Effect of anode/filter combination on average glandular dose in mammography
A comparative analysis of the mean glandular doses was conducted in 100 female patients who underwent screening mammography in 2011 and 2013. Siemens Mammomat Novation with the application of the W/Rh anode/filter combination was used in 2011, whereas in 2013 anode/filter combination was Mo/Mo or Mo...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Medknow Publications & Media Pvt Ltd
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4471644/ https://www.ncbi.nlm.nih.gov/pubmed/26150687 http://dx.doi.org/10.4103/0971-6203.152251 |
_version_ | 1782376946503516160 |
---|---|
author | Biegała, Michał Jakubowska, Teresa Markowska, Karolina |
author_facet | Biegała, Michał Jakubowska, Teresa Markowska, Karolina |
author_sort | Biegała, Michał |
collection | PubMed |
description | A comparative analysis of the mean glandular doses was conducted in 100 female patients who underwent screening mammography in 2011 and 2013. Siemens Mammomat Novation with the application of the W/Rh anode/filter combination was used in 2011, whereas in 2013 anode/filter combination was Mo/Mo or Mo/Rh. The functioning of mammography was checked and the effectiveness of the automatic exposure control (AEC) system was verified by measuring compensation of changes in the phantom thickness and measuring tube voltage. On the base of exposure parameters, an average glandular dose for each of 100 female patients was estimated. The images obtained by using AEC system had the acceptable threshold contrast visibility irrespective of the applied anode/filter combination. Mean glandular doses in the females, examined with the application of the W/Rh anode/filter combination, were on average 23.6% lower than that of the Mo/Mo or Mo/Rh anode/filter combinations. It is recommended to use a combination of the W/Rh anode /filter which exhibited lower mean glandular doses. |
format | Online Article Text |
id | pubmed-4471644 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | Medknow Publications & Media Pvt Ltd |
record_format | MEDLINE/PubMed |
spelling | pubmed-44716442015-07-06 Effect of anode/filter combination on average glandular dose in mammography Biegała, Michał Jakubowska, Teresa Markowska, Karolina J Med Phys Technical Note A comparative analysis of the mean glandular doses was conducted in 100 female patients who underwent screening mammography in 2011 and 2013. Siemens Mammomat Novation with the application of the W/Rh anode/filter combination was used in 2011, whereas in 2013 anode/filter combination was Mo/Mo or Mo/Rh. The functioning of mammography was checked and the effectiveness of the automatic exposure control (AEC) system was verified by measuring compensation of changes in the phantom thickness and measuring tube voltage. On the base of exposure parameters, an average glandular dose for each of 100 female patients was estimated. The images obtained by using AEC system had the acceptable threshold contrast visibility irrespective of the applied anode/filter combination. Mean glandular doses in the females, examined with the application of the W/Rh anode/filter combination, were on average 23.6% lower than that of the Mo/Mo or Mo/Rh anode/filter combinations. It is recommended to use a combination of the W/Rh anode /filter which exhibited lower mean glandular doses. Medknow Publications & Media Pvt Ltd 2015 /pmc/articles/PMC4471644/ /pubmed/26150687 http://dx.doi.org/10.4103/0971-6203.152251 Text en Copyright: © Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Technical Note Biegała, Michał Jakubowska, Teresa Markowska, Karolina Effect of anode/filter combination on average glandular dose in mammography |
title | Effect of anode/filter combination on average glandular dose in mammography |
title_full | Effect of anode/filter combination on average glandular dose in mammography |
title_fullStr | Effect of anode/filter combination on average glandular dose in mammography |
title_full_unstemmed | Effect of anode/filter combination on average glandular dose in mammography |
title_short | Effect of anode/filter combination on average glandular dose in mammography |
title_sort | effect of anode/filter combination on average glandular dose in mammography |
topic | Technical Note |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4471644/ https://www.ncbi.nlm.nih.gov/pubmed/26150687 http://dx.doi.org/10.4103/0971-6203.152251 |
work_keys_str_mv | AT biegałamichał effectofanodefiltercombinationonaverageglandulardoseinmammography AT jakubowskateresa effectofanodefiltercombinationonaverageglandulardoseinmammography AT markowskakarolina effectofanodefiltercombinationonaverageglandulardoseinmammography |