BACKGROUND: Although NaI(Tl) gamma counters play an important role in many quantitative positron emission tomography (PET) protocols, their calibration for positron-emitting samples has not been standardized across imaging sites. In this study, we characterized the operational range of a gamma count...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer International Publishing
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4452125/ https://www.ncbi.nlm.nih.gov/pubmed/26046011 http://dx.doi.org/10.1186/s40658-015-0114-3 |
_version_ | 1782374254172438528 |
---|---|
author | Lodge, Martin A Holt, Daniel P Kinahan, Paul E Wong, Dean F Wahl, Richard L |
author_facet | Lodge, Martin A Holt, Daniel P Kinahan, Paul E Wong, Dean F Wahl, Richard L |
author_sort | Lodge, Martin A |
collection | PubMed |
description | BACKGROUND: Although NaI(Tl) gamma counters play an important role in many quantitative positron emission tomography (PET) protocols, their calibration for positron-emitting samples has not been standardized across imaging sites. In this study, we characterized the operational range of a gamma counter specifically for positron-emitting radionuclides, and we assessed the role of traceable (68)Ge/(68)Ga sources for standardizing system calibration. METHODS: A NaI(Tl) gamma counter was characterized with respect to count rate performance, adequacy of detector shielding, system stability, and sample volume effects using positron-emitting radionuclides (409- to 613-keV energy window). System efficiency was measured using (18)F and compared with corresponding data obtained using a long-lived (68)Ge/(68)Ga source that was implicitly traceable to a national standard. RESULTS: One percent count loss was measured at 450 × 10(3) counts per minute. Penetration of the detector shielding by 511-keV photons gave rise to a negligible background count rate. System stability tests showed a coefficient of variation of 0.13% over 100 days. For a sample volume of 4 mL, the efficiencies relative to those at 0.1 mL were 0.96, 0.94, 0.91, 0.78, and 0.72 for (11)C, (18)F, (125)I, (99m)Tc, and (51)Cr, respectively. The efficiency of a traceable (68)Ge/(68)Ga source was 30.1% ± 0.07% and was found to be in close agreement with the efficiency for (18)F after consideration of the different positron fractions. CONCLUSIONS: Long-lived (68)Ge/(68)Ga reference sources, implicitly traceable to a national metrology institute, can aid standardization of gamma counter calibration for (18)F. A characteristic feature of positron emitters meant that accurate calibration could be maintained over a wide range of sample volumes by using a narrow energy window centered on the 511-keV peak. |
format | Online Article Text |
id | pubmed-4452125 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | Springer International Publishing |
record_format | MEDLINE/PubMed |
spelling | pubmed-44521252015-06-02 Performance assessment of a NaI(Tl) gamma counter for PET applications with methods for improved quantitative accuracy and greater standardization Lodge, Martin A Holt, Daniel P Kinahan, Paul E Wong, Dean F Wahl, Richard L EJNMMI Phys Original Research BACKGROUND: Although NaI(Tl) gamma counters play an important role in many quantitative positron emission tomography (PET) protocols, their calibration for positron-emitting samples has not been standardized across imaging sites. In this study, we characterized the operational range of a gamma counter specifically for positron-emitting radionuclides, and we assessed the role of traceable (68)Ge/(68)Ga sources for standardizing system calibration. METHODS: A NaI(Tl) gamma counter was characterized with respect to count rate performance, adequacy of detector shielding, system stability, and sample volume effects using positron-emitting radionuclides (409- to 613-keV energy window). System efficiency was measured using (18)F and compared with corresponding data obtained using a long-lived (68)Ge/(68)Ga source that was implicitly traceable to a national standard. RESULTS: One percent count loss was measured at 450 × 10(3) counts per minute. Penetration of the detector shielding by 511-keV photons gave rise to a negligible background count rate. System stability tests showed a coefficient of variation of 0.13% over 100 days. For a sample volume of 4 mL, the efficiencies relative to those at 0.1 mL were 0.96, 0.94, 0.91, 0.78, and 0.72 for (11)C, (18)F, (125)I, (99m)Tc, and (51)Cr, respectively. The efficiency of a traceable (68)Ge/(68)Ga source was 30.1% ± 0.07% and was found to be in close agreement with the efficiency for (18)F after consideration of the different positron fractions. CONCLUSIONS: Long-lived (68)Ge/(68)Ga reference sources, implicitly traceable to a national metrology institute, can aid standardization of gamma counter calibration for (18)F. A characteristic feature of positron emitters meant that accurate calibration could be maintained over a wide range of sample volumes by using a narrow energy window centered on the 511-keV peak. Springer International Publishing 2015-05-06 /pmc/articles/PMC4452125/ /pubmed/26046011 http://dx.doi.org/10.1186/s40658-015-0114-3 Text en © Lodge et al.; licensee Springer. 2015 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. |
spellingShingle | Original Research Lodge, Martin A Holt, Daniel P Kinahan, Paul E Wong, Dean F Wahl, Richard L Performance assessment of a NaI(Tl) gamma counter for PET applications with methods for improved quantitative accuracy and greater standardization |
title | Performance assessment of a NaI(Tl) gamma counter for PET applications with methods for improved quantitative accuracy and greater standardization |
title_full | Performance assessment of a NaI(Tl) gamma counter for PET applications with methods for improved quantitative accuracy and greater standardization |
title_fullStr | Performance assessment of a NaI(Tl) gamma counter for PET applications with methods for improved quantitative accuracy and greater standardization |
title_full_unstemmed | Performance assessment of a NaI(Tl) gamma counter for PET applications with methods for improved quantitative accuracy and greater standardization |
title_short | Performance assessment of a NaI(Tl) gamma counter for PET applications with methods for improved quantitative accuracy and greater standardization |
title_sort | performance assessment of a nai(tl) gamma counter for pet applications with methods for improved quantitative accuracy and greater standardization |
topic | Original Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4452125/ https://www.ncbi.nlm.nih.gov/pubmed/26046011 http://dx.doi.org/10.1186/s40658-015-0114-3 |
work_keys_str_mv | AT lodgemartina performanceassessmentofanaitlgammacounterforpetapplicationswithmethodsforimprovedquantitativeaccuracyandgreaterstandardization AT holtdanielp performanceassessmentofanaitlgammacounterforpetapplicationswithmethodsforimprovedquantitativeaccuracyandgreaterstandardization AT kinahanpaule performanceassessmentofanaitlgammacounterforpetapplicationswithmethodsforimprovedquantitativeaccuracyandgreaterstandardization AT wongdeanf performanceassessmentofanaitlgammacounterforpetapplicationswithmethodsforimprovedquantitativeaccuracyandgreaterstandardization AT wahlrichardl performanceassessmentofanaitlgammacounterforpetapplicationswithmethodsforimprovedquantitativeaccuracyandgreaterstandardization |