Cargando…
Transection of a Coopdech bronchial blocker tip during bronchial resection for right upper lobectomy: a case report
A bronchial blocker (BB) is preferred for lung separation in patients with difficult airways. However, BBs, unlike double-lumen tubes, must be placed in the bronchus of the lung being operated on, hence can be damaged by surgical manipulation. Intubation was unexpectedly difficult in this male patie...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Korean Society of Anesthesiologists
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4452674/ https://www.ncbi.nlm.nih.gov/pubmed/26045933 http://dx.doi.org/10.4097/kjae.2015.68.3.287 |
_version_ | 1782374337582465024 |
---|---|
author | Lee, Yong-Hun Yang, Hye Mo Kim, Hyun-Chang Bahk, Jae-Hyon Seo, Jeong-Hwa |
author_facet | Lee, Yong-Hun Yang, Hye Mo Kim, Hyun-Chang Bahk, Jae-Hyon Seo, Jeong-Hwa |
author_sort | Lee, Yong-Hun |
collection | PubMed |
description | A bronchial blocker (BB) is preferred for lung separation in patients with difficult airways. However, BBs, unlike double-lumen tubes, must be placed in the bronchus of the lung being operated on, hence can be damaged by surgical manipulation. Intubation was unexpectedly difficult in this male patient, so a Coopdech BB was placed in the right mainstem bronchus through a single-lumen tracheoscopic ventilation tube for a thoracoscopic right upper lobectomy. During the bronchial resection, however, the distal tip of the BB was transected and pinched in the staple line, so the staple line was partially opened, and the BB was withdrawn into the trachea. The opened bronchial stump was sutured manually under apnea without conversion to an open thoracotomy, and there was no significant air leakage through the suture line. This case underlines the importance of frequently evaluating the position of a BB during lung surgery. |
format | Online Article Text |
id | pubmed-4452674 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | The Korean Society of Anesthesiologists |
record_format | MEDLINE/PubMed |
spelling | pubmed-44526742015-06-04 Transection of a Coopdech bronchial blocker tip during bronchial resection for right upper lobectomy: a case report Lee, Yong-Hun Yang, Hye Mo Kim, Hyun-Chang Bahk, Jae-Hyon Seo, Jeong-Hwa Korean J Anesthesiol Case Report A bronchial blocker (BB) is preferred for lung separation in patients with difficult airways. However, BBs, unlike double-lumen tubes, must be placed in the bronchus of the lung being operated on, hence can be damaged by surgical manipulation. Intubation was unexpectedly difficult in this male patient, so a Coopdech BB was placed in the right mainstem bronchus through a single-lumen tracheoscopic ventilation tube for a thoracoscopic right upper lobectomy. During the bronchial resection, however, the distal tip of the BB was transected and pinched in the staple line, so the staple line was partially opened, and the BB was withdrawn into the trachea. The opened bronchial stump was sutured manually under apnea without conversion to an open thoracotomy, and there was no significant air leakage through the suture line. This case underlines the importance of frequently evaluating the position of a BB during lung surgery. The Korean Society of Anesthesiologists 2015-06 2015-05-28 /pmc/articles/PMC4452674/ /pubmed/26045933 http://dx.doi.org/10.4097/kjae.2015.68.3.287 Text en Copyright © the Korean Society of Anesthesiologists, 2015 http://creativecommons.org/licenses/by-nc/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0/), which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Case Report Lee, Yong-Hun Yang, Hye Mo Kim, Hyun-Chang Bahk, Jae-Hyon Seo, Jeong-Hwa Transection of a Coopdech bronchial blocker tip during bronchial resection for right upper lobectomy: a case report |
title | Transection of a Coopdech bronchial blocker tip during bronchial resection for right upper lobectomy: a case report |
title_full | Transection of a Coopdech bronchial blocker tip during bronchial resection for right upper lobectomy: a case report |
title_fullStr | Transection of a Coopdech bronchial blocker tip during bronchial resection for right upper lobectomy: a case report |
title_full_unstemmed | Transection of a Coopdech bronchial blocker tip during bronchial resection for right upper lobectomy: a case report |
title_short | Transection of a Coopdech bronchial blocker tip during bronchial resection for right upper lobectomy: a case report |
title_sort | transection of a coopdech bronchial blocker tip during bronchial resection for right upper lobectomy: a case report |
topic | Case Report |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4452674/ https://www.ncbi.nlm.nih.gov/pubmed/26045933 http://dx.doi.org/10.4097/kjae.2015.68.3.287 |
work_keys_str_mv | AT leeyonghun transectionofacoopdechbronchialblockertipduringbronchialresectionforrightupperlobectomyacasereport AT yanghyemo transectionofacoopdechbronchialblockertipduringbronchialresectionforrightupperlobectomyacasereport AT kimhyunchang transectionofacoopdechbronchialblockertipduringbronchialresectionforrightupperlobectomyacasereport AT bahkjaehyon transectionofacoopdechbronchialblockertipduringbronchialresectionforrightupperlobectomyacasereport AT seojeonghwa transectionofacoopdechbronchialblockertipduringbronchialresectionforrightupperlobectomyacasereport |