Mean Expected Error in Prediction of Total Body Water: A True Accuracy Comparison between Bioimpedance Spectroscopy and Single Frequency Regression Equations

For several decades electrical bioimpedance (EBI) has been used to assess body fluid distribution and body composition. Despite the development of several different approaches for assessing total body water (TBW), it remains uncertain whether bioimpedance spectroscopic (BIS) approaches are more accu...

Descripción completa

Detalles Bibliográficos
Autores principales: Seoane, Fernando, Abtahi, Shirin, Abtahi, Farhad, Ellegård, Lars, Johannsson, Gudmundur, Bosaeus, Ingvar, Ward, Leigh C.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hindawi Publishing Corporation 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4468285/
https://www.ncbi.nlm.nih.gov/pubmed/26137489
http://dx.doi.org/10.1155/2015/656323
_version_ 1782376482929115136
author Seoane, Fernando
Abtahi, Shirin
Abtahi, Farhad
Ellegård, Lars
Johannsson, Gudmundur
Bosaeus, Ingvar
Ward, Leigh C.
author_facet Seoane, Fernando
Abtahi, Shirin
Abtahi, Farhad
Ellegård, Lars
Johannsson, Gudmundur
Bosaeus, Ingvar
Ward, Leigh C.
author_sort Seoane, Fernando
collection PubMed
description For several decades electrical bioimpedance (EBI) has been used to assess body fluid distribution and body composition. Despite the development of several different approaches for assessing total body water (TBW), it remains uncertain whether bioimpedance spectroscopic (BIS) approaches are more accurate than single frequency regression equations. The main objective of this study was to answer this question by calculating the expected accuracy of a single measurement for different EBI methods. The results of this study showed that all methods produced similarly high correlation and concordance coefficients, indicating good accuracy as a method. Even the limits of agreement produced from the Bland-Altman analysis indicated that the performance of single frequency, Sun's prediction equations, at population level was close to the performance of both BIS methods; however, when comparing the Mean Absolute Percentage Error value between the single frequency prediction equations and the BIS methods, a significant difference was obtained, indicating slightly better accuracy for the BIS methods. Despite the higher accuracy of BIS methods over 50 kHz prediction equations at both population and individual level, the magnitude of the improvement was small. Such slight improvement in accuracy of BIS methods is suggested insufficient to warrant their clinical use where the most accurate predictions of TBW are required, for example, when assessing over-fluidic status on dialysis. To reach expected errors below 4-5%, novel and individualized approaches must be developed to improve the accuracy of bioimpedance-based methods for the advent of innovative personalized health monitoring applications.
format Online
Article
Text
id pubmed-4468285
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Hindawi Publishing Corporation
record_format MEDLINE/PubMed
spelling pubmed-44682852015-07-01 Mean Expected Error in Prediction of Total Body Water: A True Accuracy Comparison between Bioimpedance Spectroscopy and Single Frequency Regression Equations Seoane, Fernando Abtahi, Shirin Abtahi, Farhad Ellegård, Lars Johannsson, Gudmundur Bosaeus, Ingvar Ward, Leigh C. Biomed Res Int Research Article For several decades electrical bioimpedance (EBI) has been used to assess body fluid distribution and body composition. Despite the development of several different approaches for assessing total body water (TBW), it remains uncertain whether bioimpedance spectroscopic (BIS) approaches are more accurate than single frequency regression equations. The main objective of this study was to answer this question by calculating the expected accuracy of a single measurement for different EBI methods. The results of this study showed that all methods produced similarly high correlation and concordance coefficients, indicating good accuracy as a method. Even the limits of agreement produced from the Bland-Altman analysis indicated that the performance of single frequency, Sun's prediction equations, at population level was close to the performance of both BIS methods; however, when comparing the Mean Absolute Percentage Error value between the single frequency prediction equations and the BIS methods, a significant difference was obtained, indicating slightly better accuracy for the BIS methods. Despite the higher accuracy of BIS methods over 50 kHz prediction equations at both population and individual level, the magnitude of the improvement was small. Such slight improvement in accuracy of BIS methods is suggested insufficient to warrant their clinical use where the most accurate predictions of TBW are required, for example, when assessing over-fluidic status on dialysis. To reach expected errors below 4-5%, novel and individualized approaches must be developed to improve the accuracy of bioimpedance-based methods for the advent of innovative personalized health monitoring applications. Hindawi Publishing Corporation 2015 2015-06-02 /pmc/articles/PMC4468285/ /pubmed/26137489 http://dx.doi.org/10.1155/2015/656323 Text en Copyright © 2015 Fernando Seoane et al. https://creativecommons.org/licenses/by/3.0/ This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Research Article
Seoane, Fernando
Abtahi, Shirin
Abtahi, Farhad
Ellegård, Lars
Johannsson, Gudmundur
Bosaeus, Ingvar
Ward, Leigh C.
Mean Expected Error in Prediction of Total Body Water: A True Accuracy Comparison between Bioimpedance Spectroscopy and Single Frequency Regression Equations
title Mean Expected Error in Prediction of Total Body Water: A True Accuracy Comparison between Bioimpedance Spectroscopy and Single Frequency Regression Equations
title_full Mean Expected Error in Prediction of Total Body Water: A True Accuracy Comparison between Bioimpedance Spectroscopy and Single Frequency Regression Equations
title_fullStr Mean Expected Error in Prediction of Total Body Water: A True Accuracy Comparison between Bioimpedance Spectroscopy and Single Frequency Regression Equations
title_full_unstemmed Mean Expected Error in Prediction of Total Body Water: A True Accuracy Comparison between Bioimpedance Spectroscopy and Single Frequency Regression Equations
title_short Mean Expected Error in Prediction of Total Body Water: A True Accuracy Comparison between Bioimpedance Spectroscopy and Single Frequency Regression Equations
title_sort mean expected error in prediction of total body water: a true accuracy comparison between bioimpedance spectroscopy and single frequency regression equations
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4468285/
https://www.ncbi.nlm.nih.gov/pubmed/26137489
http://dx.doi.org/10.1155/2015/656323
work_keys_str_mv AT seoanefernando meanexpectederrorinpredictionoftotalbodywateratrueaccuracycomparisonbetweenbioimpedancespectroscopyandsinglefrequencyregressionequations
AT abtahishirin meanexpectederrorinpredictionoftotalbodywateratrueaccuracycomparisonbetweenbioimpedancespectroscopyandsinglefrequencyregressionequations
AT abtahifarhad meanexpectederrorinpredictionoftotalbodywateratrueaccuracycomparisonbetweenbioimpedancespectroscopyandsinglefrequencyregressionequations
AT ellegardlars meanexpectederrorinpredictionoftotalbodywateratrueaccuracycomparisonbetweenbioimpedancespectroscopyandsinglefrequencyregressionequations
AT johannssongudmundur meanexpectederrorinpredictionoftotalbodywateratrueaccuracycomparisonbetweenbioimpedancespectroscopyandsinglefrequencyregressionequations
AT bosaeusingvar meanexpectederrorinpredictionoftotalbodywateratrueaccuracycomparisonbetweenbioimpedancespectroscopyandsinglefrequencyregressionequations
AT wardleighc meanexpectederrorinpredictionoftotalbodywateratrueaccuracycomparisonbetweenbioimpedancespectroscopyandsinglefrequencyregressionequations