Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing

BACKGROUND: The ability to determine airflow during nasal CPAP (NCPAP) treatment without adding dead space or resistance would be useful when investigating the physiologic effects of different NCPAP systems on breathing. The aim of this study was to investigate the effect on pressure stability of di...

Descripción completa

Detalles Bibliográficos
Autores principales: Donaldsson, Snorri, Falk, Markus, Jonsson, Baldvin, Drevhammar, Thomas
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4507850/
https://www.ncbi.nlm.nih.gov/pubmed/26192188
http://dx.doi.org/10.1371/journal.pone.0133432
_version_ 1782381857637138432
author Donaldsson, Snorri
Falk, Markus
Jonsson, Baldvin
Drevhammar, Thomas
author_facet Donaldsson, Snorri
Falk, Markus
Jonsson, Baldvin
Drevhammar, Thomas
author_sort Donaldsson, Snorri
collection PubMed
description BACKGROUND: The ability to determine airflow during nasal CPAP (NCPAP) treatment without adding dead space or resistance would be useful when investigating the physiologic effects of different NCPAP systems on breathing. The aim of this study was to investigate the effect on pressure stability of different flow measuring devices at the in-line and flow-through position, using simulated neonatal breathing. METHODS: Six different flow measure devices were evaluated by recording pressure changes and imposed work of breathing for breaths with 16 and 32 ml tidal volumes. The tests were performed initially with the devices in an in line position and with 5 and 10 L/min using flow through technique, without CPAP. The flow meters were then subsequently tested with an Infant Flow CPAP system at 3, 5 and 8 cm H2O pressure using flow through technique. The quality of the recorded signals was compared graphically. RESULTS: The resistance of the measuring devices generated pressure swings and imposed work of breathing. With bias flow, the resistance also generated CPAP pressure. Three of the devices had low resistance and generated no changes in pressure stability or CPAP pressure. The two devices intended for neonatal use had the highest measured resistance. CONCLUSION: The importance of pressure stability and increased work of breathing during non-invasive respiratory support are insufficiently studied. Clinical trials using flow-through technique have not focused on pressure stability. Our results indicate that a flow-through technique might be a way forward in obtaining a sufficiently high signal quality without the added effects of rebreathing and increased work of breathing. The results should stimulate further research and the development of equipment for dynamic flow measurements in neonates.
format Online
Article
Text
id pubmed-4507850
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-45078502015-07-24 Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing Donaldsson, Snorri Falk, Markus Jonsson, Baldvin Drevhammar, Thomas PLoS One Research Article BACKGROUND: The ability to determine airflow during nasal CPAP (NCPAP) treatment without adding dead space or resistance would be useful when investigating the physiologic effects of different NCPAP systems on breathing. The aim of this study was to investigate the effect on pressure stability of different flow measuring devices at the in-line and flow-through position, using simulated neonatal breathing. METHODS: Six different flow measure devices were evaluated by recording pressure changes and imposed work of breathing for breaths with 16 and 32 ml tidal volumes. The tests were performed initially with the devices in an in line position and with 5 and 10 L/min using flow through technique, without CPAP. The flow meters were then subsequently tested with an Infant Flow CPAP system at 3, 5 and 8 cm H2O pressure using flow through technique. The quality of the recorded signals was compared graphically. RESULTS: The resistance of the measuring devices generated pressure swings and imposed work of breathing. With bias flow, the resistance also generated CPAP pressure. Three of the devices had low resistance and generated no changes in pressure stability or CPAP pressure. The two devices intended for neonatal use had the highest measured resistance. CONCLUSION: The importance of pressure stability and increased work of breathing during non-invasive respiratory support are insufficiently studied. Clinical trials using flow-through technique have not focused on pressure stability. Our results indicate that a flow-through technique might be a way forward in obtaining a sufficiently high signal quality without the added effects of rebreathing and increased work of breathing. The results should stimulate further research and the development of equipment for dynamic flow measurements in neonates. Public Library of Science 2015-07-20 /pmc/articles/PMC4507850/ /pubmed/26192188 http://dx.doi.org/10.1371/journal.pone.0133432 Text en © 2015 Donaldsson et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited.
spellingShingle Research Article
Donaldsson, Snorri
Falk, Markus
Jonsson, Baldvin
Drevhammar, Thomas
Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing
title Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing
title_full Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing
title_fullStr Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing
title_full_unstemmed Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing
title_short Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing
title_sort imposed work of breathing for flow meters with in-line versus flow-through technique during simulated neonatal breathing
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4507850/
https://www.ncbi.nlm.nih.gov/pubmed/26192188
http://dx.doi.org/10.1371/journal.pone.0133432
work_keys_str_mv AT donaldssonsnorri imposedworkofbreathingforflowmeterswithinlineversusflowthroughtechniqueduringsimulatedneonatalbreathing
AT falkmarkus imposedworkofbreathingforflowmeterswithinlineversusflowthroughtechniqueduringsimulatedneonatalbreathing
AT jonssonbaldvin imposedworkofbreathingforflowmeterswithinlineversusflowthroughtechniqueduringsimulatedneonatalbreathing
AT drevhammarthomas imposedworkofbreathingforflowmeterswithinlineversusflowthroughtechniqueduringsimulatedneonatalbreathing