Assessment of the feasibility of an ultra-low power, wireless digital patch for the continuous ambulatory monitoring of vital signs
BACKGROUND AND OBJECTIVES: Vital signs are usually recorded at 4–8 h intervals in hospital patients, and deterioration between measurements can have serious consequences. The primary study objective was to assess agreement between a new ultra-low power, wireless and wearable surveillance system for...
Autores principales: | , , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BMJ Publishing Group
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4442176/ https://www.ncbi.nlm.nih.gov/pubmed/25991447 http://dx.doi.org/10.1136/bmjopen-2014-006606 |
_version_ | 1782372867724279808 |
---|---|
author | Hernandez-Silveira, Miguel Ahmed, Kamran Ang, Su-Shin Zandari, Fahriya Mehta, Tinaz Weir, Rebecca Burdett, Alison Toumazou, Chris Brett, Stephen J |
author_facet | Hernandez-Silveira, Miguel Ahmed, Kamran Ang, Su-Shin Zandari, Fahriya Mehta, Tinaz Weir, Rebecca Burdett, Alison Toumazou, Chris Brett, Stephen J |
author_sort | Hernandez-Silveira, Miguel |
collection | PubMed |
description | BACKGROUND AND OBJECTIVES: Vital signs are usually recorded at 4–8 h intervals in hospital patients, and deterioration between measurements can have serious consequences. The primary study objective was to assess agreement between a new ultra-low power, wireless and wearable surveillance system for continuous ambulatory monitoring of vital signs and a widely used clinical vital signs monitor. The secondary objective was to examine the system's ability to automatically identify and reject invalid physiological data. SETTING: Single hospital centre. PARTICIPANTS: Heart and respiratory rate were recorded over 2 h in 20 patients undergoing elective surgery and a second group of 41 patients with comorbid conditions, in the general ward. OUTCOME MEASURES: Primary outcome measures were limits of agreement and bias. The secondary outcome measure was proportion of data rejected. RESULTS: The digital patch provided reliable heart rate values in the majority of patients (about 80%) with normal sinus rhythm, and in the presence of abnormal ECG recordings (excluding aperiodic arrhythmias such as atrial fibrillation). The mean difference between systems was less than ±1 bpm in all patient groups studied. Although respiratory data were more frequently rejected as invalid because of the high sensitivity of impedance pneumography to motion artefacts, valid rates were reported for 50% of recordings with a mean difference of less than ±1 brpm compared with the bedside monitor. Correlation between systems was statistically significant (p<0.0001) for heart and respiratory rate, apart from respiratory rate in patients with atrial fibrillation (p=0.02). CONCLUSIONS: Overall agreement between digital patch and clinical monitor was satisfactory, as was the efficacy of the system for automatic rejection of invalid data. Wireless monitoring technologies, such as the one tested, may offer clinical value when implemented as part of wider hospital systems that integrate and support existing clinical protocols and workflows. |
format | Online Article Text |
id | pubmed-4442176 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | BMJ Publishing Group |
record_format | MEDLINE/PubMed |
spelling | pubmed-44421762015-05-28 Assessment of the feasibility of an ultra-low power, wireless digital patch for the continuous ambulatory monitoring of vital signs Hernandez-Silveira, Miguel Ahmed, Kamran Ang, Su-Shin Zandari, Fahriya Mehta, Tinaz Weir, Rebecca Burdett, Alison Toumazou, Chris Brett, Stephen J BMJ Open Medical Management BACKGROUND AND OBJECTIVES: Vital signs are usually recorded at 4–8 h intervals in hospital patients, and deterioration between measurements can have serious consequences. The primary study objective was to assess agreement between a new ultra-low power, wireless and wearable surveillance system for continuous ambulatory monitoring of vital signs and a widely used clinical vital signs monitor. The secondary objective was to examine the system's ability to automatically identify and reject invalid physiological data. SETTING: Single hospital centre. PARTICIPANTS: Heart and respiratory rate were recorded over 2 h in 20 patients undergoing elective surgery and a second group of 41 patients with comorbid conditions, in the general ward. OUTCOME MEASURES: Primary outcome measures were limits of agreement and bias. The secondary outcome measure was proportion of data rejected. RESULTS: The digital patch provided reliable heart rate values in the majority of patients (about 80%) with normal sinus rhythm, and in the presence of abnormal ECG recordings (excluding aperiodic arrhythmias such as atrial fibrillation). The mean difference between systems was less than ±1 bpm in all patient groups studied. Although respiratory data were more frequently rejected as invalid because of the high sensitivity of impedance pneumography to motion artefacts, valid rates were reported for 50% of recordings with a mean difference of less than ±1 brpm compared with the bedside monitor. Correlation between systems was statistically significant (p<0.0001) for heart and respiratory rate, apart from respiratory rate in patients with atrial fibrillation (p=0.02). CONCLUSIONS: Overall agreement between digital patch and clinical monitor was satisfactory, as was the efficacy of the system for automatic rejection of invalid data. Wireless monitoring technologies, such as the one tested, may offer clinical value when implemented as part of wider hospital systems that integrate and support existing clinical protocols and workflows. BMJ Publishing Group 2015-05-19 /pmc/articles/PMC4442176/ /pubmed/25991447 http://dx.doi.org/10.1136/bmjopen-2014-006606 Text en Published by the BMJ Publishing Group Limited. For permission to use (where not already granted under a licence) please go to http://group.bmj.com/group/rights-licensing/permissions This is an Open Access article distributed in accordance with the Creative Commons Attribution Non Commercial (CC BY-NC 4.0) license, which permits others to distribute, remix, adapt, build upon this work non-commercially, and license their derivative works on different terms, provided the original work is properly cited and the use is non-commercial. See: http://creativecommons.org/licenses/by-nc/4.0/ |
spellingShingle | Medical Management Hernandez-Silveira, Miguel Ahmed, Kamran Ang, Su-Shin Zandari, Fahriya Mehta, Tinaz Weir, Rebecca Burdett, Alison Toumazou, Chris Brett, Stephen J Assessment of the feasibility of an ultra-low power, wireless digital patch for the continuous ambulatory monitoring of vital signs |
title | Assessment of the feasibility of an ultra-low power, wireless digital patch for the continuous ambulatory monitoring of vital signs |
title_full | Assessment of the feasibility of an ultra-low power, wireless digital patch for the continuous ambulatory monitoring of vital signs |
title_fullStr | Assessment of the feasibility of an ultra-low power, wireless digital patch for the continuous ambulatory monitoring of vital signs |
title_full_unstemmed | Assessment of the feasibility of an ultra-low power, wireless digital patch for the continuous ambulatory monitoring of vital signs |
title_short | Assessment of the feasibility of an ultra-low power, wireless digital patch for the continuous ambulatory monitoring of vital signs |
title_sort | assessment of the feasibility of an ultra-low power, wireless digital patch for the continuous ambulatory monitoring of vital signs |
topic | Medical Management |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4442176/ https://www.ncbi.nlm.nih.gov/pubmed/25991447 http://dx.doi.org/10.1136/bmjopen-2014-006606 |
work_keys_str_mv | AT hernandezsilveiramiguel assessmentofthefeasibilityofanultralowpowerwirelessdigitalpatchforthecontinuousambulatorymonitoringofvitalsigns AT ahmedkamran assessmentofthefeasibilityofanultralowpowerwirelessdigitalpatchforthecontinuousambulatorymonitoringofvitalsigns AT angsushin assessmentofthefeasibilityofanultralowpowerwirelessdigitalpatchforthecontinuousambulatorymonitoringofvitalsigns AT zandarifahriya assessmentofthefeasibilityofanultralowpowerwirelessdigitalpatchforthecontinuousambulatorymonitoringofvitalsigns AT mehtatinaz assessmentofthefeasibilityofanultralowpowerwirelessdigitalpatchforthecontinuousambulatorymonitoringofvitalsigns AT weirrebecca assessmentofthefeasibilityofanultralowpowerwirelessdigitalpatchforthecontinuousambulatorymonitoringofvitalsigns AT burdettalison assessmentofthefeasibilityofanultralowpowerwirelessdigitalpatchforthecontinuousambulatorymonitoringofvitalsigns AT toumazouchris assessmentofthefeasibilityofanultralowpowerwirelessdigitalpatchforthecontinuousambulatorymonitoringofvitalsigns AT brettstephenj assessmentofthefeasibilityofanultralowpowerwirelessdigitalpatchforthecontinuousambulatorymonitoringofvitalsigns |