Mean corpuscular volume of control red blood cells determines the interpretation of eosin-5′-maleimide (EMA) test result in infants aged less than 6 months
Eosin-5′-maleimide (EMA) binding test is a flow cytometric test used to detect hereditary spherocytosis (HS). To perform the test sample from patients, 5–6 reference samples of red blood are needed. Our aim was to investigate how the mean corpuscular volume (MCV) of red blood cells influences on the...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer Berlin Heidelberg
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4488455/ https://www.ncbi.nlm.nih.gov/pubmed/25907377 http://dx.doi.org/10.1007/s00277-015-2377-0 |
Sumario: | Eosin-5′-maleimide (EMA) binding test is a flow cytometric test used to detect hereditary spherocytosis (HS). To perform the test sample from patients, 5–6 reference samples of red blood are needed. Our aim was to investigate how the mean corpuscular volume (MCV) of red blood cells influences on the value of fluorescence of bounded EMA dye and how the choice of reference samples affects the test result. EMA test was performed in peripheral blood from 404 individuals, including 31 children suffering from HS. Mean fluorescence channel of EMA-RBCs was measured with Cytomics FC500 flow cytometer. Mean corpuscular volume of RBCs was assessed with LH750 Beckman Coulter. Statistical analysis was performed using Graph Pad Prism. The correlation Spearman coefficient between mean channel of fluorescence of EMA-RBCs and MCV was r = 0.39, p < 0.0001. Interpretation of EMA test depends on MCV of the reference samples. If reference blood samples have lower MCV than the patients MCV, EMA test result might be negative. Due to different MCV values of RBCs in infancy and ca. Three months later, EMA test in neonates might be interpreted falsely negative. Samples from children younger than 3 months old had EMA test result 86.1 ± 11.7 %, whereas same samples that analyzed 4.1 ± 2.1 later had results of 75.4 ± 4.5 %, p < 0.05. Mean fluorescence of EMA-bound RBC depends on RBC’s volume. MCV of reference samples affects EMA test results; thus, we recommend selection of reference samples with MCV in range of ±2 fL compared to MCV of patient RBC’s. |
---|