A multicentre comparison of quantitative (90)Y PET/CT for dosimetric purposes after radioembolization with resin microspheres: The QUEST Phantom Study

PURPOSE: To investigate and compare the quantitative accuracy of (90)Y imaging across different generation PET/CT scanners, for the purpose of dosimetry after radioembolization with resin microspheres. METHODS: A strict experimental and imaging protocol was followed by 47 international sites using t...

Descripción completa

Detalles Bibliográficos
Autores principales: Willowson, Kathy P., Tapner, Michael, Bailey, Dale L.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Berlin Heidelberg 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4480824/
https://www.ncbi.nlm.nih.gov/pubmed/25967868
http://dx.doi.org/10.1007/s00259-015-3059-9
_version_ 1782378196140818432
author Willowson, Kathy P.
Tapner, Michael
Bailey, Dale L.
author_facet Willowson, Kathy P.
Tapner, Michael
Bailey, Dale L.
author_sort Willowson, Kathy P.
collection PubMed
description PURPOSE: To investigate and compare the quantitative accuracy of (90)Y imaging across different generation PET/CT scanners, for the purpose of dosimetry after radioembolization with resin microspheres. METHODS: A strict experimental and imaging protocol was followed by 47 international sites using the NEMA 2007/IEC 2008 PET body phantom with an 8-to-1 sphere-to-background ratio of (90)Y solution. The phantom was imaged over a 7-day period (activity ranging from 0.5 to 3.0 GBq) and all reconstructed data were analysed at a core laboratory for consistent processing. Quantitative accuracy was assessed through measures of total phantom activity, activity concentration in background and hot spheres, misplaced counts in a nonradioactive insert, and background variability. RESULTS: Of the 69 scanners assessed, 37 had both time-of-flight (ToF) and resolution recovery (RR) capability. These current generation scanners from GE, Philips and Siemens could reconstruct background concentration measures to within 10 % of true values over the evaluated range, with greater deviations on the Philips systems at low count rates, and demonstrated typical partial volume effects on hot sphere recovery, which dominated spheres of diameter <20 mm. For spheres >20 mm in diameter, activity concentrations were consistently underestimated by about 20 %. Non-ToF scanners from GE Healthcare and Siemens were capable of producing accurate measures, but with inferior quantitative recovery compared with ToF systems. CONCLUSION: Current generation ToF scanners can consistently reconstruct (90)Y activity concentrations, but they underestimate activity concentrations in small structures (≤37 mm diameter) within a warm background due to partial volume effects and constraints of the reconstruction algorithm. At the highest count rates investigated, measures of background concentration (about 300 kBq/ml) could be estimated on average to within 1 %, 5 % and 2 % for GE Healthcare (all-pass filter, RR + ToF), Philips (4i8s ToF) and Siemens (2i21s all-pass filter, RR + ToF) ToF systems, respectively. Over the range of activities investigated, comparable performance between GE Healthcare and Siemens ToF systems suggests suitability for quantitative analysis in a scenario analogous to that of postradioembolization imaging for treatment of liver cancer.
format Online
Article
Text
id pubmed-4480824
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Springer Berlin Heidelberg
record_format MEDLINE/PubMed
spelling pubmed-44808242015-07-02 A multicentre comparison of quantitative (90)Y PET/CT for dosimetric purposes after radioembolization with resin microspheres: The QUEST Phantom Study Willowson, Kathy P. Tapner, Michael Bailey, Dale L. Eur J Nucl Med Mol Imaging Original Article PURPOSE: To investigate and compare the quantitative accuracy of (90)Y imaging across different generation PET/CT scanners, for the purpose of dosimetry after radioembolization with resin microspheres. METHODS: A strict experimental and imaging protocol was followed by 47 international sites using the NEMA 2007/IEC 2008 PET body phantom with an 8-to-1 sphere-to-background ratio of (90)Y solution. The phantom was imaged over a 7-day period (activity ranging from 0.5 to 3.0 GBq) and all reconstructed data were analysed at a core laboratory for consistent processing. Quantitative accuracy was assessed through measures of total phantom activity, activity concentration in background and hot spheres, misplaced counts in a nonradioactive insert, and background variability. RESULTS: Of the 69 scanners assessed, 37 had both time-of-flight (ToF) and resolution recovery (RR) capability. These current generation scanners from GE, Philips and Siemens could reconstruct background concentration measures to within 10 % of true values over the evaluated range, with greater deviations on the Philips systems at low count rates, and demonstrated typical partial volume effects on hot sphere recovery, which dominated spheres of diameter <20 mm. For spheres >20 mm in diameter, activity concentrations were consistently underestimated by about 20 %. Non-ToF scanners from GE Healthcare and Siemens were capable of producing accurate measures, but with inferior quantitative recovery compared with ToF systems. CONCLUSION: Current generation ToF scanners can consistently reconstruct (90)Y activity concentrations, but they underestimate activity concentrations in small structures (≤37 mm diameter) within a warm background due to partial volume effects and constraints of the reconstruction algorithm. At the highest count rates investigated, measures of background concentration (about 300 kBq/ml) could be estimated on average to within 1 %, 5 % and 2 % for GE Healthcare (all-pass filter, RR + ToF), Philips (4i8s ToF) and Siemens (2i21s all-pass filter, RR + ToF) ToF systems, respectively. Over the range of activities investigated, comparable performance between GE Healthcare and Siemens ToF systems suggests suitability for quantitative analysis in a scenario analogous to that of postradioembolization imaging for treatment of liver cancer. Springer Berlin Heidelberg 2015-05-13 2015 /pmc/articles/PMC4480824/ /pubmed/25967868 http://dx.doi.org/10.1007/s00259-015-3059-9 Text en © The Author(s) 2015 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.
spellingShingle Original Article
Willowson, Kathy P.
Tapner, Michael
Bailey, Dale L.
A multicentre comparison of quantitative (90)Y PET/CT for dosimetric purposes after radioembolization with resin microspheres: The QUEST Phantom Study
title A multicentre comparison of quantitative (90)Y PET/CT for dosimetric purposes after radioembolization with resin microspheres: The QUEST Phantom Study
title_full A multicentre comparison of quantitative (90)Y PET/CT for dosimetric purposes after radioembolization with resin microspheres: The QUEST Phantom Study
title_fullStr A multicentre comparison of quantitative (90)Y PET/CT for dosimetric purposes after radioembolization with resin microspheres: The QUEST Phantom Study
title_full_unstemmed A multicentre comparison of quantitative (90)Y PET/CT for dosimetric purposes after radioembolization with resin microspheres: The QUEST Phantom Study
title_short A multicentre comparison of quantitative (90)Y PET/CT for dosimetric purposes after radioembolization with resin microspheres: The QUEST Phantom Study
title_sort multicentre comparison of quantitative (90)y pet/ct for dosimetric purposes after radioembolization with resin microspheres: the quest phantom study
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4480824/
https://www.ncbi.nlm.nih.gov/pubmed/25967868
http://dx.doi.org/10.1007/s00259-015-3059-9
work_keys_str_mv AT willowsonkathyp amulticentrecomparisonofquantitative90ypetctfordosimetricpurposesafterradioembolizationwithresinmicrospheresthequestphantomstudy
AT tapnermichael amulticentrecomparisonofquantitative90ypetctfordosimetricpurposesafterradioembolizationwithresinmicrospheresthequestphantomstudy
AT amulticentrecomparisonofquantitative90ypetctfordosimetricpurposesafterradioembolizationwithresinmicrospheresthequestphantomstudy
AT baileydalel amulticentrecomparisonofquantitative90ypetctfordosimetricpurposesafterradioembolizationwithresinmicrospheresthequestphantomstudy
AT willowsonkathyp multicentrecomparisonofquantitative90ypetctfordosimetricpurposesafterradioembolizationwithresinmicrospheresthequestphantomstudy
AT tapnermichael multicentrecomparisonofquantitative90ypetctfordosimetricpurposesafterradioembolizationwithresinmicrospheresthequestphantomstudy
AT multicentrecomparisonofquantitative90ypetctfordosimetricpurposesafterradioembolizationwithresinmicrospheresthequestphantomstudy
AT baileydalel multicentrecomparisonofquantitative90ypetctfordosimetricpurposesafterradioembolizationwithresinmicrospheresthequestphantomstudy