Cargando…

Significant Artifact Reduction at 1.5T and 3T MRI by the Use of a Cochlear Implant with Removable Magnet: An Experimental Human Cadaver Study

OBJECTIVE: Cochlear implants (CIs) are standard treatment for postlingually deafened individuals and prelingually deafened children. This human cadaver study evaluated diagnostic usefulness, image quality and artifacts in 1.5T and 3T magnetic resonance (MR) brain scans after CI with a removable magn...

Descripción completa

Detalles Bibliográficos
Autores principales: Wagner, Franca, Wimmer, Wilhelm, Leidolt, Lars, Vischer, Mattheus, Weder, Stefan, Wiest, Roland, Mantokoudis, Georgios, Caversaccio, Marco D.
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4511764/
https://www.ncbi.nlm.nih.gov/pubmed/26200775
http://dx.doi.org/10.1371/journal.pone.0132483