Cargando…

A Case Report of Implant Fracture of Extensively Porous-Coated, Distally Fixated Cementless Long Stem: Detailed Course of Stem Bending Development

We report the fracture of a cementless long stem in a distally fixated, extensively porous-coated femoral prosthesis used for revisional total hip arthroplasty (THA) 9 years previously in a 48-year-old woman, measuring 58 kg in weight and 155 cm in height. Following resolution of an infection 7 year...

Descripción completa

Detalles Bibliográficos
Autor principal: Mimura, Tomohiro
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4454702/
https://www.ncbi.nlm.nih.gov/pubmed/26078901
http://dx.doi.org/10.1155/2015/895214
_version_ 1782374639952986112
author Mimura, Tomohiro
author_facet Mimura, Tomohiro
author_sort Mimura, Tomohiro
collection PubMed
description We report the fracture of a cementless long stem in a distally fixated, extensively porous-coated femoral prosthesis used for revisional total hip arthroplasty (THA) 9 years previously in a 48-year-old woman, measuring 58 kg in weight and 155 cm in height. Following resolution of an infection 7 years after the revisional THA, a series of posterior dislocations occurred up until 7 months before sudden stem fracture. Extensive radiographic imaging evidence indicated a gradual progression of bending in the stem, and scanning electron microscope energy dispersive X-ray spectroscopy revealed oxygen in the medial and lateral sections, suspecting in vivo corrosion. We retrospectively examined the detailed course of stem bending development prior to stem fracture. The stem bending immediately after the revisional THA, at the first dislocation, and immediately before the stem fracture was 0°, 1.9°, and 5.2°, respectively. We consider that the cyclic loading with poor proximal bone support, distal fixation, and small diameter were the potential risk of this implant fracture. However, the course of stem bending development suggested that repeated operations and several dislocations might have aggravated the implant fatigue in the present case.
format Online
Article
Text
id pubmed-4454702
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-44547022015-06-15 A Case Report of Implant Fracture of Extensively Porous-Coated, Distally Fixated Cementless Long Stem: Detailed Course of Stem Bending Development Mimura, Tomohiro Case Rep Orthop Case Report We report the fracture of a cementless long stem in a distally fixated, extensively porous-coated femoral prosthesis used for revisional total hip arthroplasty (THA) 9 years previously in a 48-year-old woman, measuring 58 kg in weight and 155 cm in height. Following resolution of an infection 7 years after the revisional THA, a series of posterior dislocations occurred up until 7 months before sudden stem fracture. Extensive radiographic imaging evidence indicated a gradual progression of bending in the stem, and scanning electron microscope energy dispersive X-ray spectroscopy revealed oxygen in the medial and lateral sections, suspecting in vivo corrosion. We retrospectively examined the detailed course of stem bending development prior to stem fracture. The stem bending immediately after the revisional THA, at the first dislocation, and immediately before the stem fracture was 0°, 1.9°, and 5.2°, respectively. We consider that the cyclic loading with poor proximal bone support, distal fixation, and small diameter were the potential risk of this implant fracture. However, the course of stem bending development suggested that repeated operations and several dislocations might have aggravated the implant fatigue in the present case. Hindawi Publishing Corporation 2015 2015-05-21 /pmc/articles/PMC4454702/ /pubmed/26078901 http://dx.doi.org/10.1155/2015/895214 Text en Copyright © 2015 Tomohiro Mimura. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Case Report
Mimura, Tomohiro
A Case Report of Implant Fracture of Extensively Porous-Coated, Distally Fixated Cementless Long Stem: Detailed Course of Stem Bending Development
title A Case Report of Implant Fracture of Extensively Porous-Coated, Distally Fixated Cementless Long Stem: Detailed Course of Stem Bending Development
title_full A Case Report of Implant Fracture of Extensively Porous-Coated, Distally Fixated Cementless Long Stem: Detailed Course of Stem Bending Development
title_fullStr A Case Report of Implant Fracture of Extensively Porous-Coated, Distally Fixated Cementless Long Stem: Detailed Course of Stem Bending Development
title_full_unstemmed A Case Report of Implant Fracture of Extensively Porous-Coated, Distally Fixated Cementless Long Stem: Detailed Course of Stem Bending Development
title_short A Case Report of Implant Fracture of Extensively Porous-Coated, Distally Fixated Cementless Long Stem: Detailed Course of Stem Bending Development
title_sort case report of implant fracture of extensively porous-coated, distally fixated cementless long stem: detailed course of stem bending development
topic Case Report
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4454702/
https://www.ncbi.nlm.nih.gov/pubmed/26078901
http://dx.doi.org/10.1155/2015/895214
work_keys_str_mv AT mimuratomohiro acasereportofimplantfractureofextensivelyporouscoateddistallyfixatedcementlesslongstemdetailedcourseofstembendingdevelopment
AT mimuratomohiro casereportofimplantfractureofextensivelyporouscoateddistallyfixatedcementlesslongstemdetailedcourseofstembendingdevelopment