Fractional Exhaled Nitric Oxide: Comparison Between Portable Devices and Correlation With Sputum Eosinophils

This study was performed to compare the 2 different portable devices measuring fractional exhaled nitric oxide (FeNO) and to see the correlation between FeNO and induced sputum eosinophil count (ISE). Forty consecutive subjects clinically suspected to have asthma underwent FeNO measurement by NIOX-M...

Descripción completa

Detalles Bibliográficos
Autores principales: Yune, Sehyo, Lee, Jin-Young, Choi, Dong-Chull, Lee, Byung-Jae
Formato: Online Artículo Texto
Lenguaje:English
Publicado: The Korean Academy of Asthma, Allergy and Clinical Immunology; The Korean Academy of Pediatric Allergy and Respiratory Disease 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4446639/
https://www.ncbi.nlm.nih.gov/pubmed/25749783
http://dx.doi.org/10.4168/aair.2015.7.4.404
_version_ 1782373470568448000
author Yune, Sehyo
Lee, Jin-Young
Choi, Dong-Chull
Lee, Byung-Jae
author_facet Yune, Sehyo
Lee, Jin-Young
Choi, Dong-Chull
Lee, Byung-Jae
author_sort Yune, Sehyo
collection PubMed
description This study was performed to compare the 2 different portable devices measuring fractional exhaled nitric oxide (FeNO) and to see the correlation between FeNO and induced sputum eosinophil count (ISE). Forty consecutive subjects clinically suspected to have asthma underwent FeNO measurement by NIOX-MINO® and NObreath® concurrently. All also had induced sputum analysis, methacholine provocation test or bronchodilator response test, and spin prick test. Agreement between the 2 devices was evaluated. The correlation between FeNO and ISE was assessed, as well as the cut-off level of FeNO to identify ISE ≥3%. The intraclass correlation coefficient (ICC) between FeNO levels measured by NIOX-MINO® (FeNO(NIOX-MINO)) and NObreath® (FeNON(Obreath)) was 0.972 with 95% confidence interval of 0.948-0.985. The 95% limits of agreement were -28.9 to 19.9 ppb. The correlation coefficient between ISE and FeNO(NIOX-MINO) was 0.733 (P<0.001), and 0.751 between ISE and FeNO(NObreath) (P<0.001). The ROC curve found that the FeNO(NIOXMINO) of 37.5 ppb and the FeNO(NObreath) of 36.5 ppb identified ISE ≥3% with 90% sensitivity and 81% specificity. Age, sex, body mass index, smoking history, atopy, and the presence of asthma did not affect the FeNO level and its correlation with ISE. The NIOX-MINO ® and NObreath® agree with each other to a high degree. Both devices showed close correlation with ISE with similar cut-off value in identifying ISE ≥3%.
format Online
Article
Text
id pubmed-4446639
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher The Korean Academy of Asthma, Allergy and Clinical Immunology; The Korean Academy of Pediatric Allergy and Respiratory Disease
record_format MEDLINE/PubMed
spelling pubmed-44466392015-07-01 Fractional Exhaled Nitric Oxide: Comparison Between Portable Devices and Correlation With Sputum Eosinophils Yune, Sehyo Lee, Jin-Young Choi, Dong-Chull Lee, Byung-Jae Allergy Asthma Immunol Res Brief Communication This study was performed to compare the 2 different portable devices measuring fractional exhaled nitric oxide (FeNO) and to see the correlation between FeNO and induced sputum eosinophil count (ISE). Forty consecutive subjects clinically suspected to have asthma underwent FeNO measurement by NIOX-MINO® and NObreath® concurrently. All also had induced sputum analysis, methacholine provocation test or bronchodilator response test, and spin prick test. Agreement between the 2 devices was evaluated. The correlation between FeNO and ISE was assessed, as well as the cut-off level of FeNO to identify ISE ≥3%. The intraclass correlation coefficient (ICC) between FeNO levels measured by NIOX-MINO® (FeNO(NIOX-MINO)) and NObreath® (FeNON(Obreath)) was 0.972 with 95% confidence interval of 0.948-0.985. The 95% limits of agreement were -28.9 to 19.9 ppb. The correlation coefficient between ISE and FeNO(NIOX-MINO) was 0.733 (P<0.001), and 0.751 between ISE and FeNO(NObreath) (P<0.001). The ROC curve found that the FeNO(NIOXMINO) of 37.5 ppb and the FeNO(NObreath) of 36.5 ppb identified ISE ≥3% with 90% sensitivity and 81% specificity. Age, sex, body mass index, smoking history, atopy, and the presence of asthma did not affect the FeNO level and its correlation with ISE. The NIOX-MINO ® and NObreath® agree with each other to a high degree. Both devices showed close correlation with ISE with similar cut-off value in identifying ISE ≥3%. The Korean Academy of Asthma, Allergy and Clinical Immunology; The Korean Academy of Pediatric Allergy and Respiratory Disease 2015-07 2015-03-27 /pmc/articles/PMC4446639/ /pubmed/25749783 http://dx.doi.org/10.4168/aair.2015.7.4.404 Text en Copyright © 2015 The Korean Academy of Asthma, Allergy and Clinical Immunology • The Korean Academy of Pediatric Allergy and Respiratory Disease http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Brief Communication
Yune, Sehyo
Lee, Jin-Young
Choi, Dong-Chull
Lee, Byung-Jae
Fractional Exhaled Nitric Oxide: Comparison Between Portable Devices and Correlation With Sputum Eosinophils
title Fractional Exhaled Nitric Oxide: Comparison Between Portable Devices and Correlation With Sputum Eosinophils
title_full Fractional Exhaled Nitric Oxide: Comparison Between Portable Devices and Correlation With Sputum Eosinophils
title_fullStr Fractional Exhaled Nitric Oxide: Comparison Between Portable Devices and Correlation With Sputum Eosinophils
title_full_unstemmed Fractional Exhaled Nitric Oxide: Comparison Between Portable Devices and Correlation With Sputum Eosinophils
title_short Fractional Exhaled Nitric Oxide: Comparison Between Portable Devices and Correlation With Sputum Eosinophils
title_sort fractional exhaled nitric oxide: comparison between portable devices and correlation with sputum eosinophils
topic Brief Communication
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4446639/
https://www.ncbi.nlm.nih.gov/pubmed/25749783
http://dx.doi.org/10.4168/aair.2015.7.4.404
work_keys_str_mv AT yunesehyo fractionalexhalednitricoxidecomparisonbetweenportabledevicesandcorrelationwithsputumeosinophils
AT leejinyoung fractionalexhalednitricoxidecomparisonbetweenportabledevicesandcorrelationwithsputumeosinophils
AT choidongchull fractionalexhalednitricoxidecomparisonbetweenportabledevicesandcorrelationwithsputumeosinophils
AT leebyungjae fractionalexhalednitricoxidecomparisonbetweenportabledevicesandcorrelationwithsputumeosinophils