Objectives: To demonstrate, using human factors engineering (HFE), that a redesigned, pre-filled, ready-to-use, pre-asembled follitropin alfa pen can be used to administer prescribed follitropin alfa doses safely and accurately. Methods: A failure modes and effects analysis identified hazards and ha...

Descripción completa

Detalles Bibliográficos
Autores principales: Mahony, Mary C, Patterson, Patricia, Hayward, Brooke, North, Robert, Green, Dawne
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Informa Healthcare 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4496816/
https://www.ncbi.nlm.nih.gov/pubmed/25895897
http://dx.doi.org/10.1517/17425247.2015.1033395
Descripción
Sumario:Objectives: To demonstrate, using human factors engineering (HFE), that a redesigned, pre-filled, ready-to-use, pre-asembled follitropin alfa pen can be used to administer prescribed follitropin alfa doses safely and accurately. Methods: A failure modes and effects analysis identified hazards and harms potentially caused by use errors; risk-control measures were implemented to ensure acceptable device use risk management. Participants were women with infertility, their significant others, and fertility nurse (FN) professionals. Preliminary testing included ‘Instructions for Use’ (IFU) and pre-validation studies. Validation studies used simulated injections in a representative use environment; participants received prior training on pen use. Results: User performance in preliminary testing led to IFU revisions and a change to outer needle cap design to mitigate needle stick potential. In the first validation study (49 users, 343 simulated injections), in the FN group, one observed critical use error resulted in a device design modification and another in an IFU change. A second validation study tested the mitigation strategies; previously reported use errors were not repeated. Conclusions: Through an iterative process involving a series of studies, modifications were made to the pen design and IFU. Simulated-use testing demonstrated that the redesigned pen can be used to administer follitropin alfa effectively and safely.