Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing
BACKGROUND: The ability to determine airflow during nasal CPAP (NCPAP) treatment without adding dead space or resistance would be useful when investigating the physiologic effects of different NCPAP systems on breathing. The aim of this study was to investigate the effect on pressure stability of di...
Autores principales: | , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Public Library of Science
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4507850/ https://www.ncbi.nlm.nih.gov/pubmed/26192188 http://dx.doi.org/10.1371/journal.pone.0133432 |
_version_ | 1782381857637138432 |
---|---|
author | Donaldsson, Snorri Falk, Markus Jonsson, Baldvin Drevhammar, Thomas |
author_facet | Donaldsson, Snorri Falk, Markus Jonsson, Baldvin Drevhammar, Thomas |
author_sort | Donaldsson, Snorri |
collection | PubMed |
description | BACKGROUND: The ability to determine airflow during nasal CPAP (NCPAP) treatment without adding dead space or resistance would be useful when investigating the physiologic effects of different NCPAP systems on breathing. The aim of this study was to investigate the effect on pressure stability of different flow measuring devices at the in-line and flow-through position, using simulated neonatal breathing. METHODS: Six different flow measure devices were evaluated by recording pressure changes and imposed work of breathing for breaths with 16 and 32 ml tidal volumes. The tests were performed initially with the devices in an in line position and with 5 and 10 L/min using flow through technique, without CPAP. The flow meters were then subsequently tested with an Infant Flow CPAP system at 3, 5 and 8 cm H2O pressure using flow through technique. The quality of the recorded signals was compared graphically. RESULTS: The resistance of the measuring devices generated pressure swings and imposed work of breathing. With bias flow, the resistance also generated CPAP pressure. Three of the devices had low resistance and generated no changes in pressure stability or CPAP pressure. The two devices intended for neonatal use had the highest measured resistance. CONCLUSION: The importance of pressure stability and increased work of breathing during non-invasive respiratory support are insufficiently studied. Clinical trials using flow-through technique have not focused on pressure stability. Our results indicate that a flow-through technique might be a way forward in obtaining a sufficiently high signal quality without the added effects of rebreathing and increased work of breathing. The results should stimulate further research and the development of equipment for dynamic flow measurements in neonates. |
format | Online Article Text |
id | pubmed-4507850 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | Public Library of Science |
record_format | MEDLINE/PubMed |
spelling | pubmed-45078502015-07-24 Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing Donaldsson, Snorri Falk, Markus Jonsson, Baldvin Drevhammar, Thomas PLoS One Research Article BACKGROUND: The ability to determine airflow during nasal CPAP (NCPAP) treatment without adding dead space or resistance would be useful when investigating the physiologic effects of different NCPAP systems on breathing. The aim of this study was to investigate the effect on pressure stability of different flow measuring devices at the in-line and flow-through position, using simulated neonatal breathing. METHODS: Six different flow measure devices were evaluated by recording pressure changes and imposed work of breathing for breaths with 16 and 32 ml tidal volumes. The tests were performed initially with the devices in an in line position and with 5 and 10 L/min using flow through technique, without CPAP. The flow meters were then subsequently tested with an Infant Flow CPAP system at 3, 5 and 8 cm H2O pressure using flow through technique. The quality of the recorded signals was compared graphically. RESULTS: The resistance of the measuring devices generated pressure swings and imposed work of breathing. With bias flow, the resistance also generated CPAP pressure. Three of the devices had low resistance and generated no changes in pressure stability or CPAP pressure. The two devices intended for neonatal use had the highest measured resistance. CONCLUSION: The importance of pressure stability and increased work of breathing during non-invasive respiratory support are insufficiently studied. Clinical trials using flow-through technique have not focused on pressure stability. Our results indicate that a flow-through technique might be a way forward in obtaining a sufficiently high signal quality without the added effects of rebreathing and increased work of breathing. The results should stimulate further research and the development of equipment for dynamic flow measurements in neonates. Public Library of Science 2015-07-20 /pmc/articles/PMC4507850/ /pubmed/26192188 http://dx.doi.org/10.1371/journal.pone.0133432 Text en © 2015 Donaldsson et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited. |
spellingShingle | Research Article Donaldsson, Snorri Falk, Markus Jonsson, Baldvin Drevhammar, Thomas Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing |
title | Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing |
title_full | Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing |
title_fullStr | Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing |
title_full_unstemmed | Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing |
title_short | Imposed Work of Breathing for Flow Meters with In-Line versus Flow-Through Technique during Simulated Neonatal Breathing |
title_sort | imposed work of breathing for flow meters with in-line versus flow-through technique during simulated neonatal breathing |
topic | Research Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4507850/ https://www.ncbi.nlm.nih.gov/pubmed/26192188 http://dx.doi.org/10.1371/journal.pone.0133432 |
work_keys_str_mv | AT donaldssonsnorri imposedworkofbreathingforflowmeterswithinlineversusflowthroughtechniqueduringsimulatedneonatalbreathing AT falkmarkus imposedworkofbreathingforflowmeterswithinlineversusflowthroughtechniqueduringsimulatedneonatalbreathing AT jonssonbaldvin imposedworkofbreathingforflowmeterswithinlineversusflowthroughtechniqueduringsimulatedneonatalbreathing AT drevhammarthomas imposedworkofbreathingforflowmeterswithinlineversusflowthroughtechniqueduringsimulatedneonatalbreathing |