Validation of Varian’s SmartAdapt® deformable image registration algorithm for clinical application
BACKGROUND: Re-contouring of structures on consecutive planning computed tomography (CT) images for patients that exhibit anatomical changes is elaborate and may negatively impact the turn-around time if this is required for many patients. This study was therefore initiated to validate the accuracy...
Autores principales: | , , , , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
BioMed Central
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4465143/ https://www.ncbi.nlm.nih.gov/pubmed/25889772 http://dx.doi.org/10.1186/s13014-015-0372-1 |
Sumario: | BACKGROUND: Re-contouring of structures on consecutive planning computed tomography (CT) images for patients that exhibit anatomical changes is elaborate and may negatively impact the turn-around time if this is required for many patients. This study was therefore initiated to validate the accuracy and usefulness of automatic contour propagation for head and neck cancer patients using SmartAdapt® which is the deformable image registration (DIR) application in Varian’s Eclipse™ treatment planning system. METHODS: CT images of eight head and neck cancer patients with multiple planning CTs were registered using SmartAdapt®. The contoured structures of target volumes and OARs of the primary planning CT were deformed accordingly and subsequently compared with a reference structure set being either: 1) a structure set independently contoured by the treating Radiation Oncologist (RO), or 2) the DIR-generated structure set after being reviewed and modified by the RO. RESULTS: Application of DIR offered a considerable time saving for ROs in delineation of structures on CTs that were acquired mid-treatment. Quantitative analysis showed that 84% of the volume of the DIR-generated structures overlapped with the independently re-contoured structures, while 94% of the volume overlapped with the DIR-generated structures after review by the RO. This apparent intra-observer variation was further investigated resulting in the identification of several causes. Qualitative analysis showed that 92% of the DIR-generated structures either need no or only minor modification during RO reviews. CONCLUSIONS: SmartAdapt is a powerful tool with sufficient accuracy that saves considerable time in re-contouring structures on re-CTs. However, careful review of the DIR-generated structures is mandatory, in particular in areas where tumour regression plays a role. |
---|