A simple and easy in vitro model to test the efficacy of IV lines' needleless connectors against contamination
BACKGROUND: Hub colonization after manipulation is responsible for 29% to 60% of catheter-related bloodstream infections (C-RBSI). Prevention can be achieved by the use of hub connectors, but its efficacy is generally based on instillation of high concentrations of microorganisms, which do not refle...
Autores principales: | , , , , |
---|---|
Formato: | Online Artículo Texto |
Lenguaje: | English |
Publicado: |
Springer International Publishing
2014
|
Materias: | |
Acceso en línea: | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4513010/ https://www.ncbi.nlm.nih.gov/pubmed/26266924 http://dx.doi.org/10.1186/s40635-014-0027-9 |
_version_ | 1782382574091370496 |
---|---|
author | Guembe, María Pérez-Granda, María Jesús Alcalá, Luis Martín-Rabadán, Pablo Bouza, Emilio |
author_facet | Guembe, María Pérez-Granda, María Jesús Alcalá, Luis Martín-Rabadán, Pablo Bouza, Emilio |
author_sort | Guembe, María |
collection | PubMed |
description | BACKGROUND: Hub colonization after manipulation is responsible for 29% to 60% of catheter-related bloodstream infections (C-RBSI). Prevention can be achieved by the use of hub connectors, but its efficacy is generally based on instillation of high concentrations of microorganisms, which do not reflect the real contamination in daily practice. Our purpose was to create an in vitro model lasting long enough to be used for the comparison of the efficacy between various connectors against contamination simulating the real daily handling. METHODS: The model consisted of 40 blood culture bottles with an inserted cannula with a needle-free closed connector. Twice a day, each line was manipulated while instilling 1 mL of two different fluids (saline and propofol). We manipulated the bottles as follows: ten bottles with clean gloves and disinfecting connectors with alcohol (controls), ten bottles with hands (no gloves), ten bottles with gloves impregnated with a 0.5 McFarland (MF) solution of Staphylococcus aureus (SA), and ten bottles with gloves impregnated with a 0.05 MF solution of SA. The bottles were incubated in a BACTEC System at 37°C under continuous agitation up to 10 days. When a bottle turned positive, 100 μL of the fluid was cultured and incubated followed by microorganism identification using standard procedures. RESULTS: Overall, all bottles in the control group were negative at the end of the incubation time. In the three contamination experiments, almost all (38/40) bottles were positive during the incubation time. We only found differences regarding the median time to positivity (interquartile range (IQR)) between saline and propofol in the manipulation with SA 0.05 MF: 240 h (154.82 to 360.00) vs. 66 h (58.01 to 69.11), p = 0.008. CONCLUSIONS: A daily connector handling with 0.05 McFarland S. aureus solution while instilling saline proved to be a useful model lasting long enough to be used for the comparison of the efficacy of different types of closed needleless connectors against contamination. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1186/s40635-014-0027-9) contains supplementary material, which is available to authorized users. |
format | Online Article Text |
id | pubmed-4513010 |
institution | National Center for Biotechnology Information |
language | English |
publishDate | 2014 |
publisher | Springer International Publishing |
record_format | MEDLINE/PubMed |
spelling | pubmed-45130102015-07-27 A simple and easy in vitro model to test the efficacy of IV lines' needleless connectors against contamination Guembe, María Pérez-Granda, María Jesús Alcalá, Luis Martín-Rabadán, Pablo Bouza, Emilio Intensive Care Med Exp Research BACKGROUND: Hub colonization after manipulation is responsible for 29% to 60% of catheter-related bloodstream infections (C-RBSI). Prevention can be achieved by the use of hub connectors, but its efficacy is generally based on instillation of high concentrations of microorganisms, which do not reflect the real contamination in daily practice. Our purpose was to create an in vitro model lasting long enough to be used for the comparison of the efficacy between various connectors against contamination simulating the real daily handling. METHODS: The model consisted of 40 blood culture bottles with an inserted cannula with a needle-free closed connector. Twice a day, each line was manipulated while instilling 1 mL of two different fluids (saline and propofol). We manipulated the bottles as follows: ten bottles with clean gloves and disinfecting connectors with alcohol (controls), ten bottles with hands (no gloves), ten bottles with gloves impregnated with a 0.5 McFarland (MF) solution of Staphylococcus aureus (SA), and ten bottles with gloves impregnated with a 0.05 MF solution of SA. The bottles were incubated in a BACTEC System at 37°C under continuous agitation up to 10 days. When a bottle turned positive, 100 μL of the fluid was cultured and incubated followed by microorganism identification using standard procedures. RESULTS: Overall, all bottles in the control group were negative at the end of the incubation time. In the three contamination experiments, almost all (38/40) bottles were positive during the incubation time. We only found differences regarding the median time to positivity (interquartile range (IQR)) between saline and propofol in the manipulation with SA 0.05 MF: 240 h (154.82 to 360.00) vs. 66 h (58.01 to 69.11), p = 0.008. CONCLUSIONS: A daily connector handling with 0.05 McFarland S. aureus solution while instilling saline proved to be a useful model lasting long enough to be used for the comparison of the efficacy of different types of closed needleless connectors against contamination. ELECTRONIC SUPPLEMENTARY MATERIAL: The online version of this article (doi:10.1186/s40635-014-0027-9) contains supplementary material, which is available to authorized users. Springer International Publishing 2014-11-07 /pmc/articles/PMC4513010/ /pubmed/26266924 http://dx.doi.org/10.1186/s40635-014-0027-9 Text en © Guembe et al.; licensee Springer. 2014 This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. |
spellingShingle | Research Guembe, María Pérez-Granda, María Jesús Alcalá, Luis Martín-Rabadán, Pablo Bouza, Emilio A simple and easy in vitro model to test the efficacy of IV lines' needleless connectors against contamination |
title | A simple and easy in vitro model to test the efficacy of IV lines' needleless connectors against contamination |
title_full | A simple and easy in vitro model to test the efficacy of IV lines' needleless connectors against contamination |
title_fullStr | A simple and easy in vitro model to test the efficacy of IV lines' needleless connectors against contamination |
title_full_unstemmed | A simple and easy in vitro model to test the efficacy of IV lines' needleless connectors against contamination |
title_short | A simple and easy in vitro model to test the efficacy of IV lines' needleless connectors against contamination |
title_sort | simple and easy in vitro model to test the efficacy of iv lines' needleless connectors against contamination |
topic | Research |
url | https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4513010/ https://www.ncbi.nlm.nih.gov/pubmed/26266924 http://dx.doi.org/10.1186/s40635-014-0027-9 |
work_keys_str_mv | AT guembemaria asimpleandeasyinvitromodeltotesttheefficacyofivlinesneedlelessconnectorsagainstcontamination AT perezgrandamariajesus asimpleandeasyinvitromodeltotesttheefficacyofivlinesneedlelessconnectorsagainstcontamination AT alcalaluis asimpleandeasyinvitromodeltotesttheefficacyofivlinesneedlelessconnectorsagainstcontamination AT martinrabadanpablo asimpleandeasyinvitromodeltotesttheefficacyofivlinesneedlelessconnectorsagainstcontamination AT bouzaemilio asimpleandeasyinvitromodeltotesttheefficacyofivlinesneedlelessconnectorsagainstcontamination AT guembemaria simpleandeasyinvitromodeltotesttheefficacyofivlinesneedlelessconnectorsagainstcontamination AT perezgrandamariajesus simpleandeasyinvitromodeltotesttheefficacyofivlinesneedlelessconnectorsagainstcontamination AT alcalaluis simpleandeasyinvitromodeltotesttheefficacyofivlinesneedlelessconnectorsagainstcontamination AT martinrabadanpablo simpleandeasyinvitromodeltotesttheefficacyofivlinesneedlelessconnectorsagainstcontamination AT bouzaemilio simpleandeasyinvitromodeltotesttheefficacyofivlinesneedlelessconnectorsagainstcontamination |