Counteracting negative venous line pressures to avoid arterial air bubbles: an experimental study comparing two different types of miniaturized extracorporeal perfusion systems

BACKGROUND: Because of its low rate of clinical complications, miniaturized extracorporeal perfusion systems (MEPS) are frequently used in heart centers worldwide. However, many recent studies refer to the higher probability of gaseous microemboli formation by MEPS, caused by subzero pressure values...

Descripción completa

Detalles Bibliográficos
Autores principales: Aboud, Anas, Mederos-Dahms, Hendrikje, Liebing, Kai, Zittermann, Armin, Schubert, Harald, Murray, Edward, Renner, Andre, Gummert, Jan, Börgermann, Jochen
Formato: Online Artículo Texto
Lenguaje:English
Publicado: BioMed Central 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4459480/
https://www.ncbi.nlm.nih.gov/pubmed/26021999
http://dx.doi.org/10.1186/s12871-015-0058-0
_version_ 1782375231664422912
author Aboud, Anas
Mederos-Dahms, Hendrikje
Liebing, Kai
Zittermann, Armin
Schubert, Harald
Murray, Edward
Renner, Andre
Gummert, Jan
Börgermann, Jochen
author_facet Aboud, Anas
Mederos-Dahms, Hendrikje
Liebing, Kai
Zittermann, Armin
Schubert, Harald
Murray, Edward
Renner, Andre
Gummert, Jan
Börgermann, Jochen
author_sort Aboud, Anas
collection PubMed
description BACKGROUND: Because of its low rate of clinical complications, miniaturized extracorporeal perfusion systems (MEPS) are frequently used in heart centers worldwide. However, many recent studies refer to the higher probability of gaseous microemboli formation by MEPS, caused by subzero pressure values. This is the main reason why various de-airing devices were developed for today’s perfusion systems. In the present study, we investigated the potential benefits of a simple one-way-valve connected to a volume replacement reservoir (OVR) for volume and pressure compensation. METHODS: In an experimental study on 26 pigs, we compared MEPS (n = 13) with MEPS plus OVR (n = 13). Except OVR, perfusion equipment was identical in both groups. Primary endpoints were pressure values in the venous line and the right atrium as well as the number and volume of air bubbles. Secondary endpoints were biochemical parameters of systemic inflammatory response, ischemia, hemodilution and hemolysis. RESULTS: One animal was lost in the MEPS + OVR group. In the MEPS + OVR group no pressure values below −150 mmHg in the venous line and no values under -100 mmHg in right atrium were noticed. On the contrary, nearly 20 % of venous pressure values in the MEPS group were below −150 and approximately 10 % of right atrial pressure values were below -100 mmHg. Compared with the MEPS group, the bubble counter device showed lower numbers of arterial air bubbles in the MEPS + OVR group (mean ± SD: 13444 ± 5709 vs. 1 ± 2, respectively; p < 0.001). In addition, bubble volume was significantly lower in the MEPS + OVR group than in the MEPS group (mean ± SD: 1522 ± 654 μl vs. 4 ± 6 μl, respectively; p < 0.001). The proinflammatory cytokine interleukin-6 and biochemical indices of cardiac ischemia (creatine kinase, and troponin I) were comparable between both groups. CONCLUSIONS: The use of a miniaturized perfusion system with a volume replacement reservoir is able to counteract excessive negative venous line pressures and to reduce the number and volume of arterial air bubbles. This approach may lead to a lower rate of neurological complications.
format Online
Article
Text
id pubmed-4459480
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher BioMed Central
record_format MEDLINE/PubMed
spelling pubmed-44594802015-06-09 Counteracting negative venous line pressures to avoid arterial air bubbles: an experimental study comparing two different types of miniaturized extracorporeal perfusion systems Aboud, Anas Mederos-Dahms, Hendrikje Liebing, Kai Zittermann, Armin Schubert, Harald Murray, Edward Renner, Andre Gummert, Jan Börgermann, Jochen BMC Anesthesiol Research Article BACKGROUND: Because of its low rate of clinical complications, miniaturized extracorporeal perfusion systems (MEPS) are frequently used in heart centers worldwide. However, many recent studies refer to the higher probability of gaseous microemboli formation by MEPS, caused by subzero pressure values. This is the main reason why various de-airing devices were developed for today’s perfusion systems. In the present study, we investigated the potential benefits of a simple one-way-valve connected to a volume replacement reservoir (OVR) for volume and pressure compensation. METHODS: In an experimental study on 26 pigs, we compared MEPS (n = 13) with MEPS plus OVR (n = 13). Except OVR, perfusion equipment was identical in both groups. Primary endpoints were pressure values in the venous line and the right atrium as well as the number and volume of air bubbles. Secondary endpoints were biochemical parameters of systemic inflammatory response, ischemia, hemodilution and hemolysis. RESULTS: One animal was lost in the MEPS + OVR group. In the MEPS + OVR group no pressure values below −150 mmHg in the venous line and no values under -100 mmHg in right atrium were noticed. On the contrary, nearly 20 % of venous pressure values in the MEPS group were below −150 and approximately 10 % of right atrial pressure values were below -100 mmHg. Compared with the MEPS group, the bubble counter device showed lower numbers of arterial air bubbles in the MEPS + OVR group (mean ± SD: 13444 ± 5709 vs. 1 ± 2, respectively; p < 0.001). In addition, bubble volume was significantly lower in the MEPS + OVR group than in the MEPS group (mean ± SD: 1522 ± 654 μl vs. 4 ± 6 μl, respectively; p < 0.001). The proinflammatory cytokine interleukin-6 and biochemical indices of cardiac ischemia (creatine kinase, and troponin I) were comparable between both groups. CONCLUSIONS: The use of a miniaturized perfusion system with a volume replacement reservoir is able to counteract excessive negative venous line pressures and to reduce the number and volume of arterial air bubbles. This approach may lead to a lower rate of neurological complications. BioMed Central 2015-05-29 /pmc/articles/PMC4459480/ /pubmed/26021999 http://dx.doi.org/10.1186/s12871-015-0058-0 Text en © Aboud et al. 2015 This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
spellingShingle Research Article
Aboud, Anas
Mederos-Dahms, Hendrikje
Liebing, Kai
Zittermann, Armin
Schubert, Harald
Murray, Edward
Renner, Andre
Gummert, Jan
Börgermann, Jochen
Counteracting negative venous line pressures to avoid arterial air bubbles: an experimental study comparing two different types of miniaturized extracorporeal perfusion systems
title Counteracting negative venous line pressures to avoid arterial air bubbles: an experimental study comparing two different types of miniaturized extracorporeal perfusion systems
title_full Counteracting negative venous line pressures to avoid arterial air bubbles: an experimental study comparing two different types of miniaturized extracorporeal perfusion systems
title_fullStr Counteracting negative venous line pressures to avoid arterial air bubbles: an experimental study comparing two different types of miniaturized extracorporeal perfusion systems
title_full_unstemmed Counteracting negative venous line pressures to avoid arterial air bubbles: an experimental study comparing two different types of miniaturized extracorporeal perfusion systems
title_short Counteracting negative venous line pressures to avoid arterial air bubbles: an experimental study comparing two different types of miniaturized extracorporeal perfusion systems
title_sort counteracting negative venous line pressures to avoid arterial air bubbles: an experimental study comparing two different types of miniaturized extracorporeal perfusion systems
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4459480/
https://www.ncbi.nlm.nih.gov/pubmed/26021999
http://dx.doi.org/10.1186/s12871-015-0058-0
work_keys_str_mv AT aboudanas counteractingnegativevenouslinepressurestoavoidarterialairbubblesanexperimentalstudycomparingtwodifferenttypesofminiaturizedextracorporealperfusionsystems
AT mederosdahmshendrikje counteractingnegativevenouslinepressurestoavoidarterialairbubblesanexperimentalstudycomparingtwodifferenttypesofminiaturizedextracorporealperfusionsystems
AT liebingkai counteractingnegativevenouslinepressurestoavoidarterialairbubblesanexperimentalstudycomparingtwodifferenttypesofminiaturizedextracorporealperfusionsystems
AT zittermannarmin counteractingnegativevenouslinepressurestoavoidarterialairbubblesanexperimentalstudycomparingtwodifferenttypesofminiaturizedextracorporealperfusionsystems
AT schubertharald counteractingnegativevenouslinepressurestoavoidarterialairbubblesanexperimentalstudycomparingtwodifferenttypesofminiaturizedextracorporealperfusionsystems
AT murrayedward counteractingnegativevenouslinepressurestoavoidarterialairbubblesanexperimentalstudycomparingtwodifferenttypesofminiaturizedextracorporealperfusionsystems
AT rennerandre counteractingnegativevenouslinepressurestoavoidarterialairbubblesanexperimentalstudycomparingtwodifferenttypesofminiaturizedextracorporealperfusionsystems
AT gummertjan counteractingnegativevenouslinepressurestoavoidarterialairbubblesanexperimentalstudycomparingtwodifferenttypesofminiaturizedextracorporealperfusionsystems
AT borgermannjochen counteractingnegativevenouslinepressurestoavoidarterialairbubblesanexperimentalstudycomparingtwodifferenttypesofminiaturizedextracorporealperfusionsystems