Comparison of the Refractive Measurements with Hand-held Autorefractometer, Table-mounted Autorefractometer and Cycloplegic Retinoscopy in Children
PURPOSE: To evaluate the performance of the hand-held and table-top autorefractokeratometer in measuring refractive errors by comparing them with cycloplegic retinoscopy. METHODS: Included in the study were 112 eyes of 112 pediatric patients whose mean age was 6.78 ± 2.61 years (range, 2 to 12 years...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Korean Ophthalmological Society
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4446558/ https://www.ncbi.nlm.nih.gov/pubmed/26028946 http://dx.doi.org/10.3341/kjo.2015.29.3.178 |
_version_ | 1782373452007604224 |
---|---|
author | Akil, Handan Keskin, Soner Çavdarli, Cemal |
author_facet | Akil, Handan Keskin, Soner Çavdarli, Cemal |
author_sort | Akil, Handan |
collection | PubMed |
description | PURPOSE: To evaluate the performance of the hand-held and table-top autorefractokeratometer in measuring refractive errors by comparing them with cycloplegic retinoscopy. METHODS: Included in the study were 112 eyes of 112 pediatric patients whose mean age was 6.78 ± 2.61 years (range, 2 to 12 years). The refractive errors of all the eyes were measured with and without cycloplegia using a hand held autorefractokeratometer (Retinomax K-plus 3), table top autorefractokeratometer (Canon RK-F1) and performing cycloplegic retinoscopy. The spherical equivalent, cylindrical axis and keratometer values were statistically compared. RESULTS: The mean spherical equivalent obtained from the Retinomax K-plus 3 was significantly less hyperopic than that of Canon RK-F1 (p = 0.004) before cycloplegia. When the Bland Altman analysis was performed in comparisons of spherical equivalent values measured with the Retinomax K-plus 3, Canon RK-F1 and cycloplegic retinoscopy, it was seen that almost all of the differences between the measurements remained within the range of ±2 standard deviation. Good agreement was found between Retinomax K-plus 3 and Canon RK-F1 for the Jackson cross-cylinder values at axis 0° and 45°; keratometer values respectively. CONCLUSIONS: The refractive error components were highly correlated between the two instruments and cycloplegic retinoscopy. |
format | Online Article Text |
id | pubmed-4446558 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | The Korean Ophthalmological Society |
record_format | MEDLINE/PubMed |
spelling | pubmed-44465582015-06-01 Comparison of the Refractive Measurements with Hand-held Autorefractometer, Table-mounted Autorefractometer and Cycloplegic Retinoscopy in Children Akil, Handan Keskin, Soner Çavdarli, Cemal Korean J Ophthalmol Original Article PURPOSE: To evaluate the performance of the hand-held and table-top autorefractokeratometer in measuring refractive errors by comparing them with cycloplegic retinoscopy. METHODS: Included in the study were 112 eyes of 112 pediatric patients whose mean age was 6.78 ± 2.61 years (range, 2 to 12 years). The refractive errors of all the eyes were measured with and without cycloplegia using a hand held autorefractokeratometer (Retinomax K-plus 3), table top autorefractokeratometer (Canon RK-F1) and performing cycloplegic retinoscopy. The spherical equivalent, cylindrical axis and keratometer values were statistically compared. RESULTS: The mean spherical equivalent obtained from the Retinomax K-plus 3 was significantly less hyperopic than that of Canon RK-F1 (p = 0.004) before cycloplegia. When the Bland Altman analysis was performed in comparisons of spherical equivalent values measured with the Retinomax K-plus 3, Canon RK-F1 and cycloplegic retinoscopy, it was seen that almost all of the differences between the measurements remained within the range of ±2 standard deviation. Good agreement was found between Retinomax K-plus 3 and Canon RK-F1 for the Jackson cross-cylinder values at axis 0° and 45°; keratometer values respectively. CONCLUSIONS: The refractive error components were highly correlated between the two instruments and cycloplegic retinoscopy. The Korean Ophthalmological Society 2015-06 2015-05-20 /pmc/articles/PMC4446558/ /pubmed/26028946 http://dx.doi.org/10.3341/kjo.2015.29.3.178 Text en © 2015 The Korean Ophthalmological Society http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Akil, Handan Keskin, Soner Çavdarli, Cemal Comparison of the Refractive Measurements with Hand-held Autorefractometer, Table-mounted Autorefractometer and Cycloplegic Retinoscopy in Children |
title | Comparison of the Refractive Measurements with Hand-held Autorefractometer, Table-mounted Autorefractometer and Cycloplegic Retinoscopy in Children |
title_full | Comparison of the Refractive Measurements with Hand-held Autorefractometer, Table-mounted Autorefractometer and Cycloplegic Retinoscopy in Children |
title_fullStr | Comparison of the Refractive Measurements with Hand-held Autorefractometer, Table-mounted Autorefractometer and Cycloplegic Retinoscopy in Children |
title_full_unstemmed | Comparison of the Refractive Measurements with Hand-held Autorefractometer, Table-mounted Autorefractometer and Cycloplegic Retinoscopy in Children |
title_short | Comparison of the Refractive Measurements with Hand-held Autorefractometer, Table-mounted Autorefractometer and Cycloplegic Retinoscopy in Children |
title_sort | comparison of the refractive measurements with hand-held autorefractometer, table-mounted autorefractometer and cycloplegic retinoscopy in children |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4446558/ https://www.ncbi.nlm.nih.gov/pubmed/26028946 http://dx.doi.org/10.3341/kjo.2015.29.3.178 |
work_keys_str_mv | AT akilhandan comparisonoftherefractivemeasurementswithhandheldautorefractometertablemountedautorefractometerandcycloplegicretinoscopyinchildren AT keskinsoner comparisonoftherefractivemeasurementswithhandheldautorefractometertablemountedautorefractometerandcycloplegicretinoscopyinchildren AT cavdarlicemal comparisonoftherefractivemeasurementswithhandheldautorefractometertablemountedautorefractometerandcycloplegicretinoscopyinchildren |