Cargando…

Computed Tomography Angiography of Carotid Arteries and Vertebrobasilar System: A Simulation Study for Radiation Dose Reduction

Computed tomography angiography (CTA) of carotid arteries and vertebrobasilar system is a standardized procedure with excellent image quality, but radiation exposure remains a matter of concern. The aim of this study is to examine to what extent radiation dose can be lowered in relation to a standar...

Descripción completa

Detalles Bibliográficos
Autores principales: Kramer, Manuel, Ellmann, Stephan, Allmendinger, Thomas, Eller, Achim, Kammerer, Ferdinand, May, Matthias S., Baigger, João F., Uder, Michael, Lell, Michael M.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Wolters Kluwer Health 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4504529/
https://www.ncbi.nlm.nih.gov/pubmed/26131822
http://dx.doi.org/10.1097/MD.0000000000001058
Descripción
Sumario:Computed tomography angiography (CTA) of carotid arteries and vertebrobasilar system is a standardized procedure with excellent image quality, but radiation exposure remains a matter of concern. The aim of this study is to examine to what extent radiation dose can be lowered in relation to a standard protocol by simulating examinations with lower tube currents applying a dedicated software. Lower tube current was simulated by a dedicated noise insertion and reconstruction software (ReconCT). In a phantom study, true scans were performed with different dose protocols and compared to the results of simulated dose reductions of the same degree, respectively. In a patient study, 30 CTAs of supra-aortic vessels were reconstructed at a level of 100%, 75%, 50%, and 25% of the initial dose. Objective and subjective image analyses were performed. No significant noise differences between true scans and simulated scans of mimicked contrasted vessels were found. In the patient study, the quality scores of the 4 dose groups differed statistically significant; this difference vanished for the comparison of the 100% and 75% datasets after dichotomization into the categories of diagnostic and nondiagnostic image quality (P = .50). This study suggests an easy-to-implement method of simulating CTAs of carotid arteries and vertebrobasilar system with lower tube current for dose reduction by artificially adding noise to the original raw data. Lowering the radiation dose in a moderate extent to 75% of the original dose levels does not significantly alter the diagnostic image quality.