Artificial disc and vertebra system: a novel motion preservation device for cervical spinal disease after vertebral corpectomy

OBJECTIVE: To determine the range of motion and stability of the human cadaveric cervical spine after the implantation of a novel artificial disc and vertebra system by comparing an intact group and a fusion group. METHODS: Biomechanical tests were conducted on 18 human cadaveric cervical specimens....

Descripción completa

Detalles Bibliográficos
Autores principales: Dong, Jun, Lu, Meng, Lu, Teng, Liang, Baobao, Xu, Junkui, Qin, Jie, Cai, Xuan, Huang, Sihua, Wang, Dong, Li, Haopeng, He, Xijing
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4496753/
https://www.ncbi.nlm.nih.gov/pubmed/26222819
http://dx.doi.org/10.6061/clinics/2015(07)06
_version_ 1782380454596313088
author Dong, Jun
Lu, Meng
Lu, Teng
Liang, Baobao
Xu, Junkui
Qin, Jie
Cai, Xuan
Huang, Sihua
Wang, Dong
Li, Haopeng
He, Xijing
author_facet Dong, Jun
Lu, Meng
Lu, Teng
Liang, Baobao
Xu, Junkui
Qin, Jie
Cai, Xuan
Huang, Sihua
Wang, Dong
Li, Haopeng
He, Xijing
author_sort Dong, Jun
collection PubMed
description OBJECTIVE: To determine the range of motion and stability of the human cadaveric cervical spine after the implantation of a novel artificial disc and vertebra system by comparing an intact group and a fusion group. METHODS: Biomechanical tests were conducted on 18 human cadaveric cervical specimens. The range of motion and the stability index range of motion were measured to study the function and stability of the artificial disc and vertebra system of the intact group compared with the fusion group. RESULTS: In all cases, the artificial disc and vertebra system maintained intervertebral motion and reestablished vertebral height at the operative level. After its implantation, there was no significant difference in the range of motion (ROM) of C3–7 in all directions in the non-fusion group compared with the intact group (p>0.05), but significant differences were detected in flexion, extension and axial rotation compared with the fusion group (p<0.05). The ROM of adjacent segments (C(3−4), C(6−7)) of the non-fusion group decreased significantly in some directions compared with the fusion group (p<0.05). Significant differences in the C(4-6) ROM in some directions were detected between the non-fusion group and the intact group. In the fusion group, the C(4−6) ROM in all directions decreased significantly compared with the intact and non-fusion groups (p<0.01). The stability index ROM (SI-ROM) of some directions was negative in the non-fusion group, and a significant difference in SI-ROM was only found in the C(4−6) segment of the non-fusion group compared with the fusion group. CONCLUSION: An artificial disc and vertebra system could restore vertebral height and preserve the dynamic function of the surgical area and could theoretically reduce the risk of adjacent segment degeneration compared with the anterior fusion procedure. However, our results should be considered with caution because of the low power of the study. The use of a larger sample should be considered in future studies.
format Online
Article
Text
id pubmed-4496753
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo
record_format MEDLINE/PubMed
spelling pubmed-44967532015-07-13 Artificial disc and vertebra system: a novel motion preservation device for cervical spinal disease after vertebral corpectomy Dong, Jun Lu, Meng Lu, Teng Liang, Baobao Xu, Junkui Qin, Jie Cai, Xuan Huang, Sihua Wang, Dong Li, Haopeng He, Xijing Clinics (Sao Paulo) Basic Research OBJECTIVE: To determine the range of motion and stability of the human cadaveric cervical spine after the implantation of a novel artificial disc and vertebra system by comparing an intact group and a fusion group. METHODS: Biomechanical tests were conducted on 18 human cadaveric cervical specimens. The range of motion and the stability index range of motion were measured to study the function and stability of the artificial disc and vertebra system of the intact group compared with the fusion group. RESULTS: In all cases, the artificial disc and vertebra system maintained intervertebral motion and reestablished vertebral height at the operative level. After its implantation, there was no significant difference in the range of motion (ROM) of C3–7 in all directions in the non-fusion group compared with the intact group (p>0.05), but significant differences were detected in flexion, extension and axial rotation compared with the fusion group (p<0.05). The ROM of adjacent segments (C(3−4), C(6−7)) of the non-fusion group decreased significantly in some directions compared with the fusion group (p<0.05). Significant differences in the C(4-6) ROM in some directions were detected between the non-fusion group and the intact group. In the fusion group, the C(4−6) ROM in all directions decreased significantly compared with the intact and non-fusion groups (p<0.01). The stability index ROM (SI-ROM) of some directions was negative in the non-fusion group, and a significant difference in SI-ROM was only found in the C(4−6) segment of the non-fusion group compared with the fusion group. CONCLUSION: An artificial disc and vertebra system could restore vertebral height and preserve the dynamic function of the surgical area and could theoretically reduce the risk of adjacent segment degeneration compared with the anterior fusion procedure. However, our results should be considered with caution because of the low power of the study. The use of a larger sample should be considered in future studies. Hospital das Clínicas da Faculdade de Medicina da Universidade de São Paulo 2015-07 2015-07 /pmc/articles/PMC4496753/ /pubmed/26222819 http://dx.doi.org/10.6061/clinics/2015(07)06 Text en Copyright © 2015 Clinics http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
spellingShingle Basic Research
Dong, Jun
Lu, Meng
Lu, Teng
Liang, Baobao
Xu, Junkui
Qin, Jie
Cai, Xuan
Huang, Sihua
Wang, Dong
Li, Haopeng
He, Xijing
Artificial disc and vertebra system: a novel motion preservation device for cervical spinal disease after vertebral corpectomy
title Artificial disc and vertebra system: a novel motion preservation device for cervical spinal disease after vertebral corpectomy
title_full Artificial disc and vertebra system: a novel motion preservation device for cervical spinal disease after vertebral corpectomy
title_fullStr Artificial disc and vertebra system: a novel motion preservation device for cervical spinal disease after vertebral corpectomy
title_full_unstemmed Artificial disc and vertebra system: a novel motion preservation device for cervical spinal disease after vertebral corpectomy
title_short Artificial disc and vertebra system: a novel motion preservation device for cervical spinal disease after vertebral corpectomy
title_sort artificial disc and vertebra system: a novel motion preservation device for cervical spinal disease after vertebral corpectomy
topic Basic Research
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4496753/
https://www.ncbi.nlm.nih.gov/pubmed/26222819
http://dx.doi.org/10.6061/clinics/2015(07)06
work_keys_str_mv AT dongjun artificialdiscandvertebrasystemanovelmotionpreservationdeviceforcervicalspinaldiseaseaftervertebralcorpectomy
AT lumeng artificialdiscandvertebrasystemanovelmotionpreservationdeviceforcervicalspinaldiseaseaftervertebralcorpectomy
AT luteng artificialdiscandvertebrasystemanovelmotionpreservationdeviceforcervicalspinaldiseaseaftervertebralcorpectomy
AT liangbaobao artificialdiscandvertebrasystemanovelmotionpreservationdeviceforcervicalspinaldiseaseaftervertebralcorpectomy
AT xujunkui artificialdiscandvertebrasystemanovelmotionpreservationdeviceforcervicalspinaldiseaseaftervertebralcorpectomy
AT qinjie artificialdiscandvertebrasystemanovelmotionpreservationdeviceforcervicalspinaldiseaseaftervertebralcorpectomy
AT caixuan artificialdiscandvertebrasystemanovelmotionpreservationdeviceforcervicalspinaldiseaseaftervertebralcorpectomy
AT huangsihua artificialdiscandvertebrasystemanovelmotionpreservationdeviceforcervicalspinaldiseaseaftervertebralcorpectomy
AT wangdong artificialdiscandvertebrasystemanovelmotionpreservationdeviceforcervicalspinaldiseaseaftervertebralcorpectomy
AT lihaopeng artificialdiscandvertebrasystemanovelmotionpreservationdeviceforcervicalspinaldiseaseaftervertebralcorpectomy
AT hexijing artificialdiscandvertebrasystemanovelmotionpreservationdeviceforcervicalspinaldiseaseaftervertebralcorpectomy