Therapeutic Plasma Exchange Using the Spectra Optia Cell Separator Compared With the COBE Spectra
BACKGROUND: The Spectra Optia (SPO) is a novel continuous-flow centrifugal apheresis system based on the COBE Spectra (CSP) platform. There have been few attempts to validate the advantages of the SPO. We performed a retrospective study comparing the two cell separators for therapeutic plasma exchan...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
The Korean Society for Laboratory Medicine
2015
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4510503/ https://www.ncbi.nlm.nih.gov/pubmed/26206687 http://dx.doi.org/10.3343/alm.2015.35.5.506 |
_version_ | 1782382179936894976 |
---|---|
author | Kim, Do-kyun Kim, Sinyoung Jeong, Seok Hoon Kim, Hyun Ok Kim, Hyung Jik |
author_facet | Kim, Do-kyun Kim, Sinyoung Jeong, Seok Hoon Kim, Hyun Ok Kim, Hyung Jik |
author_sort | Kim, Do-kyun |
collection | PubMed |
description | BACKGROUND: The Spectra Optia (SPO) is a novel continuous-flow centrifugal apheresis system based on the COBE Spectra (CSP) platform. There have been few attempts to validate the advantages of the SPO. We performed a retrospective study comparing the two cell separators for therapeutic plasma exchange (TPE) procedures in kidney transplant (KT) patients and seeing efficacy and safety. METHODS: We analyzed 720 TPE procedures performed between August 2012 and July 2014. Procedures included desensitization TPE before KT and TPE for the management of acute and chronic antibody-mediated graft rejection. Demographic characteristics, operational TPE variables, and laboratory data were analyzed. RESULTS: Demographic characteristics for the SPO (n=389) and CSP (n=331) groups did not differ significantly. The procedure time to exchange one plasma volume was 94.2±10.3 min in the SPO group and 100.4±11.2 min in the CSP group (P<0.001). The plasma removal efficiency (PRE) was 92.5±4.9% in the SPO group and 83.2±3.7% in the CSP group (P<0.001). There were no significant differences across the two apheresis systems for changes in hematologic parameters. CONCLUSIONS: Compared with the CSP, the SPO was associated with an improved PRE and a shorter procedure time to exchange one plasma volume. Our results in KT patients show that the SPO is superior to the CSP in TPE procedures. |
format | Online Article Text |
id | pubmed-4510503 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2015 |
publisher | The Korean Society for Laboratory Medicine |
record_format | MEDLINE/PubMed |
spelling | pubmed-45105032015-09-01 Therapeutic Plasma Exchange Using the Spectra Optia Cell Separator Compared With the COBE Spectra Kim, Do-kyun Kim, Sinyoung Jeong, Seok Hoon Kim, Hyun Ok Kim, Hyung Jik Ann Lab Med Original Article BACKGROUND: The Spectra Optia (SPO) is a novel continuous-flow centrifugal apheresis system based on the COBE Spectra (CSP) platform. There have been few attempts to validate the advantages of the SPO. We performed a retrospective study comparing the two cell separators for therapeutic plasma exchange (TPE) procedures in kidney transplant (KT) patients and seeing efficacy and safety. METHODS: We analyzed 720 TPE procedures performed between August 2012 and July 2014. Procedures included desensitization TPE before KT and TPE for the management of acute and chronic antibody-mediated graft rejection. Demographic characteristics, operational TPE variables, and laboratory data were analyzed. RESULTS: Demographic characteristics for the SPO (n=389) and CSP (n=331) groups did not differ significantly. The procedure time to exchange one plasma volume was 94.2±10.3 min in the SPO group and 100.4±11.2 min in the CSP group (P<0.001). The plasma removal efficiency (PRE) was 92.5±4.9% in the SPO group and 83.2±3.7% in the CSP group (P<0.001). There were no significant differences across the two apheresis systems for changes in hematologic parameters. CONCLUSIONS: Compared with the CSP, the SPO was associated with an improved PRE and a shorter procedure time to exchange one plasma volume. Our results in KT patients show that the SPO is superior to the CSP in TPE procedures. The Korean Society for Laboratory Medicine 2015-09 2015-07-15 /pmc/articles/PMC4510503/ /pubmed/26206687 http://dx.doi.org/10.3343/alm.2015.35.5.506 Text en © The Korean Society for Laboratory Medicine. http://creativecommons.org/licenses/by-nc/3.0/ This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/3.0/) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited. |
spellingShingle | Original Article Kim, Do-kyun Kim, Sinyoung Jeong, Seok Hoon Kim, Hyun Ok Kim, Hyung Jik Therapeutic Plasma Exchange Using the Spectra Optia Cell Separator Compared With the COBE Spectra |
title | Therapeutic Plasma Exchange Using the Spectra Optia Cell Separator Compared With the COBE Spectra |
title_full | Therapeutic Plasma Exchange Using the Spectra Optia Cell Separator Compared With the COBE Spectra |
title_fullStr | Therapeutic Plasma Exchange Using the Spectra Optia Cell Separator Compared With the COBE Spectra |
title_full_unstemmed | Therapeutic Plasma Exchange Using the Spectra Optia Cell Separator Compared With the COBE Spectra |
title_short | Therapeutic Plasma Exchange Using the Spectra Optia Cell Separator Compared With the COBE Spectra |
title_sort | therapeutic plasma exchange using the spectra optia cell separator compared with the cobe spectra |
topic | Original Article |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4510503/ https://www.ncbi.nlm.nih.gov/pubmed/26206687 http://dx.doi.org/10.3343/alm.2015.35.5.506 |
work_keys_str_mv | AT kimdokyun therapeuticplasmaexchangeusingthespectraoptiacellseparatorcomparedwiththecobespectra AT kimsinyoung therapeuticplasmaexchangeusingthespectraoptiacellseparatorcomparedwiththecobespectra AT jeongseokhoon therapeuticplasmaexchangeusingthespectraoptiacellseparatorcomparedwiththecobespectra AT kimhyunok therapeuticplasmaexchangeusingthespectraoptiacellseparatorcomparedwiththecobespectra AT kimhyungjik therapeuticplasmaexchangeusingthespectraoptiacellseparatorcomparedwiththecobespectra |