A multicentre comparison of quantitative (90)Y PET/CT for dosimetric purposes after radioembolization with resin microspheres: The QUEST Phantom Study
PURPOSE: To investigate and compare the quantitative accuracy of (90)Y imaging across different generation PET/CT scanners, for the purpose of dosimetry after radioembolization with resin microspheres. METHODS: A strict experimental and imaging protocol was followed by 47 international sites using t...
Autores principales: | , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Berlin Heidelberg
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4480824/ https://www.ncbi.nlm.nih.gov/pubmed/25967868 http://dx.doi.org/10.1007/s00259-015-3059-9 |
_version_ | 1782378196140818432 |
---|---|
author | Willowson, Kathy P. Tapner, Michael Bailey, Dale L. |
author_facet | Willowson, Kathy P. Tapner, Michael Bailey, Dale L. |
author_sort | Willowson, Kathy P. |
collection | PubMed |
description | PURPOSE: To investigate and compare the quantitative accuracy of (90)Y imaging across different generation PET/CT scanners, for the purpose of dosimetry after radioembolization with resin microspheres. METHODS: A strict experimental and imaging protocol was followed by 47 international sites using the NEMA 2007/IEC 2008 PET body phantom with an 8-to-1 sphere-to-background ratio of (90)Y solution. The phantom was imaged over a 7-day period (activity ranging from 0.5 to 3.0 GBq) and all reconstructed data were analysed at a core laboratory for consistent processing. Quantitative accuracy was assessed through measures of total phantom activity, activity concentration in background and hot spheres, misplaced counts in a nonradioactive insert, and background variability. RESULTS: Of the 69 scanners assessed, 37 had both time-of-flight (ToF) and resolution recovery (RR) capability. These current generation scanners from GE, Philips and Siemens could reconstruct background concentration measures to within 10 % of true values over the evaluated range, with greater deviations on the Philips systems at low count rates, and demonstrated typical partial volume effects on hot sphere recovery, which dominated spheres of diameter <20 mm. For spheres >20 mm in diameter, activity concentrations were consistently underestimated by about 20 %. Non-ToF scanners from GE Healthcare and Siemens were capable of producing accurate measures, but with inferior quantitative recovery compared with ToF systems. CONCLUSION: Current generation ToF scanners can consistently reconstruct (90)Y activity concentrations, but they underestimate activity concentrations in small structures (≤37 mm diameter) within a warm background due to partial volume effects and constraints of the reconstruction algorithm. At the highest count rates investigated, measures of background concentration (about 300 kBq/ml) could be estimated on average to within 1 %, 5 % and 2 % for GE Healthcare (all-pass filter, RR + ToF), Philips (4i8s ToF) and Siemens (2i21s all-pass filter, RR + ToF) ToF systems, respectively. Over the range of activities investigated, comparable performance between GE Healthcare and Siemens ToF systems suggests suitability for quantitative analysis in a scenario analogous to that of postradioembolization imaging for treatment of liver cancer. |
format | Online Article Text |
id | pubmed-4480824 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | Springer Berlin Heidelberg |
record_format | MEDLINE/PubMed |
spelling | pubmed-44808242015-07-02 A multicentre comparison of quantitative (90)Y PET/CT for dosimetric purposes after radioembolization with resin microspheres: The QUEST Phantom Study Willowson, Kathy P. Tapner, Michael Bailey, Dale L. Eur J Nucl Med Mol Imaging Original Article PURPOSE: To investigate and compare the quantitative accuracy of (90)Y imaging across different generation PET/CT scanners, for the purpose of dosimetry after radioembolization with resin microspheres. METHODS: A strict experimental and imaging protocol was followed by 47 international sites using the NEMA 2007/IEC 2008 PET body phantom with an 8-to-1 sphere-to-background ratio of (90)Y solution. The phantom was imaged over a 7-day period (activity ranging from 0.5 to 3.0 GBq) and all reconstructed data were analysed at a core laboratory for consistent processing. Quantitative accuracy was assessed through measures of total phantom activity, activity concentration in background and hot spheres, misplaced counts in a nonradioactive insert, and background variability. RESULTS: Of the 69 scanners assessed, 37 had both time-of-flight (ToF) and resolution recovery (RR) capability. These current generation scanners from GE, Philips and Siemens could reconstruct background concentration measures to within 10 % of true values over the evaluated range, with greater deviations on the Philips systems at low count rates, and demonstrated typical partial volume effects on hot sphere recovery, which dominated spheres of diameter <20 mm. For spheres >20 mm in diameter, activity concentrations were consistently underestimated by about 20 %. Non-ToF scanners from GE Healthcare and Siemens were capable of producing accurate measures, but with inferior quantitative recovery compared with ToF systems. CONCLUSION: Current generation ToF scanners can consistently reconstruct (90)Y activity concentrations, but they underestimate activity concentrations in small structures (≤37 mm diameter) within a warm background due to partial volume effects and constraints of the reconstruction algorithm. At the highest count rates investigated, measures of background concentration (about 300 kBq/ml) could be estimated on average to within 1 %, 5 % and 2 % for GE Healthcare (all-pass filter, RR + ToF), Philips (4i8s ToF) and Siemens (2i21s all-pass filter, RR + ToF) ToF systems, respectively. Over the range of activities investigated, comparable performance between GE Healthcare and Siemens ToF systems suggests suitability for quantitative analysis in a scenario analogous to that of postradioembolization imaging for treatment of liver cancer. Springer Berlin Heidelberg 2015-05-13 2015 /pmc/articles/PMC4480824/ /pubmed/25967868 http://dx.doi.org/10.1007/s00259-015-3059-9 Text en © The Author(s) 2015 Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. |
spellingShingle | Original Article Willowson, Kathy P. Tapner, Michael Bailey, Dale L. A multicentre comparison of quantitative (90)Y PET/CT for dosimetric purposes after radioembolization with resin microspheres: The QUEST Phantom Study |
title | A multicentre comparison of quantitative (90)Y PET/CT for dosimetric purposes after radioembolization with resin microspheres: The QUEST Phantom Study |
title_full | A multicentre comparison of quantitative (90)Y PET/CT for dosimetric purposes after radioembolization with resin microspheres: The QUEST Phantom Study |
title_fullStr | A multicentre comparison of quantitative (90)Y PET/CT for dosimetric purposes after radioembolization with resin microspheres: The QUEST Phantom Study |
title_full_unstemmed | A multicentre comparison of quantitative (90)Y PET/CT for dosimetric purposes after radioembolization with resin microspheres: The QUEST Phantom Study |
title_short | A multicentre comparison of quantitative (90)Y PET/CT for dosimetric purposes after radioembolization with resin microspheres: The QUEST Phantom Study |
title_sort | multicentre comparison of quantitative (90)y pet/ct for dosimetric purposes after radioembolization with resin microspheres: the quest phantom study |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4480824/ https://www.ncbi.nlm.nih.gov/pubmed/25967868 http://dx.doi.org/10.1007/s00259-015-3059-9 |
work_keys_str_mv | AT willowsonkathyp amulticentrecomparisonofquantitative90ypetctfordosimetricpurposesafterradioembolizationwithresinmicrospheresthequestphantomstudy AT tapnermichael amulticentrecomparisonofquantitative90ypetctfordosimetricpurposesafterradioembolizationwithresinmicrospheresthequestphantomstudy AT amulticentrecomparisonofquantitative90ypetctfordosimetricpurposesafterradioembolizationwithresinmicrospheresthequestphantomstudy AT baileydalel amulticentrecomparisonofquantitative90ypetctfordosimetricpurposesafterradioembolizationwithresinmicrospheresthequestphantomstudy AT willowsonkathyp multicentrecomparisonofquantitative90ypetctfordosimetricpurposesafterradioembolizationwithresinmicrospheresthequestphantomstudy AT tapnermichael multicentrecomparisonofquantitative90ypetctfordosimetricpurposesafterradioembolizationwithresinmicrospheresthequestphantomstudy AT multicentrecomparisonofquantitative90ypetctfordosimetricpurposesafterradioembolizationwithresinmicrospheresthequestphantomstudy AT baileydalel multicentrecomparisonofquantitative90ypetctfordosimetricpurposesafterradioembolizationwithresinmicrospheresthequestphantomstudy |