PURPOSE: This observational study was designed to evaluate the clinical value of cardiac output (CO) obtained via bioreactance (NICOM™) as compared with values of CO obtained via thermodilution (using pulmonary artery catheter, Vigilance™) and the thoracic bioimpedance (BioZ.com™), in patients under...

Descripción completa

Detalles Bibliográficos
Autores principales: Cheung, Hoiyin, Dong, Quan, Dong, Rong, Yu, Buwei
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Springer Japan 2014
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4488496/
https://www.ncbi.nlm.nih.gov/pubmed/25381090
http://dx.doi.org/10.1007/s00540-014-1938-z
_version_ 1782379169036894208
author Cheung, Hoiyin
Dong, Quan
Dong, Rong
Yu, Buwei
author_facet Cheung, Hoiyin
Dong, Quan
Dong, Rong
Yu, Buwei
author_sort Cheung, Hoiyin
collection PubMed
description PURPOSE: This observational study was designed to evaluate the clinical value of cardiac output (CO) obtained via bioreactance (NICOM™) as compared with values of CO obtained via thermodilution (using pulmonary artery catheter, Vigilance™) and the thoracic bioimpedance (BioZ.com™), in patients undergoing off-pump coronary artery bypass surgery. METHODS: Fifty American Society of Anesthesiologists physical status I–III patients, aged 38–81 years, scheduled for off-pump coronary artery bypass surgery were enrolled in this study. CO data (NCO, BCO, PCO) were recorded during the operative period at ten time points after stable hemodynamic conditions were achieved. RESULTS: The equation of the relationship between the PCO and NCO is PCO = 0.945 × NCO + 0.328 (r = 0.77), and that of PCO and BCO is PCO = 0.965 × BCO + 0.729 (r = 0.63). Furthermore, no statistical difference was found between PCO versus NCO (mean (SD): 4.4 (1.1) versus 4.4 (0.9), p = 0.431). A significant correlation was found between PCO and NCO (r = 0.77, p < 0.001). Correlation was also found between PCO and BCO (r = 0.63, p < 0.001). CONCLUSIONS: The NICOM device is a safe, convenient, and reliable device for measuring continuous non-invasive cardiac output and cardiac index, and the trends of change in CO during the surgery are similar between NICOM and PAC.
format Online
Article
Text
id pubmed-4488496
institution National Center for Biotechnology Information
language English
publishDate 2014
publisher Springer Japan
record_format MEDLINE/PubMed
spelling pubmed-44884962015-07-07 Correlation of cardiac output measured by non-invasive continuous cardiac output monitoring (NICOM) and thermodilution in patients undergoing off-pump coronary artery bypass surgery Cheung, Hoiyin Dong, Quan Dong, Rong Yu, Buwei J Anesth Original Article PURPOSE: This observational study was designed to evaluate the clinical value of cardiac output (CO) obtained via bioreactance (NICOM™) as compared with values of CO obtained via thermodilution (using pulmonary artery catheter, Vigilance™) and the thoracic bioimpedance (BioZ.com™), in patients undergoing off-pump coronary artery bypass surgery. METHODS: Fifty American Society of Anesthesiologists physical status I–III patients, aged 38–81 years, scheduled for off-pump coronary artery bypass surgery were enrolled in this study. CO data (NCO, BCO, PCO) were recorded during the operative period at ten time points after stable hemodynamic conditions were achieved. RESULTS: The equation of the relationship between the PCO and NCO is PCO = 0.945 × NCO + 0.328 (r = 0.77), and that of PCO and BCO is PCO = 0.965 × BCO + 0.729 (r = 0.63). Furthermore, no statistical difference was found between PCO versus NCO (mean (SD): 4.4 (1.1) versus 4.4 (0.9), p = 0.431). A significant correlation was found between PCO and NCO (r = 0.77, p < 0.001). Correlation was also found between PCO and BCO (r = 0.63, p < 0.001). CONCLUSIONS: The NICOM device is a safe, convenient, and reliable device for measuring continuous non-invasive cardiac output and cardiac index, and the trends of change in CO during the surgery are similar between NICOM and PAC. Springer Japan 2014-11-08 2015 /pmc/articles/PMC4488496/ /pubmed/25381090 http://dx.doi.org/10.1007/s00540-014-1938-z Text en © The Author(s) 2014 https://creativecommons.org/licenses/by/4.0/ Open AccessThis article is distributed under the terms of the Creative Commons Attribution License which permits any use, distribution, and reproduction in any medium, provided the original author(s) and the source are credited.
spellingShingle Original Article
Cheung, Hoiyin
Dong, Quan
Dong, Rong
Yu, Buwei
Correlation of cardiac output measured by non-invasive continuous cardiac output monitoring (NICOM) and thermodilution in patients undergoing off-pump coronary artery bypass surgery
title Correlation of cardiac output measured by non-invasive continuous cardiac output monitoring (NICOM) and thermodilution in patients undergoing off-pump coronary artery bypass surgery
title_full Correlation of cardiac output measured by non-invasive continuous cardiac output monitoring (NICOM) and thermodilution in patients undergoing off-pump coronary artery bypass surgery
title_fullStr Correlation of cardiac output measured by non-invasive continuous cardiac output monitoring (NICOM) and thermodilution in patients undergoing off-pump coronary artery bypass surgery
title_full_unstemmed Correlation of cardiac output measured by non-invasive continuous cardiac output monitoring (NICOM) and thermodilution in patients undergoing off-pump coronary artery bypass surgery
title_short Correlation of cardiac output measured by non-invasive continuous cardiac output monitoring (NICOM) and thermodilution in patients undergoing off-pump coronary artery bypass surgery
title_sort correlation of cardiac output measured by non-invasive continuous cardiac output monitoring (nicom) and thermodilution in patients undergoing off-pump coronary artery bypass surgery
topic Original Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4488496/
https://www.ncbi.nlm.nih.gov/pubmed/25381090
http://dx.doi.org/10.1007/s00540-014-1938-z
work_keys_str_mv AT cheunghoiyin correlationofcardiacoutputmeasuredbynoninvasivecontinuouscardiacoutputmonitoringnicomandthermodilutioninpatientsundergoingoffpumpcoronaryarterybypasssurgery
AT dongquan correlationofcardiacoutputmeasuredbynoninvasivecontinuouscardiacoutputmonitoringnicomandthermodilutioninpatientsundergoingoffpumpcoronaryarterybypasssurgery
AT dongrong correlationofcardiacoutputmeasuredbynoninvasivecontinuouscardiacoutputmonitoringnicomandthermodilutioninpatientsundergoingoffpumpcoronaryarterybypasssurgery
AT yubuwei correlationofcardiacoutputmeasuredbynoninvasivecontinuouscardiacoutputmonitoringnicomandthermodilutioninpatientsundergoingoffpumpcoronaryarterybypasssurgery