A novel method based on two cameras for accurate estimation of arterial oxygen saturation

BACKGROUND: Photoplethysmographic imaging (PPGi) that is based on camera allows acquiring photoplethysmogram and measuring physiological parameters such as pulse rate, respiration rate and perfusion level. It has also shown potential for estimation of arterial oxygen saturation (SaO(2)). However, th...

Descripción completa

Detalles Bibliográficos
Autores principales: Liu, He, Ivanov, Kamen, Wang, Yadong, Wang, Lei
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4449570/
https://www.ncbi.nlm.nih.gov/pubmed/26025439
http://dx.doi.org/10.1186/s12938-015-0045-1
_version_ 1782373878481289216
author Liu, He
Ivanov, Kamen
Wang, Yadong
Wang, Lei
author_facet Liu, He
Ivanov, Kamen
Wang, Yadong
Wang, Lei
author_sort Liu, He
collection PubMed
description BACKGROUND: Photoplethysmographic imaging (PPGi) that is based on camera allows acquiring photoplethysmogram and measuring physiological parameters such as pulse rate, respiration rate and perfusion level. It has also shown potential for estimation of arterial oxygen saturation (SaO(2)). However, there are some technical limitations such as optical shunting, different camera sensitivity to different light spectra, different AC-to-DC ratios (the peak-to-peak amplitude to baseline ratio) of the PPGi signal for different portions of the sensor surface area, the low sampling rate and the inconsistency of contact force between the fingertip and camera lens. METHODS: In this paper, we take full account of the above-mentioned design challenges and present an accurate SaO(2) estimation method based on two cameras. The hardware system we used consisted of an FPGA development board (XC6SLX150T-3FGG676 from Xilinx), with connected to it two commercial cameras and an SD card. The two cameras were placed back to back, one camera acquired PPGi signal from the right index fingertip under 660 nm light illumination while the other camera acquired PPGi signal from the thumb fingertip using an 800 nm light illumination. The both PPGi signals were captured simultaneously, recorded in a text file on the SD card and processed offline using MATLAB®. The calculation of SaO(2) was based on the principle of pulse oximetry. The AC-to-DC ratio was acquired by the ratio of powers of AC and DC components of the PPGi signal in the time–frequency domain using the smoothed pseudo Wigner–Ville distribution. The calibration curve required for SaO(2) measurement was obtained by linear regression analysis. RESULTS: The results of our estimation method from 12 subjects showed a high correlation and accuracy with those of conventional pulse oximetry for the range from 90 to 100%. CONCLUSIONS: Our method is suitable for mobile applications implemented in smartphones, which could allow SaO(2) measurement in a pervasive environment.
format Online
Article
Text
id pubmed-4449570
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-44495702015-05-31 A novel method based on two cameras for accurate estimation of arterial oxygen saturation Liu, He Ivanov, Kamen Wang, Yadong Wang, Lei Biomed Eng Online Research BACKGROUND: Photoplethysmographic imaging (PPGi) that is based on camera allows acquiring photoplethysmogram and measuring physiological parameters such as pulse rate, respiration rate and perfusion level. It has also shown potential for estimation of arterial oxygen saturation (SaO(2)). However, there are some technical limitations such as optical shunting, different camera sensitivity to different light spectra, different AC-to-DC ratios (the peak-to-peak amplitude to baseline ratio) of the PPGi signal for different portions of the sensor surface area, the low sampling rate and the inconsistency of contact force between the fingertip and camera lens. METHODS: In this paper, we take full account of the above-mentioned design challenges and present an accurate SaO(2) estimation method based on two cameras. The hardware system we used consisted of an FPGA development board (XC6SLX150T-3FGG676 from Xilinx), with connected to it two commercial cameras and an SD card. The two cameras were placed back to back, one camera acquired PPGi signal from the right index fingertip under 660 nm light illumination while the other camera acquired PPGi signal from the thumb fingertip using an 800 nm light illumination. The both PPGi signals were captured simultaneously, recorded in a text file on the SD card and processed offline using MATLAB®. The calculation of SaO(2) was based on the principle of pulse oximetry. The AC-to-DC ratio was acquired by the ratio of powers of AC and DC components of the PPGi signal in the time–frequency domain using the smoothed pseudo Wigner–Ville distribution. The calibration curve required for SaO(2) measurement was obtained by linear regression analysis. RESULTS: The results of our estimation method from 12 subjects showed a high correlation and accuracy with those of conventional pulse oximetry for the range from 90 to 100%. CONCLUSIONS: Our method is suitable for mobile applications implemented in smartphones, which could allow SaO(2) measurement in a pervasive environment. BioMed Central 2015-05-30 /pmc/articles/PMC4449570/ /pubmed/26025439 http://dx.doi.org/10.1186/s12938-015-0045-1 Text en © Liu et al. 2015 Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research
Liu, He
Ivanov, Kamen
Wang, Yadong
Wang, Lei
A novel method based on two cameras for accurate estimation of arterial oxygen saturation
title A novel method based on two cameras for accurate estimation of arterial oxygen saturation
title_full A novel method based on two cameras for accurate estimation of arterial oxygen saturation
title_fullStr A novel method based on two cameras for accurate estimation of arterial oxygen saturation
title_full_unstemmed A novel method based on two cameras for accurate estimation of arterial oxygen saturation
title_short A novel method based on two cameras for accurate estimation of arterial oxygen saturation
title_sort novel method based on two cameras for accurate estimation of arterial oxygen saturation
topic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4449570/
https://www.ncbi.nlm.nih.gov/pubmed/26025439
http://dx.doi.org/10.1186/s12938-015-0045-1
work_keys_str_mv AT liuhe anovelmethodbasedontwocamerasforaccurateestimationofarterialoxygensaturation
AT ivanovkamen anovelmethodbasedontwocamerasforaccurateestimationofarterialoxygensaturation
AT wangyadong anovelmethodbasedontwocamerasforaccurateestimationofarterialoxygensaturation
AT wanglei anovelmethodbasedontwocamerasforaccurateestimationofarterialoxygensaturation
AT liuhe novelmethodbasedontwocamerasforaccurateestimationofarterialoxygensaturation
AT ivanovkamen novelmethodbasedontwocamerasforaccurateestimationofarterialoxygensaturation
AT wangyadong novelmethodbasedontwocamerasforaccurateestimationofarterialoxygensaturation
AT wanglei novelmethodbasedontwocamerasforaccurateestimationofarterialoxygensaturation