Fractional Exhaled Nitric Oxide: Comparison Between Portable Devices and Correlation With Sputum Eosinophils
This study was performed to compare the 2 different portable devices measuring fractional exhaled nitric oxide (FeNO) and to see the correlation between FeNO and induced sputum eosinophil count (ISE). Forty consecutive subjects clinically suspected to have asthma underwent FeNO measurement by NIOX-M...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Korean Academy of Asthma, Allergy and Clinical Immunology; The Korean Academy of Pediatric Allergy and Respiratory Disease
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4446639/ https://www.ncbi.nlm.nih.gov/pubmed/25749783 http://dx.doi.org/10.4168/aair.2015.7.4.404 |
_version_ | 1782373470568448000 |
---|---|
author | Yune, Sehyo Lee, Jin-Young Choi, Dong-Chull Lee, Byung-Jae |
author_facet | Yune, Sehyo Lee, Jin-Young Choi, Dong-Chull Lee, Byung-Jae |
author_sort | Yune, Sehyo |
collection | PubMed |
description | This study was performed to compare the 2 different portable devices measuring fractional exhaled nitric oxide (FeNO) and to see the correlation between FeNO and induced sputum eosinophil count (ISE). Forty consecutive subjects clinically suspected to have asthma underwent FeNO measurement by NIOX-MINO® and NObreath® concurrently. All also had induced sputum analysis, methacholine provocation test or bronchodilator response test, and spin prick test. Agreement between the 2 devices was evaluated. The correlation between FeNO and ISE was assessed, as well as the cut-off level of FeNO to identify ISE ≥3%. The intraclass correlation coefficient (ICC) between FeNO levels measured by NIOX-MINO® (FeNO(NIOX-MINO)) and NObreath® (FeNON(Obreath)) was 0.972 with 95% confidence interval of 0.948-0.985. The 95% limits of agreement were -28.9 to 19.9 ppb. The correlation coefficient between ISE and FeNO(NIOX-MINO) was 0.733 (P<0.001), and 0.751 between ISE and FeNO(NObreath) (P<0.001). The ROC curve found that the FeNO(NIOXMINO) of 37.5 ppb and the FeNO(NObreath) of 36.5 ppb identified ISE ≥3% with 90% sensitivity and 81% specificity. Age, sex, body mass index, smoking history, atopy, and the presence of asthma did not affect the FeNO level and its correlation with ISE. The NIOX-MINO ® and NObreath® agree with each other to a high degree. Both devices showed close correlation with ISE with similar cut-off value in identifying ISE ≥3%. |
format | Online Article Text |
id | pubmed-4446639 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | The Korean Academy of Asthma, Allergy and Clinical Immunology; The Korean Academy of Pediatric Allergy and Respiratory Disease |
record_format | MEDLINE/PubMed |
spelling | pubmed-44466392015-07-01 Fractional Exhaled Nitric Oxide: Comparison Between Portable Devices and Correlation With Sputum Eosinophils Yune, Sehyo Lee, Jin-Young Choi, Dong-Chull Lee, Byung-Jae Allergy Asthma Immunol Res Brief Communication This study was performed to compare the 2 different portable devices measuring fractional exhaled nitric oxide (FeNO) and to see the correlation between FeNO and induced sputum eosinophil count (ISE). Forty consecutive subjects clinically suspected to have asthma underwent FeNO measurement by NIOX-MINO® and NObreath® concurrently. All also had induced sputum analysis, methacholine provocation test or bronchodilator response test, and spin prick test. Agreement between the 2 devices was evaluated. The correlation between FeNO and ISE was assessed, as well as the cut-off level of FeNO to identify ISE ≥3%. The intraclass correlation coefficient (ICC) between FeNO levels measured by NIOX-MINO® (FeNO(NIOX-MINO)) and NObreath® (FeNON(Obreath)) was 0.972 with 95% confidence interval of 0.948-0.985. The 95% limits of agreement were -28.9 to 19.9 ppb. The correlation coefficient between ISE and FeNO(NIOX-MINO) was 0.733 (P<0.001), and 0.751 between ISE and FeNO(NObreath) (P<0.001). The ROC curve found that the FeNO(NIOXMINO) of 37.5 ppb and the FeNO(NObreath) of 36.5 ppb identified ISE ≥3% with 90% sensitivity and 81% specificity. Age, sex, body mass index, smoking history, atopy, and the presence of asthma did not affect the FeNO level and its correlation with ISE. The NIOX-MINO ® and NObreath® agree with each other to a high degree. Both devices showed close correlation with ISE with similar cut-off value in identifying ISE ≥3%. The Korean Academy of Asthma, Allergy and Clinical Immunology; The Korean Academy of Pediatric Allergy and Respiratory Disease 2015-07 2015-03-27 /pmc/articles/PMC4446639/ /pubmed/25749783 http://dx.doi.org/10.4168/aair.2015.7.4.404 Text en Copyright © 2015 The Korean Academy of Asthma, Allergy and Clinical Immunology • The Korean Academy of Pediatric Allergy and Respiratory Disease http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Brief Communication Yune, Sehyo Lee, Jin-Young Choi, Dong-Chull Lee, Byung-Jae Fractional Exhaled Nitric Oxide: Comparison Between Portable Devices and Correlation With Sputum Eosinophils |
title | Fractional Exhaled Nitric Oxide: Comparison Between Portable Devices and Correlation With Sputum Eosinophils |
title_full | Fractional Exhaled Nitric Oxide: Comparison Between Portable Devices and Correlation With Sputum Eosinophils |
title_fullStr | Fractional Exhaled Nitric Oxide: Comparison Between Portable Devices and Correlation With Sputum Eosinophils |
title_full_unstemmed | Fractional Exhaled Nitric Oxide: Comparison Between Portable Devices and Correlation With Sputum Eosinophils |
title_short | Fractional Exhaled Nitric Oxide: Comparison Between Portable Devices and Correlation With Sputum Eosinophils |
title_sort | fractional exhaled nitric oxide: comparison between portable devices and correlation with sputum eosinophils |
topic | Brief Communication |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4446639/ https://www.ncbi.nlm.nih.gov/pubmed/25749783 http://dx.doi.org/10.4168/aair.2015.7.4.404 |
work_keys_str_mv | AT yunesehyo fractionalexhalednitricoxidecomparisonbetweenportabledevicesandcorrelationwithsputumeosinophils AT leejinyoung fractionalexhalednitricoxidecomparisonbetweenportabledevicesandcorrelationwithsputumeosinophils AT choidongchull fractionalexhalednitricoxidecomparisonbetweenportabledevicesandcorrelationwithsputumeosinophils AT leebyungjae fractionalexhalednitricoxidecomparisonbetweenportabledevicesandcorrelationwithsputumeosinophils |