Possible Contrast Media Reduction with Low keV Monoenergetic Images in the Detection of Focal Liver Lesions: A Dual-Energy CT Animal Study

OBJECTIVE: To investigate the feasibility of dual-energy CT for contrast media (CM) reduction in the diagnosis of hypervascular and hypovascular focal liver lesions (FLL). SUBJECTS AND METHODS: The Institutional Animal Care and Use Committee approved this study. VX2 tumors were implanted in two diff...

Descripción completa

Detalles Bibliográficos
Autores principales: Chung, Yong Eun, You, Je Sung, Lee, Hye-Jeong, Lim, Joon Seok, Lee, Hye Sun, Baek, Song-Ee, Kim, Myeong-Jin
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Public Library of Science 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4512723/
https://www.ncbi.nlm.nih.gov/pubmed/26203652
http://dx.doi.org/10.1371/journal.pone.0133170
_version_ 1782382552402624512
author Chung, Yong Eun
You, Je Sung
Lee, Hye-Jeong
Lim, Joon Seok
Lee, Hye Sun
Baek, Song-Ee
Kim, Myeong-Jin
author_facet Chung, Yong Eun
You, Je Sung
Lee, Hye-Jeong
Lim, Joon Seok
Lee, Hye Sun
Baek, Song-Ee
Kim, Myeong-Jin
author_sort Chung, Yong Eun
collection PubMed
description OBJECTIVE: To investigate the feasibility of dual-energy CT for contrast media (CM) reduction in the diagnosis of hypervascular and hypovascular focal liver lesions (FLL). SUBJECTS AND METHODS: The Institutional Animal Care and Use Committee approved this study. VX2 tumors were implanted in two different segments of the liver in 13 rabbits. After 2 weeks, two phase contrast enhanced CT scans including the arterial phase (AP) and portal-venous phase (PVP) were performed three times with 24-hour intervals with three different concentrations of iodine, 300 (I(300)), 150 (I(150)) and 75 mg I/mL (I(75)). The mean HU and standard deviation (SD) were measured in the liver, the hypervascular portion of the VX2 tumor which represented hypervascular tumors, and the central necrotic area of the VX2 tumor which represented hypovascular tumors in 140kVp images with I(300) as a reference standard and in monoenergetic images (between 40keV and 140keV) with I(150) and I(75). The contrast-to-noise ratio (CNR) for FLLs and the ratio of the CNRs (CNR(ratio)) between monoenergetic image sets with I(150) and I(75), and the reference standard were calculated. RESULTS: For hypervascular lesions, the CNR(ratio) was not statistically different from 1.0 between 40keV and 70keV images with I(150), whereas the CNR(ratio) was significantly lower than 1.0 in all keV images with I(75). For hypovascular lesions, the CNR(ratio) was similar to or higher than 1.0 between 40keV and 80keV with I(150) and between 40keV and 70keV with I(75). CONCLUSIONS: With dual-energy CT, the total amount of CM might be halved in the diagnosis of hypervascular FLLs and reduced to one-fourth in the diagnosis of hypovascular FLLs, while still preserving CNRs.
format Online
Article
Text
id pubmed-4512723
institution National Center for Biotechnology Information
language English
publishDate 2015
publisher Public Library of Science
record_format MEDLINE/PubMed
spelling pubmed-45127232015-07-24 Possible Contrast Media Reduction with Low keV Monoenergetic Images in the Detection of Focal Liver Lesions: A Dual-Energy CT Animal Study Chung, Yong Eun You, Je Sung Lee, Hye-Jeong Lim, Joon Seok Lee, Hye Sun Baek, Song-Ee Kim, Myeong-Jin PLoS One Research Article OBJECTIVE: To investigate the feasibility of dual-energy CT for contrast media (CM) reduction in the diagnosis of hypervascular and hypovascular focal liver lesions (FLL). SUBJECTS AND METHODS: The Institutional Animal Care and Use Committee approved this study. VX2 tumors were implanted in two different segments of the liver in 13 rabbits. After 2 weeks, two phase contrast enhanced CT scans including the arterial phase (AP) and portal-venous phase (PVP) were performed three times with 24-hour intervals with three different concentrations of iodine, 300 (I(300)), 150 (I(150)) and 75 mg I/mL (I(75)). The mean HU and standard deviation (SD) were measured in the liver, the hypervascular portion of the VX2 tumor which represented hypervascular tumors, and the central necrotic area of the VX2 tumor which represented hypovascular tumors in 140kVp images with I(300) as a reference standard and in monoenergetic images (between 40keV and 140keV) with I(150) and I(75). The contrast-to-noise ratio (CNR) for FLLs and the ratio of the CNRs (CNR(ratio)) between monoenergetic image sets with I(150) and I(75), and the reference standard were calculated. RESULTS: For hypervascular lesions, the CNR(ratio) was not statistically different from 1.0 between 40keV and 70keV images with I(150), whereas the CNR(ratio) was significantly lower than 1.0 in all keV images with I(75). For hypovascular lesions, the CNR(ratio) was similar to or higher than 1.0 between 40keV and 80keV with I(150) and between 40keV and 70keV with I(75). CONCLUSIONS: With dual-energy CT, the total amount of CM might be halved in the diagnosis of hypervascular FLLs and reduced to one-fourth in the diagnosis of hypovascular FLLs, while still preserving CNRs. Public Library of Science 2015-07-23 /pmc/articles/PMC4512723/ /pubmed/26203652 http://dx.doi.org/10.1371/journal.pone.0133170 Text en © 2015 Chung et al http://creativecommons.org/licenses/by/4.0/ This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are properly credited.
spellingShingle Research Article
Chung, Yong Eun
You, Je Sung
Lee, Hye-Jeong
Lim, Joon Seok
Lee, Hye Sun
Baek, Song-Ee
Kim, Myeong-Jin
Possible Contrast Media Reduction with Low keV Monoenergetic Images in the Detection of Focal Liver Lesions: A Dual-Energy CT Animal Study
title Possible Contrast Media Reduction with Low keV Monoenergetic Images in the Detection of Focal Liver Lesions: A Dual-Energy CT Animal Study
title_full Possible Contrast Media Reduction with Low keV Monoenergetic Images in the Detection of Focal Liver Lesions: A Dual-Energy CT Animal Study
title_fullStr Possible Contrast Media Reduction with Low keV Monoenergetic Images in the Detection of Focal Liver Lesions: A Dual-Energy CT Animal Study
title_full_unstemmed Possible Contrast Media Reduction with Low keV Monoenergetic Images in the Detection of Focal Liver Lesions: A Dual-Energy CT Animal Study
title_short Possible Contrast Media Reduction with Low keV Monoenergetic Images in the Detection of Focal Liver Lesions: A Dual-Energy CT Animal Study
title_sort possible contrast media reduction with low kev monoenergetic images in the detection of focal liver lesions: a dual-energy ct animal study
topic Research Article
url https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4512723/
https://www.ncbi.nlm.nih.gov/pubmed/26203652
http://dx.doi.org/10.1371/journal.pone.0133170
work_keys_str_mv AT chungyongeun possiblecontrastmediareductionwithlowkevmonoenergeticimagesinthedetectionoffocalliverlesionsadualenergyctanimalstudy
AT youjesung possiblecontrastmediareductionwithlowkevmonoenergeticimagesinthedetectionoffocalliverlesionsadualenergyctanimalstudy
AT leehyejeong possiblecontrastmediareductionwithlowkevmonoenergeticimagesinthedetectionoffocalliverlesionsadualenergyctanimalstudy
AT limjoonseok possiblecontrastmediareductionwithlowkevmonoenergeticimagesinthedetectionoffocalliverlesionsadualenergyctanimalstudy
AT leehyesun possiblecontrastmediareductionwithlowkevmonoenergeticimagesinthedetectionoffocalliverlesionsadualenergyctanimalstudy
AT baeksongee possiblecontrastmediareductionwithlowkevmonoenergeticimagesinthedetectionoffocalliverlesionsadualenergyctanimalstudy
AT kimmyeongjin possiblecontrastmediareductionwithlowkevmonoenergeticimagesinthedetectionoffocalliverlesionsadualenergyctanimalstudy