Small fields measurements with radiochromic films

The small fields in radiotherapy are widely used due to the development of techniques such as intensity-modulated radiotherapy and stereotactic radio surgery. The measurement of the dose distributions for small fields is a challenge. A perfect dosimeter should be independent of the radiation energy...

Descripción completa

Detalles Bibliográficos
Autores principales: Gonzalez-Lopez, Antonio, Vera-Sanchez, Juan-Antonio, Lago-Martin, Jose-Domingo
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Medknow Publications & Media Pvt Ltd 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4478646/
https://www.ncbi.nlm.nih.gov/pubmed/26170551
http://dx.doi.org/10.4103/0971-6203.158667
_version_ 1782377917645324288
author Gonzalez-Lopez, Antonio
Vera-Sanchez, Juan-Antonio
Lago-Martin, Jose-Domingo
author_facet Gonzalez-Lopez, Antonio
Vera-Sanchez, Juan-Antonio
Lago-Martin, Jose-Domingo
author_sort Gonzalez-Lopez, Antonio
collection PubMed
description The small fields in radiotherapy are widely used due to the development of techniques such as intensity-modulated radiotherapy and stereotactic radio surgery. The measurement of the dose distributions for small fields is a challenge. A perfect dosimeter should be independent of the radiation energy and the dose rate and should have a negligible volume effect. The radiochromic (RC) film characteristics fit well to these requirements. However, the response of RC films and their digitizing processes present a significant spatial inhomogeneity problem. The present work uses a method for two-dimensional (2D) measurement with RC films based on the reduction of the spatial inhomogeneity of both the film and the film digitizing process. By means of registering and averaging several measurements of the same field, the inhomogeneities are mostly canceled. Measurements of output factors (OFs), dose profiles (in-plane and cross-plane), and 2D dose distributions are presented. The field sizes investigated are 0.5 × 0.5 cm(2), 0.7 × 0.7 cm(2), 1 × 1 cm(2), 2 × 2 cm(2), 3 × 3 cm(2), 6 × 6 cm(2), and 10 × 10 cm(2) for 6 and 15 MV photon beams. The OFs measured with the RC film are compared with the measurements carried out with a PinPoint ionization chamber (IC) and a Semiflex IC, while the measured transversal dose profiles were compared with Monte Carlo simulations. The results obtained for the OFs measurements show a good agreement with the values obtained from RC films and the PinPoint and Semiflex chambers when the field size is greater or equal than 2 × 2 cm(2). These agreements give confidence on the accuracy of the method as well as on the results obtained for smaller fields. Also, good agreement was found between the measured profiles and the Monte Carlo calculated profiles for the field size of 1 × 1 cm(2). We expect, therefore, that the presented method can be used to perform accurate measurements of small fields.
format Online
Article
Text
id pubmed-4478646
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Medknow Publications & Media Pvt Ltd
record_format MEDLINE/PubMed
spelling pubmed-44786462015-07-13 Small fields measurements with radiochromic films Gonzalez-Lopez, Antonio Vera-Sanchez, Juan-Antonio Lago-Martin, Jose-Domingo J Med Phys Original Article The small fields in radiotherapy are widely used due to the development of techniques such as intensity-modulated radiotherapy and stereotactic radio surgery. The measurement of the dose distributions for small fields is a challenge. A perfect dosimeter should be independent of the radiation energy and the dose rate and should have a negligible volume effect. The radiochromic (RC) film characteristics fit well to these requirements. However, the response of RC films and their digitizing processes present a significant spatial inhomogeneity problem. The present work uses a method for two-dimensional (2D) measurement with RC films based on the reduction of the spatial inhomogeneity of both the film and the film digitizing process. By means of registering and averaging several measurements of the same field, the inhomogeneities are mostly canceled. Measurements of output factors (OFs), dose profiles (in-plane and cross-plane), and 2D dose distributions are presented. The field sizes investigated are 0.5 × 0.5 cm(2), 0.7 × 0.7 cm(2), 1 × 1 cm(2), 2 × 2 cm(2), 3 × 3 cm(2), 6 × 6 cm(2), and 10 × 10 cm(2) for 6 and 15 MV photon beams. The OFs measured with the RC film are compared with the measurements carried out with a PinPoint ionization chamber (IC) and a Semiflex IC, while the measured transversal dose profiles were compared with Monte Carlo simulations. The results obtained for the OFs measurements show a good agreement with the values obtained from RC films and the PinPoint and Semiflex chambers when the field size is greater or equal than 2 × 2 cm(2). These agreements give confidence on the accuracy of the method as well as on the results obtained for smaller fields. Also, good agreement was found between the measured profiles and the Monte Carlo calculated profiles for the field size of 1 × 1 cm(2). We expect, therefore, that the presented method can be used to perform accurate measurements of small fields. Medknow Publications & Media Pvt Ltd 2015 /pmc/articles/PMC4478646/ /pubmed/26170551 http://dx.doi.org/10.4103/0971-6203.158667 Text en Copyright: © Journal of Medical Physics http://creativecommons.org/licenses/by-nc-sa/3.0 This is an open-access article distributed under the terms of the Creative Commons Attribution-Noncommercial-Share Alike 3.0 Unported, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Original Article
Gonzalez-Lopez, Antonio
Vera-Sanchez, Juan-Antonio
Lago-Martin, Jose-Domingo
Small fields measurements with radiochromic films
title Small fields measurements with radiochromic films
title_full Small fields measurements with radiochromic films
title_fullStr Small fields measurements with radiochromic films
title_full_unstemmed Small fields measurements with radiochromic films
title_short Small fields measurements with radiochromic films
title_sort small fields measurements with radiochromic films
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4478646/
https://www.ncbi.nlm.nih.gov/pubmed/26170551
http://dx.doi.org/10.4103/0971-6203.158667
work_keys_str_mv AT gonzalezlopezantonio smallfieldsmeasurementswithradiochromicfilms
AT verasanchezjuanantonio smallfieldsmeasurementswithradiochromicfilms
AT lagomartinjosedomingo smallfieldsmeasurementswithradiochromicfilms