Ultrahigh-resolution Cerenkov-light imaging system for positron radionuclides: potential applications and limitations

OBJECTIVE: Cerenkov-light imaging provides inherently high resolution because the light is emitted near the positron radionuclide. However, the magnitude for the high spatial resolution of Cerenkov-light imaging is unclear. Its potential molecular imaging applications also remain unclear. We develop...

Descripción completa

Detalles Bibliográficos
Autores principales: Yamamoto, Seiichi, Watabe, Tadashi, Ikeda, Hayato, Kanai, Yasukazu, Watabe, Hiroshi, Ogata, Yoshimune, Kato, Katsuhiko, Hatazawa, Jun
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Japan 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4483184/
https://www.ncbi.nlm.nih.gov/pubmed/25103137
http://dx.doi.org/10.1007/s12149-014-0892-z
Descripción
Sumario:OBJECTIVE: Cerenkov-light imaging provides inherently high resolution because the light is emitted near the positron radionuclide. However, the magnitude for the high spatial resolution of Cerenkov-light imaging is unclear. Its potential molecular imaging applications also remain unclear. We developed an ultrahigh-resolution Cerenkov-light imaging system, measured its spatial resolution, and explored its applications to molecular imaging research. METHODS: Our Cerenkov-light imaging system consists of a high-sensitivity charged-coupled device camera (Hamamatsu Photonics ORCA2-ER) and a bright lens (Xenon 0.95/25). An extension ring was inserted between them to magnify the subject. A ~100-μm-diameter (22)Na point source was made and imaged by the system. For applications of Cerenkov-light imaging, we conducted (18)F-FDG administered in vivo, ex vivo whole brain, and sliced brain imaging of rats. RESULTS: We obtained spatial resolution of ~220 μm for a (22)Na point source with our developed imaging system. The (18)F-FDG rat head images showed high light intensity in the eyes for the Cerenkov-light images, although there was no accumulation in these parts in the PET images. The sliced rat brain showed much higher spatial resolution for the Cerenkov-light images compared with CdWO(4) scintillator-based autoradiography, although some contrast decrease was observed for them. CONCLUSION: Even though the Cerenkov-light images showed ultrahigh resolution of ~220 μm, their distribution and contrast were sometimes different from the actual positron accumulation in the subjects. Care must be taken when evaluating positron distribution from Cerenkov-light images. However, the ultrahigh resolution of Cerenkov-light imaging will be useful for transparent subjects including phantom studies.