhageinrabbitsbytimedifferencemagneticinductivephaseshiftspectroscopy
AT sunjian detectionofcerebralhemorrhageinrabbitsbytimedifferencemagneticinductivephaseshiftspectroscopy
AT ningxu detectionofcerebralhemorrhageinrabbitsbytimedifferencemagneticinductivephaseshiftspectroscopy
AT zhuangwei detectionofcerebralhemorrhageinrabbitsbytimedifferencemagneticinductivephaseshiftspectroscopy
AT pengbin detectionofcerebralhemorrhageinrabbitsbytimedifferencemagneticinductivephaseshiftspectroscopy
AT ligen detectionofcerebralhemorrhageinrabbitsbytimedifferencemagneticinductivephaseshiftspectroscopy