Comparison of cumulative dissipated energy between the Infiniti and Centurion phacoemulsification systems

PURPOSE: To compare cumulative dissipated energy between two phacoemulsification machines. SETTING: An ambulatory surgical center, Honolulu, Hawaii, USA. DESIGN: Retrospective chart review. METHODS: A total of 2,077 consecutive cases of cataract extraction by phacoemulsification performed by five su...

Descripción completa

Detalles Bibliográficos
Autores principales: Chen, Ming, Anderson, Erik, Hill, Geoffrey, Chen, John J, Patrianakos, Thomas
Formato: Online Artículo Texto
Lenguaje:English
Publicado: Dove Medical Press 2015
Materias:
Acceso en línea:https://www.ncbi.nlm.nih.gov/pmc/articles/PMC4516204/
https://www.ncbi.nlm.nih.gov/pubmed/26229430
http://dx.doi.org/10.2147/OPTH.S88225
Descripción
Sumario:PURPOSE: To compare cumulative dissipated energy between two phacoemulsification machines. SETTING: An ambulatory surgical center, Honolulu, Hawaii, USA. DESIGN: Retrospective chart review. METHODS: A total of 2,077 consecutive cases of cataract extraction by phacoemulsification performed by five surgeons from November 2012 to November 2014 were included in the study; 1,021 consecutive cases were performed using the Infiniti Vision System, followed by 1,056 consecutive cases performed using the Centurion Vision System. RESULTS: The Centurion phacoemulsification system required less energy to remove a cataractous lens with an adjusted average energy reduction of 38% (5.09 percent-seconds) (P<0.001) across all surgeons in comparison to the Infiniti phacoemulsification system. The reduction in cumulative dissipated energy was statistically significant for each surgeon, with a range of 29%–45% (2.25–12.54 percent-seconds) (P=0.005–<0.001). Cumulative dissipated energy for both the Infiniti and Centurion systems varied directly with patient age, increasing an average of 2.38 percent-seconds/10 years. CONCLUSION: The Centurion phacoemulsification system required less energy to remove a cataractous lens in comparison to the Infiniti phacoemulsification system.